CC BY 4.0 · Eur J Dent 2024; 18(02): 579-586
DOI: 10.1055/s-0043-1774319
Original Article

In Vitro Evaluation of Candida albicans Adhesion and Related Surface Properties of CAD/CAM Denture Base Resins

1   Department of Substitutive Dental Sciences, College of Dentistry, Imam Abdulrahman Bin Faisal University, Dammam, Saudi Arabia
,
1   Department of Substitutive Dental Sciences, College of Dentistry, Imam Abdulrahman Bin Faisal University, Dammam, Saudi Arabia
,
1   Department of Substitutive Dental Sciences, College of Dentistry, Imam Abdulrahman Bin Faisal University, Dammam, Saudi Arabia
,
1   Department of Substitutive Dental Sciences, College of Dentistry, Imam Abdulrahman Bin Faisal University, Dammam, Saudi Arabia
,
1   Department of Substitutive Dental Sciences, College of Dentistry, Imam Abdulrahman Bin Faisal University, Dammam, Saudi Arabia
,
Faraz A. Farooqi
2   Department of Dental Education, College of Dentistry, Imam Abdulrahman Bin Faisal University, Dammam, Saudi Arabia
,
Asif Matin
3   IRC Membranes and Water Security, King Fahd University of Petroleum and Minerals, Dhahran, Saudi Arabia
,
Doaa M. Al-Eraky
4   Department of Biomedical Sciences, College of Medicine, Gulf Medical University, Ajman, United Arab Emirates
,
1   Department of Substitutive Dental Sciences, College of Dentistry, Imam Abdulrahman Bin Faisal University, Dammam, Saudi Arabia
,
1   Department of Substitutive Dental Sciences, College of Dentistry, Imam Abdulrahman Bin Faisal University, Dammam, Saudi Arabia
› Author Affiliations
 

Abstract

Objective The aim of this study was to evaluate the surface roughness, contact angle, and adhesion of Candida albicans to computer-aided designing/computer-aided manufacturing (CAD/CAM) and heat-polymerized (HP) denture base materials.

Materials and Methods Specimens were allocated to six groups based on the composition of studied denture base materials, HP acrylic resin, milled resins (AvaDent and IvoCad), and 3D-printed resins (ASIGA, FormLabs, and NextDent). Ten specimens per group were used for each test (n = 10/test). Surface roughness and contact angles were analyzed using profilometer and goniometer, respectively. Adhesion of C. albicans was counted using colony-forming unit (CFU/mL). Means and standard deviations were calculated, and then one-way analysis of variance (ANOVA), followed by Tukey's post hoc test. Correlation of Candida adhesion and surface parameters was determined by using Pearson's correlation analysis.

Results No statistically significant difference was noted in surface roughness between HP, milled, and 3D-printed denture base resins except NextDent, which showed significantly higher roughness in comparison to all other resins (p = 0.001). In terms of contact angle, milled resins had the lowest value, followed by HP, ASIGA, and FormLabs, whereas NextDent showed the highest contact angle (p = 0.001). C. albicans adhesion showed no significant difference between all denture base resins. A positive and significant correlation was found between C. albicans adhesion and contact angle (p = 0.003), while no correlation was reported between C. albicans adhesion and surface roughness (p = 0.523).

Conclusion Adhesion of C. albicans was similar in all tested specimens. Surface roughness showed no significant difference between all groups except NextDent, which had the highest value. Milled denture base resins had the lowest contact angle among all groups.


#

Introduction

Polymethyl methacrylate (PMMA) resin was proven satisfactory for the fabrication of removable dentures due to its mechanical, optical, aesthetic, and biological characteristics.[1] [2] However, low surface hardness, porosities, surface roughness, and contact angles could enhance microbial adhesion.[2] Thus, epidemiological studies report a 70% prevalence of denture-associated stomatitis in patients wearing removable prosthesis, with Candida albicans being the main pathogen.[3]

Candida adhesion to acrylic surface has been correlated with the surface properties (roughness and hydrophobicity) of the material, the Candida species, and the surrounding environment.[4] [5] The surface roughness of acrylic resin is dependent on many factors including the material's structure, manufacturing process such as polymerization, and polishing procedures.[6] Roughness and grooves on the resin specimen provide more hideout places for microorganisms away from the normal cleaning process.[4]

