Zusammenfassung.

Ziel: Erarbeitung der Vorteile und Nachteile bei Anwendung von runden und spitzen Laserapplikatorsystemen im Rahmen der laserinduzierten interstitiellen Thermotherapie. Material/Methode: Neben 6 In-vitro-Versuchen wurden 12 Patienten mit Lebermetastasen kolorektaler Karzinome mit Lokalisation unter dem Zwerchfell bzw. am Leberrand oder in der Nähe von großen Lebergefäßen mit der laserinduzierten Thermotherapie (LITT) behandelt. 5 Patienten wurden mit spitzen, 7 Patienten mit runden Laser-Applikationssets behandelt. Nach CT-gesteuerter Punktion der Metastasen erfolgte die LITT unter MRT-Kontrolle mit durchschnittlich 25 W über 20 Minuten. Ergebnisse: Bei entsprechender Lokalisation der Lebermetastasen unter dem Zwerchfell oder am Leberrand führt die Anwendung der runden Laserapplikatorsysteme durch den verringerten Totraum durch Wegfall der Spitze (0,5-1 cm bei spitzen Applikator-Systemen) zu einer besseren Positionierung des Lasers in der Metastase und damit zu einem besseren Therapieergebnis i. S. einer vollständigen Tumorablation. Bei allen Patienten, die mit runden Applikatorsystemen behandelt wurden, konnte die Metastase vollständig entfernt werden. Bei 3 der 5 Patienten mit spitzem Applikatorsystem zeigte die Nachkontrolle am 2. Tag nach der LITT einen vitalen Tumorrand auf der dem Zwerchfell zugewandten oder dem Gefäß anliegenden Seite. Schlussfolgerung: Bei Lokalisation von Lebermetastasen an Grenzflächen der Leber und unterm Zwerchfell führt die Anwendung von runden Laserapplikationssystemen zu einer sichereren Metastasenablation als bei spitzen Applikationssets durch die Möglichkeit einer besseren Positionierung des Lasers in der Mitte der Metastase.

Laser-induced Thermotherapy (LITT). Application of Round and Sharp Laser Systems.

Purpose: To assess the advantages and disadvantages by using round and sharp laser application systems during laser-induced thermotherapy. Method: 6 in vitro examinations were performed and 12 patients with liver metastases of colorectal carcinoma were treated with laser-induced thermotherapy. All lesions were closely localised to the diaphragm, liver capsule and large intrahepatic vessels. Five patients were treated with the sharp and 7 patients with the round applicator system. After CT-guided insertion of the catheter system into the metastasis laser therapy was performed under MRI control (approx. 25 W, 20 min). Results: Reduction of the distance between the laser applicator and rounded laser catheter tipp (approx. 0.5-1 cm) leads to better placement of the laser applicator in metastases located near the diaphragm, liver capsule and large intrahepatic vessels. Improved therapy results due to complete ablation of metastases were obtained. All patients treated with the round system had complete ablation of metastasis. In 3 of 5 patients treated with the sharp system, MRI control 2 days after therapy showed a residual tumour margin close to the diaphragm or to an intrahepatic vessel. Discussion: Improved ablation can be obtained by closer placement of a round laser catheter in liver metastasis located near the diaphragm, liver capsule and large intrahepatic vessels.

Literatur

  • 1 Vogl T J, Muller P K, Hammerstingl R, Weinhold N, Mack M G, Philipp C, Deimling M, Beuthan J, Pegios W, Riess H. et al . Malignant liver tumors treated with MR imaging-guided laser-induced thermotherapy: technique and prospective results.  Radiology. 1995;  196 257-265
  • 2 Bremer C, Alikemper T, Menzel J, Sulkowski U, Rummeny E, Reimer P. Preliminary clinical experience with laser-induced interstitial thermotherapy in patients with hepatocellular carcinoma.  Magn Reson Imaging. 1998;  8 235-239
  • 3 Vogl T J, Mack M, Straub R, Eichler K, Engelmann K, Roggan A, Zangos S. Percutaneous interstitial thermotherapy of malignant liver tumors.  RöFo. 2000;  172 12-22
  • 4 Vogl T J. Verification of MR-Thermometrie.  RöFo. 1988;  169 (2) 182-188
  • 5 Hosten N, Kreissig R, Puls R, Amthauer H, Beier J, Rohlfing T, Stroszczynski C, Herbel A, Lemke A J, Felix R. Fusion of CT and PET data: methods and clinical relevance for planning laser-induced thermotherapy of liver metastases.  RöFo. 2000;  172 630-635
  • 6 Reither K, Wacker F, Ritz J P, Isbert C, Germer C T, Roggan A, Wendt M, Wolf K J. Laser-induced thermotherapy (LITT) for liver metastasis in an open 0.2 T MRI.  RöFo. 2000;  172 175-178

Dr. Ralf Puls

Humboldt Universität zu Berlin
Med. Fakultät Charité, Campus Virchow Klinikum
Strahlenklinik und Poliklinik

Augustenburger Platz 1

13353 Berlin

Phone: ++49 (030) 450 57001

Fax: ++49 (030) 450 57901

Email: ralf.puls@charite.de

    >