Int J Angiol 2017; 26(02): 116-120
DOI: 10.1055/s-0037-1598252
Original Article
Thieme Medical Publishers 333 Seventh Avenue, New York, NY 10001, USA.

Safety and Efficacy of a Novel “Hybrid Closure” Technique in Large-Bore Arteriotomies

Michael K. Amponsah
1   Division of Cardiology, Newark Beth Israel Medical Center, Newark, New Jersey
,
Rajiv Tayal
1   Division of Cardiology, Newark Beth Israel Medical Center, Newark, New Jersey
,
Zain Khakwani
1   Division of Cardiology, Newark Beth Israel Medical Center, Newark, New Jersey
,
Michael Sinclair
1   Division of Cardiology, Newark Beth Israel Medical Center, Newark, New Jersey
,
Najam Wasty
1   Division of Cardiology, Newark Beth Israel Medical Center, Newark, New Jersey
› Author Affiliations
Further Information

Publication History

Publication Date:
25 February 2017 (online)

Abstract

The “preclose” technique employing two Perclose (P) devices is well established for large-bore artery (LBA) hemostasis. Occasionally, only one Perclose deploys successfully during the initial preclose because of arterial calcification necessitating the use of the crossover balloon technique to achieve hemostasis at the LBA. We sought to determine if the combined use of one Perclose and either one Angioseal or one Mynx vascular closure device (VCD) is a safe alternative closure technique large-bore arteriotomy closure.

In total, 40 patients underwent high-risk percutaneous coronary intervention (HRPCI) with Impella support, of whom 38 had common femoral artery (CFA) arteriotomies and 2 underwent percutaneous axillary arteriotomy (AA). Prior to Impella insertion, one Perclose device was predeployed. At the end of HRPCI, Impella was removed and a 0.035″ wire was inserted through the Impella sheath. This sheath was then withdrawn over the wire, and partially deployed Perclose was fully deployed. A 6-Fr sheath was advanced over a 0.035″ wire into the CFA or AA, achieving hemostasis and reducing the LBA to a 6-Fr size. The 6-Fr arteriotomy was closed with a 6-Fr Mynx or Angioseal VCD. Patients were followed at day 1 and day 30.

Hybrid closure was successful in 38 of 40 cases. In one case of Mynx balloon rupture, hemostasis was achieved with heparin reversal and manual compression. In the case of Perclose failure, crossover balloon tamponade at arteriotomy site and external manual compression achieved hemostasis. Patients were free of complications at day 1 and day 30.

Hybrid closure with one Perclose and either one Mynx or one Angioseal VCD is safe and effective for LBA closure.

 
  • References

  • 1 Scheinert D, Sievert H, Turco MA , et al. The safety and efficacy of an extravascular, water-soluble sealant for vascular closure: initial clinical results for Mynx. Catheter Cardiovasc Interv 2007; 70 (5) 627-633
  • 2 Veasey RA, Large JK, Silberbauer J , et al. A randomised controlled trial comparing StarClose and AngioSeal vascular closure devices in a district general hospital--the SCOAST study. Int J Clin Pract 2008; 62 (6) 912-918
  • 3 Martin JL, Pratsos A, Magargee E , et al. A randomized trial comparing compression, Perclose Proglide and Angio-Seal VIP for arterial closure following percutaneous coronary intervention: the CAP trial. Catheter Cardiovasc Interv 2008; 71 (1) 1-5
  • 4 Jaffan AA, Prince EA, Hampson CO, Murphy TP. The preclose technique in percutaneous endovascular aortic repair: a systematic literature review and meta-analysis. Cardiovasc Intervent Radiol 2013; 36 (3) 567-577
  • 5 Wei R, Xiong J, Guo W, Liu X, Zhang H, Jia X. Effectiveness comparison between double Perclose ProGlide crossing suture and traditional suture for closure of puncture sites in endovascular aortic repair [in Chinese]. Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi 2012; 26 (8) 968-971
  • 6 Genereux P, Kodali S, Leon MB , et al. Clinical outcomes using a new crossover balloon occlusion technique for percutaneous closure after transfemoral aortic valve implantation. JACC Cardiovasc Interv 2011; 4 (8) 861-867
  • 7 Hermanides RS, Ottervanger JP, Dambrink JH , et al; Zwolle Myocardial Infarction Study Group. Closure device or manual compression in patients undergoing percutaneous coronary intervention: a randomized comparison. J Invasive Cardiol 2010; 22 (12) 562-566
  • 8 Lee WA, Brown MP, Nelson PR, Huber TS. Total percutaneous access for endovascular aortic aneurysm repair (“Preclose” technique). J Vasc Surg 2007; 45 (6) 1095-1101
  • 9 Abi Rafeh N, Quevedo HC, DeAndrade KB, Yalvac EA, Dehghani H, Arain SA. The double Angio-Seal technique for arterial closure following large-bore access. J Invasive Cardiol 2013; 25 (8) 412-414