Additionally, the microbial adhesion to a material is correlated to its surface hydrophobicity and free energy.[7] The hydrophobic Candida easily adheres to hydrophobic resin surface. Therefore, increasing the hydrophilicity and reducing contact angle could lower Candida adhesion.[4] Murat et al[8] described a significant positive association between surface roughness and Candida adhesion with no correlation between hydrophobicity and Candida adhesion. On the contrary, da Silva et al[1] stated that as the hydrophilicity of a denture base material is changed, microbial colonization is altered. In addition to the impact of surface properties on Candida adhesion, the residual monomer that is released from denture base material over time may lead to porosity formation and enhance the adhesion of Candida and biofilm formation thereafter.[9]

As surface properties play an essential role in Candida adhesion, altering these characteristics may make the dentures less prone to the adhesion.[2] These alterations may comprise surface coating of the resin, chemical composition modification, or the addition of fillers. All of these treatment modalities proved effective in reducing Candida adhesion when contact angle and surface roughness were reduced.[4] A simpler approach may involve the use of computer-aided designing/computer-aided manufacturing (CAD/CAM) PMMA as a substitute to heat polymerization. The manufacturing technique of CAD/CAM milled PMMA creates a highly cross-linked structure that is less porous with minimal residual monomer.[9] Additionally, milled dentures have a better fit, which reduces dead spaces under the denture that acts as a Candida reservoir.[8] [10]

On the other hand, dentures can be made utilizing the 3D-printing technology, where they are virtually designed using CAD software and then 3D printed using the desired resin material.[11] Nevertheless, this technology is relatively new to the removable prosthesis field and has not been extensively investigated. Few studies evaluated the surface characteristics of various kinds of denture base resins and reported conflicting results. Di Fiore et al[12] reported insignificant differences between the surface roughness of heat-polymerized (HP), milled, and 3D-printed denture base resins after regular polishing procedures, while Gad et al[13] reported lower surface roughness of 3D-printed PMMA compared with HP resin. Other studies reported significant differences in roughness and contact angle between various types of 3D-printed resins and in comparison to milled or HP resins.[14] [15]

The surface roughness and wettability might vary according to the brands of CAD/CAM PMMA used.[16] Accordingly, the present study assessed the surface roughness, contact angle, and C. albicans adhesion among different brands of CAD/CAM denture base resins manufactured by different CAM technologies (milling and 3D printing) in relation to conventional HP denture base resins. The study's null hypothesis stated that there will be no difference in surface roughness, contact angles, and C. albicans adhesion between CAD/CAM and conventional HP denture base resins.


#

Materials and Methods

Sample Size Calculation and Test Groups

The sample size for this study was determined through power analysis. For this purpose, the formula was adopted from the World Health Organization (WHO), keeping 0.05 as the level of significance, power at 80%, and marginal error set at 5%, which demonstrated the need for 10 specimens for each group to estimate the presumed effect size. The total number of required specimens was 180 divided as follow; 60 specimens per tested property with 10 specimens of each material. Rectangular acrylic specimens with dimension of 10 × 12 × 2.5 mm were prepared from different resins: HP acrylic resin, prepolymerized acrylic disks for milling, AvaDent and IvoCad and 3D-printed resins, ASIGA, FormLabs, and Denture 3D+ (see [Table 1] for details).

Table 1

Materials used in the present study

Group

Material/equipment

Heat-polymerized acrylic resin

Major.Base.20, Major Prodotti Dentari, Moncalieri, Italy

AvaDent

AvaDent denture base puck (AvaDent, Global Dental Science Europe, Tilburg, the Netherlands)

IvoCad

IvoBase CAD (Ivoclar Vivadent, Schaan, Liechtenstein)

ASIGA

(Resin) ASIGA DentaBASE (Asiga pty Ltd, Alexandria, NSW, Australia)

(Printer) ASIGA MAX Printer (DLP)

FormLabs

(Resin) FormLabs Denture Base LP (FormLabs, Somerville, MA, United States)

(Printer) Form 2 Printer (SLA)

NextDent

(Resin) Denture 3D+ (NextDent B.V., Soesterberg, the Netherlands)

(Printer) NextDent 5100 3D Printer (SLA)


#

Fabrication of HP, Milled, and 3D-Printed Specimens

HP (control) acrylic resin specimens were fabricated by the use of conventional water bath method as mentioned in an earlier study.[17] Investing of wax specimens in dental stone was done, followed by wax elimination. Packing of acrylic resin mix was done at the dough phase following the application of separating medium on stone surfaces. After that, processing of acrylic resin was achieved in water bath polymerization unit (KaVo Elektrotechnisches Werk GmbH, Leutkirch, Germany) at 73°C for 90 minutes, then at 100°C for an additional 30 minutes. Finishing of specimens was accomplished by the use of tungsten carbide bur (HM 79GX-040 HP; Meisinger, Centennial, CO, United States) to remove excess resin.

