Microvascular imaging versus CEUS in the characterization of renal masses: preliminary experience in a tertiary care referral university hospital
Mikrovaskuläre Bildgebung versus CEUS bei der Charakterisierung renaler Raumforderungen: Erste Erfahrungen in einem Universitätsklinikum der Maximalversorgung
Abstract
Purpose
Evaluation of the possible role of microvascular imaging (MI) for the detection of vascularization of renal lesions, while comparing it with contrast-enhanced ultrasound (CEUS).
Materials and Methods
137 patients (160 renal lesions, 64 solid and 96 cystic) were enrolled in this single-center, retrospective, observational study. For solid masses, both the intralesional and the peripheral vascularization was analyzed and quantified by MI and then compared to peak enhancement on CEUS. Regarding cystic lesions, we evaluated the presence or absence of vascularization within the cystic wall and/or septa. MI results were compared with CEUS findings using Pearson’s Chi square test. An optimal size cut-off for renal lesions was determined with the Youden test.
Results
For solid lesions, a statistically significant correlation (p <0.05) was observed between the MI parameters and the peak enhancement on CEUS. The detection rate (DR) for lesional vascularization on MI was 87.5%, while if we consider only lesions larger than the optimal cut-off (14mm), the DR increases to 98%. In cystic lesions, the MI showed a high specificity (93.9%) in predicting CEUS results and a high positive predictive value (84.2%). The concordance was 100% in Bosniak I lesions and 80% in Bosniak IV lesions, while it was lower for the other classes. Furthermore, we found a statistically significant correlation (p <0.05) between Bosniak grade and lesional vascularity on MI.
Conclusion
Our preliminary study shows that MI cannot replace CEUS, but could reduce its use, especially in solid lesions larger than 14mm and in cysts classified as Bosniak IV, a goal that is particularly important in an active surveillance setting.
Zusammenfassung
Ziel
Bewertung der Rolle der mikrovaskulären Bildgebung (MI) bei der Erkennung der Vaskularisierung von Nierenläsionen im Vergleich zum kontrastverstärkten Ultraschall (CEUS).
Material und Methoden
137 Patienten (160 Nierenläsionen, 64 solide und 96 zystische) wurden in diese retrospektive Beobachtungsstudie in einem einzelnen Zentrum aufgenommen. Bei soliden Raumforderungen wurden sowohl die intraläsionale als auch die periphere Vaskularisierung mittels MI analysiert und quantifiziert und dann mit dem Peak-Enhancement im CEUS verglichen. Bei zystischen Läsionen wurde das Vorhandensein oder das Fehlen einer Vaskularisierung innerhalb der zystischen Wand und/oder Septen untersucht. Die MI-Ergebnisse wurden mit den CEUS-Befunden unter Verwendung des Chi-Quadrat-Tests nach Pearson verglichen. Ein optimaler Cut-off für die Größe von Nierenläsionen wurde mit dem Youden-Test bestimmt.
Ergebnisse
Bei soliden Läsionen wurde eine statistisch signifikante Korrelation (p < 0,05) zwischen den MI-Parametern und dem Peak-Enhancement im CEUS beobachtet. Die Detektionsrate (DR) für die vaskularisierte Läsion bei der MI betrug 87,5%, während die DR bei alleiniger Betrachtung von Läsionen, die größer als der optimale Cut-off (14mm) sind, auf 98% steigt. Bei zystischen Läsionen zeigte die MI eine hohe Spezifität (93,9%) für die Prädiktion von CEUS-Befunden und einen hohen positiven Vorhersagewert (84,2%). Die Übereinstimmung betrug 100% bei Bosniak-I-Läsionen und 80% bei Bosniak-IV-Läsionen, während sie bei den anderen Klassen niedriger war. Darüber hinaus fanden wir bei der MI eine statistisch signifikante Korrelation zwischen dem Bosniak-Grad und der Vaskularität der Läsion (p < 0,05).
Schlussfolgerung
Unsere vorläufige Studie zeigt, dass MI den CEUS nicht ersetzen kann, aber dessen Einsatz reduzieren könnte. Dies ist insbesondere bei soliden Läsionen, die größer als 14mm sind und bei Zysten mit Bosniak-IV-Klassifikation der Fall – was insbesondere in einer aktiven Überwachungssituation wichtig sein könnte.
Publication History
Received: 12 August 2024
Accepted after revision: 28 January 2025
Article published online:
31 March 2025
© 2025. Thieme. All rights reserved.
