Thromb Haemost 1982; 47(02): 101-103
DOI: 10.1055/s-0038-1657138
Original Article
Schattauer GmbH Stuttgart

The Effects of Inaccurate Blood Sample Volume on Prothrombin Time (PT) and Activated Partial Thromboplastin Time (aPTT)

Powers Peterson
The Department of Pathology, The New York Hospital-Cornell University Medical College, New York, N.Y., USA
,
Eugene L Gottfried
The Department of Pathology, The New York Hospital-Cornell University Medical College, New York, N.Y., USA
› Author Affiliations
Further Information

Publication History

Received 21 December 1981

Accepted 18 January 1982

Publication Date:
13 July 2018 (online)

Summary

The results of determinations of the prothrombin time (PT) and the activated partial thromboplastin time (aPTT) are frequently used to assess hemostatic function. Accurate results for these laboratory tests depend on many variables, one of which is the ratio of plasma to anticoagulant. We studied 12 patients and 4 normal subjects to determine the effects of sample volume on PT and aPTT. We conclude that underfilling may produce profound effects, particularly on the aPTT. In contrast, overfilling rarely affects the results. The greatest effects of sample volume were observed in specimens in which the true PT or aPTT was elevated. A normal PT or aPTT result on any specimen, regardless of sample volume, strongly suggests that the true value is normal.

 
  • References

  • 1 Denson KWE. The preparation of general reagents and coagulation factors. In: Human Blood Coagulation, Haemostasis and Thrombosis. Biggs R. (ed). 2. 659 Appendix I (11) Blackwell Scientific Publications; Oxford: 1976
  • 2 Linke EG, Henry JB, Statland E. Theory and practice of laboratory technique. In: Clinical Diagnosis and Management by Laboratory Methods. Henry JB. (ed). 16. saunders; Philadelphia: 53-56 1979
  • 3 Quick AJ. Hemorrhagic Diseases and Thrombosis. 2. Lea and Febiger; Philadelphia: 391-395 1966
  • 4 Lancer Coagulyzer Instruction Manual. pp 4-10-4-13,4-19-4-22. Sherwood Medical Industries. St Louis: 1974
  • 5 Spaet TH. Letter to the editor. Case 20-1979: False prolongation of prothrombin time in polycythemia. N Engl J Med 1979; 301: 503
  • 6 Bunn HF. Letter to the Editor. (Reply.). N Engl J Med 1979; 301: 503
  • 7 Humphreys RE, McPhedran P. False elevation of partial thromboplastin time and prothrombin time. JAMA 1970; 214: 1702-1704
  • 8 Ingram GIC, Hills M. The prothrombin time test: effect of varying citrate concentration. Thromb Haemostas 1976; 36: 230-236
  • 9 Brandt JT, Triplett DA. Laboratory monitoring of heparin: effect of reagents and instruments on the activated partial thromboplastin time. Am J Clin Pathol 1981; 76 (Suppl. 00) 530-537