For both milled groups (AvaDent and IvoCad), prepolymerized PMMA pucks were cut to the required dimensions, where each disk was positioned and fixed in precision cutting machine (IsoMet 5000 Linear Precision Saw, Buehler, Lake Bluff, IL, United States) and sliced using diamond disk under constant water coolant.[18]

For the printed specimens, the stereolithography (STL) file of the design was created using an open software (123D design, Autodesk, version 2.2.14, San Francisco, California, United States) and sent to each material's corresponding printer ([Table 1]). For ASIGA and NextDent resins, the resin containers were shaken for 30 minutes and then poured into the resin tank, while for FormLabs, the resin tank was mounted on the printer directly. The printing orientation of all specimens was set at 90 degrees to the platform and 50-µm layer thickness. Following the printing procedure, the specimens were immersed in 99.9% isopropyl alcohol (IPA) to remove uncured resin. To complete the polymerization of printed specimens, additional postcuring cycle was done according to manufacturer's recommendations. Specimens were placed in the glycerin path within the postcuring machine.

All specimens (HP, milled, and 3D printed) were polished using 1,200-grit sandpaper disks (MicroCut PSA; Buehler) by the use of a polishing machine (MetaServ 250 Grinder Polisher, Buehler) in wet settings to ensure standardized polishing methods. A single investigator performed the polishing procedure of all the specimens and reassessed the specimens' dimensions to 0.01-µm accuracy using a digital caliber. Specimens with acceptable dimensions were incubated in distilled water at 37°C for 48 hours before assessing the desired properties.


#

Measurement of Surface Roughness and Contact Angle

A noncontact optical interferometric profilometer (Contour GT; Bruker Nano GmbH, Berlin, Germany) was utilized in measuring the surface roughness (Ra ) of each specimen at five distant areas and 0.01-mm resolution. The average Ra value per specimen was then calculated.[19]

An automated goniometer (DM-501; Kyowa Interface Science Co., Niiza-City, Saitama, Japan) measured the contact angle (degrees) at four areas on each specimen followed by mean value calculation per specimen. The sessile drop technique was followed using an autopipette to dispense (2-μL) droplets of distilled water on the specimen's dry surface. The images were interpreted using the FAMAS software (Kyowa Interface Science Co.).[20]


#

Microbiological Analysis of the Biofilm

Frozen culture of C. albicans reference strain (ATCC 10231) was inoculated onto Sabouraud dextrose agar (SDA; MOLEQULE-ON, New Lynn, Auckland, New Zealand) for 48 hours at 37°C. Isolated colonies were added to Sabouraud dextrose broth (SDB; MOLEQULE-ON) for overnight incubation at 37°C and then the broth was diluted and adjusted to approximately 00.5 McFarland (1 × 107 CFU/mL; DensiCHEK TM Plus, Durham, NC, United States).

The biofilm formation was evaluated using the protocol of colony-forming unit (CFU) assay according to Gulati et al[21] with a slight modification. Briefly, after sterilization of specimens with 70% IPA, each specimen was placed in 12-well cell culture plates (Nunclon Delta Surface, Thermo Fisher Scientific, Roskilde, Denmark), and a volume of 1,000 µL of the adjusted yeast suspension was added to each well and incubated at 37°C for 48 hours. To remove nonadherent cells, the specimens were washed three times with phosphate buffer saline (PBS), then scraped and vortex for 2 minutes at 3,000 rpm to dislodge the adherent cells from the specimens.[22] [23] [24] To enumerate CFU count, 10-fold dilution in PBS was performed, before a volume of 100 µL was directly platted onto SDA plates and incubated at 30°C for 48 hours. The experiment was performed blindly in triplicates with positive and negative controls to ensure reproducibility.[25] [26] [27]


#

Statistical Analysis

The normality of the data was evaluated using the Shapiro–Wilk test and p-values greater than 0.05 indicated that the data were normally distributed. Comparison of means between the groups (HP, AvaDent, IvoCad, ASIGA, FormLabs, and NextDent) for each tested property was done using one-way analysis of variance (ANOVA). Significant ANOVA results necessitated the use of Tukey's post hoc test to identify the pairwise differences. Pearson's correlation analysis was used to correlate C. albicans adhesion and related surface parameters. Statistical package for social science (SPSS, IBM, New York, United States) version 24 was used for statistical analysis and p-values ≤0.05 were considered statistically significant.