Georg Thieme Verlag KG
Oswald-Hesse-Straße 50, 70469 Stuttgart, Germany
-
References
- 1 Capitanio U, Bensalah K, Bex A. et al. Epidemiology of Renal Cell Carcinoma. Eur Urol 2019; 75 (01) 74-84
- 2 Kim JH, Li S, Khandwala Y. et al. Association of Prevalence of Benign Pathologic Findings After Partial Nephrectomy With Preoperative Imaging Patterns in the United States From 2007 to 2014. JAMA Surg 2019; 154 (03) 225-231
- 3 Patel HD, Johnson MH, Pierorazio PM. et al. Diagnostic Accuracy and Risks of Biopsy in the Diagnosis of a Renal Mass Suspicious for Localized Renal Cell Carcinoma: Systematic Review of the Literature. J Urol 2016; 195 (05) 1340-1347
- 4 Marconi L, Dabestani S, Lam TB. et al. Systematic Review and Meta-analysis of Diagnostic Accuracy of Percutaneous Renal Tumour Biopsy. Eur Urol 2016; 69 (04) 660-673
- 5 EAU Guidelines. Edn. presented at the EAU Annual Congress Paris 2024. ISBN 978–94–92671–23–3. EAU Guidelines Office, Arnhem, The Netherlands. https://uroweb.org/guidelines
- 6 Cantisani V, Bertolotto M, Clevert DA. et al. EFSUMB 2020 Proposal for a Contrast-Enhanced Ultrasound-Adapted Bosniak Cyst Categorization – Position Statement. EFSUMB 2020 – Vorschlag für eine an den kontrastverstärkten Ultraschall adaptierte Bosniak-Klassifikation von Zysten – Eine Stellungnahme. Ultraschall in Med 2021; 42 (02) 154-166
- 7 Furrer MA, Spycher SCJ, Büttiker SM. et al. Comparison of the Diagnostic Performance of Contrast-enhanced Ultrasound with That of Contrast-enhanced Computed Tomography and Contrast-enhanced Magnetic Resonance Imaging in the Evaluation of Renal Masses: A Systematic Review and Meta-analysis. Eur Urol Oncol 2020; 3 (04) 464-473
- 8 Qiu X, Zhao Q, Ye Z. et al. How does contrast-enhanced ultrasonography influence Bosniak classification for complex cystic renal mass compared with conventional ultrasonography?. Medicine (Baltimore) 2020; 99 (07) e19190
- 9 Bertelli E, Palombella A, Sessa F. et al. Contrast-enhanced ultrasound (CEUS) imaging for active surveillance of small renal masses. World J Urol 2021; 39 (08) 2853-2860
- 10 Visalli C, Vinci SL, Mondello S. et al. Microvascular imaging ultrasound (MicroV) and power Doppler vascularization analysis in a pediatric population with early scrotal pain onset. Jpn J Radiol 2022; 40: 192-201
- 11 Spiesecke P, Schmidt J, Peters R. et al. Assessment of quantitative microflow Vascular Index in testicular cancer. Eur J Radiol 2024; 176: 111513
- 12 Shen TT, Xue JL. Impact of a novel ultrasound microvascular imaging and elastography on prostate cancer classification. Transl Androl Urol 2019; 8 (06) 696-702
- 13 Leong JY, Wessner CE, Kramer MR. et al. Superb Microvascular Imaging Improves Detection of Vascularity in Indeterminate Renal Masses. J Ultrasound Med 2020; 39 (10) 1947-1955
- 14 Chen M, Fu X, Shen Y. Evaluation of Multimode Color Doppler Flow Imaging in the Diagnosis of Solid Renal Tumor. Contrast Media Mol Imaging 2021; 2021: 6656877
- 15 Saman H, Raza SS, Uddin S. et al. Inducing angiogenesis, a key step in cancer vascularization, and treatment approaches. Cancers (Basel) 2020; 12 (05) 1172
- 16 Mao Y, Mu J, Zhao J. et al. The value of superb microvascular imaging in differentiating benign renal mass from malignant renal tumor: a retrospective study. Br J Radiol 2018;
- 17 Mu J, Mao Y, Li F. et al. Superb microvascular imaging is a rational choice for accurate Bosniak classification of renal cystic masses. Br J Radiol 2019;
- 18 Roussel E, Campi R, Amparore D. et al. Expanding the Role of Ultrasound for the Characterization of Renal Masses. J Clin Med 2022; 11 (04) 1112
- 19 Mao Y, Xia T, Wang H. et al. Evaluation of image-pro plus assisted superb microvascular imaging for differential diagnosis of renal masses. Urol Oncol 2024; 42 (09) 292.e9-292.e16
- 20 Park AY, Seo BK. Up-to-date Doppler techniques for breast tumor vascularity: superb microvascular imaging and contrast-enhanced ultrasound. Ultrasonography 2018; 37 (02) 98-106