#
#

Results

[Table 2] presents the means, standard deviations, and significance of surface roughness (Ra , µm) between tested materials. ANOVA results for Ra exhibited a significant difference between the groups (p = 0.001). For pairwise comparisons, no significant differences in surface roughness were found between any of the groups except with NextDent, which showed the highest Ra value (1.68 ± 0.22 µm) among the groups (p < 0.05), whereas ASIGA showed the lowest Ra value (0.92 ± 0.23 µm).

Table 2

Mean, SD, and significances between groups for all tested properties

HP

AvaDent

IvoCad

ASIGA

FormLabs

NextDent

p-value

Mean ± SD

Mean ± SD

Mean ± SD

Mean ± SD

Mean ± SD

Mean ± SD

Ra (µm)

1.09 ± 0.16[a]

1.28 ± 0.41[a]

1.15 ± 0.12[a]

0.92 ± 0.23[a]

1.23 ± 0.33[a]

1.68 ± 0.22

F = 9.931

p = 0.001[b]

Contact angle (degrees)

79.44 ± 3.84[a]

70.01 ± 2.61[c]

72.4 ± 3.74[c]

81.63 ± 3.13[a]

80.62 ± 8.35[a]

89.91 ± 3.61

F = 23.709

p = 0.001[b]

Candida count (log10 CFU/mL)[d]

4.22 ± 0.166

4.14 ± 0.089

4.11 ± 0.101

4.21 ± 0.118

4.21 ± 0.066

4.27 ± 0.203

F = 1.897

p = 0.110

Abbreviations: CFU, colony-forming unit; HP, heat-polymerized; SD, standard deviation.


a No statistical difference between the groups.


b Statistically significant at p = 0.05.


c No statistical difference between the groups.


d ANOVA (analysis of variance) results were not statistically significant; therefore, post hoc was not performed.


For contact angle, the values are summarized in [Table 2], and representative images of contact angles are shown in [Fig. 1]. The ANOVA results showed significant differences between the materials (p = 0.001). For intergroup comparisons, NextDent significantly showed the highest contact angle (89.91 ± 3.61 degrees; p = 0.001) among the groups. Compared with the control material (HP), the milled groups (AvaDent/IvoCad) showed significantly lower contact angles (p = 0.001 and 0.015, respectively), while ASIGA and FormLabs showed no significant differences (p = 0.895 and 0.992, respectively).

Zoom Image
Fig. 1 Representative images of contact angle of all tested resins. (A) heat-polymerized (HP), (B) AvaDent, (C) IvoCad, (D) ASIGA, (E) FormLabs, and (F) NextDent.

The C. albicans colony counts per material are presented in [Table 2] and [Fig. 2]. The overall results showed that there was no significant difference in C. albicans count between all tested materials (p = 0.074). The highest CFU (log10 CFU/mL) count of C. albicans adhered to the NextDent-printed specimens (4.27 ± 0.203), while the lowest count was recorded with IvoCad (4.11 ± 0.101).

Zoom Image
Fig. 2 The colony forming unit assay of Candida albicans biofilm recovered from each tested group: (A) heat-polymerized (HP), (B) AvaDent, (C) IvoCad, (D) ASIGA, (E) FormLabs, and (F) NextDent.

[Table 3] displays the level of association between C. albicans count and tested surface parameters that were analyzed using Pearson's correlation. The analysis showed that C. albicans count was not associated with surface roughness, but was significantly associated with contact angle.

Table 3

Pearson's correlation among Candida adhesion and surface parameters

Candida count

Ra

Contact angle

Candida count

1

r = 0.492

p = 0.149

r = 0.825

p = 0.003[a]

Ra

1

r = 0.230

p = 0.523

Contact angle

1

a Significant at p = 0.05.



#

Discussion

The present study tested the surface roughness, contact angle, and C. albicans adhesion of HP denture base acrylic resin compared with the CAD/CAM counterparts fabricated by milling or 3D-printing technologies. The study hypothesis was partly rejected since the contact angle and surface roughness showed variation between the tested denture base resins, while C. albicans adhesion showed no significant difference.

The results demonstrated no significant differences between the surface roughness of HP, milled, and 3D-printed resins, except NextDent specimens, which showed the highest value among all tested groups. Similarly, Di Fiore et al[12] found that HP and CAD/CAM denture base resins (milled and 3D printed) had similar surface roughness after polishing, while the milled resin showed lower surface roughness before polishing. Also, Al-Dwairi et al[15] mentioned that 3D-printed resins showed various surface properties between different brands where ASIGA exhibited the lowest surface roughness among the studied 3D-printed resins, which is in agreement with our results. In contrast, previous studies reported lower surface roughness of milled and 3D-printed resins, in comparison to HP.[4] [8] [16] [18] Also, Helal et al[14] and Alharethi[28] found lower surface roughness of milled resin compared with printed resins. Variation in surface roughness results might be related to differences between the tested materials, tested manufacturing techniques, or the polishing methods used in these studies. Also, some studies did not compare between milled and 3D-printed resins,[15] [16] or included only one type (brand) of denture base resin for each fabrication technique (milling or 3D printing).[12] Murat et al[8] subjected the specimens to thermal cycling before testing the surface properties, which could be the reason for the variation in results when compared with our findings. In another study by Srinivasan et al,[29] milled (AvaDent) and printed (NextDent) resins exhibited comparable surface roughness, which is in disagreement with the current study. This might be due to the difference in polishing techniques, printing orientation, and layer thickness.

The type of printer, printing technologies, and printing resins were the variables available among the three groups of 3D-printed specimens tested in the current study. The printing technologies of the three printers used were as follows: the NextDent and FormLabs printers were based on SLA, and the ASIGA printer was based on digital light projection (DLP) technology. NextDent specimens exhibited higher surface roughness than those of FormLabs and ASIGA. Since the printing parameters (layer thickness and printing orientation) were similar in all 3D-printed groups, and the printing technology of NextDent and FormLabs was the same (SLA), therefore the variation in surface roughness might be related to other parameters and limitations within each printer or printing material.

The contact angles of milled resins were significantly lower than those of HP and 3D-printed resins in the present study. However, Al-Dwairi et al[15] [16] found that HP had a lower contact angle than milled and 3D-printed resins. Differences in results could be related to the variations in the study designs and tested materials. In addition, they compared the properties of milled and 3D-printed resins to HP resins in two separate studies.[15] [16] Comparison of contact angle results between our study and previously published research was difficult due to the low number of studies testing contact angles of CAD/CAM materials.

C. albicans adhesion in the present study was not significantly affected by the manufacturing technique, as there was no significant difference in Candida adherence to the surfaces of the tested materials. However, the lowest count was found with milled CAD/CAM resins followed by printed resin, except NextDent specimens, which showed higher Candida count than HP PMMA. A previous study[12] compared C. albicans adhesion between different CAD/CAM resins (milled and 3D printed) and HP PMMA after 16 hours of incubation and reported similar results. They demonstrated that the time of incubation affected Candida adhesion. Increasing the incubation time resulted in microbial biofilm formation and increased Candida adhesion independent of the surface roughness.[12] In the present study, the incubation time was 48 hours. This could explain the results of no difference between the tested groups regarding C. albicans adhesion. Previous studies reported lower Candida adhesion on milled resin compared with HP resin after short incubation periods (90 minutes and 2 hours) than the one followed in our study.[8] [18] Also, the tested materials were different than those tested in the present study. A recent report noted that the printing technology (SLA and DLP) does not influence C. albicans adhesion to 3D-printed denture bases, supporting the results in the present study.[30] Shim et al[31] investigated the effect of printing orientation on Candida adhesion and found a significantly lower Candida count when a 90-degree orientation was used in comparison to 0 and 45 degrees. Accordingly, the vertical printing orientation followed in our study might be the reason for no differences of C. albicans adhesion among 3D-printed, HP, and milled resins, even in NextDent specimens, which showed a significantly higher surface roughness.

The results of the present study showed that contact angle was significantly correlated with C. albicans adhesion, while surface roughness showed no significant correlation. Milled resins tested in this study had the lowest values of contact angle compared with HP and 3D-printed resins. In addition, they also had the least number of C. albicans count but without significant difference compared with printed and HP resins. The correlation between surface hydrophobicity and Candida adhesion has been confirmed previously.[4] It was demonstrated that a decrease in contact angle was associated with less Candida adhesion.[4] Murat et al[8] reported similar findings of lower contact angle and C. albicans adhesion with milled denture base resin than HP. However, contrary to our results, they reported lower surface roughness of milled resin than HP, and significant correlation between the increase in surface roughness and the number of attached C. albicans cells, while hydrophobicity (high contact angle) showed no correlation. Variation in results might be related to differences in the tested materials, polishing techniques, and then exposing the specimens to thermal cycling. Moreover, printed resins were not evaluated in their study.

The findings of the current study presented no difference regarding C. albicans adhesion to CAD/CAM and HP denture base resins. However, the lowest count was found with milled resins. High contact angle was significantly correlated with higher C. albicans adhesion, while surface roughness showed no significant correlation. In between printed resins and compared with milled and HP resins, NextDent showed the highest surface roughness and contact angle. However, other tested 3D-printed resins showed similar properties to HP. Looking at the results of this study, the tested 3D-printed resins could be used clinically for the construction of complete dentures with similar possibility of C. albicans adhesion as HP and milled resins. Thus, incidence of denture stomatitis is expected to be similar among denture base materials manufactured by conventional heat polymerization, milling, or 3D-printing technologies. Based on the present results, the tested denture base materials when used clinically for construction of denture bases would show comparable adhesion of C. albicans irrespective to their method of fabrication (conventional, milled, and 3D printed). However, the present findings should be interpreted with caution due to its in vitro nature. Exposure to artificial saliva with various pH affects the mechanical and surface roughness of conventional and CAD/CAM denture base resins.[32] Therefore, clinical studies are required to support the present findings, after exposing the tested materials to masticatory forces, oral flora, saliva, denture cleansing, and food and beverages with varying temperatures.

The present study included more than one brand from each CAD/CAM manufacturing techniques, in addition to the use of two printing technologies, SLA and DLP, which would add validity to the results reported by each manufacturing technique. The biofilm assay by determination of CFU is a common microbiological research technique; however, this process is laborious, tedious, and time-consuming.[23] Future research should focus on studying the biofilm of C. albicans using other methods such as Cell Proliferation Assay Kits. This study is limited as the effect of aging or denture cleaning routines on the tested properties was not tested. Further investigations are required to test the effect of aging and beverage consumption on different characteristics of CAD/CAM resins. The differences between 3D-printed resins tested in the present study require further investigations to understand the factors causing these variations.


#

Conclusion

The adhesion of C. albicans to the surfaces of milled, 3D-printed, and HP denture base resins was similar; however, the lowest count was found in milled resins. Surface roughness of milled and 3D-printed resins was the same as that of HP, except NextDent, which showed the highest value. Milled resins had significantly lower contact angles compared with other tested groups.


#
#

Conflict of Interest

None declared.

Acknowledgment

The authors would like to thank Mr. Thanga Raju for his assistance in preparing the specimens and Robert Caravana for assisting in the surface roughness test.

  • References

  • 1 da Silva WJ, Leal CMB, Viu FC, Gonçalves LM, Barbosa CMR, Del Bel Cury AA. Influence of surface free energy of denture base and liner materials on Candida albicans biofilms. J Investig Clin Dent 2015; 6 (02) 141-146
  • 2 Queiroz JRC, Fissmer SF, Koga-Ito CY. et al. Effect of diamond-like carbon thin film coated acrylic resin on Candida albicans biofilm formation. J Prosthodont 2013; 22 (06) 451-455
  • 3 Koch C, Bürgers R, Hahnel S. Candida albicans adherence and proliferation on the surface of denture base materials. Gerodontology 2013; 30 (04) 309-313
  • 4 Gad MM, Abualsaud R, Khan SQ. Hydrophobicity of denture base resins: a systematic review and meta-analysis. J Int Soc Prev Community Dent 2022; 12 (02) 139-159
  • 5 Mayahara M, Kataoka R, Arimoto T. et al. Effects of surface roughness and dimorphism on the adhesion of Candida albicans to the surface of resins: scanning electron microscope analyses of mode and number of adhesions. J Investig Clin Dent 2014; 5 (04) 307-312
  • 6 Al-Harbi FA, Abdel-Halim MS, Gad MM. et al. Effect of nanodiamond addition on flexural strength, impact strength, and surface roughness of PMMA denture base. J Prosthodont 2019; 28 (01) e417-e425
  • 7 Schubert A, Wassmann T, Holtappels M, Kurbad O, Krohn S, Bürgers R. Predictability of microbial adhesion to dental materials by roughness parameters. Coatings 2019; 9 (07) 456-467
  • 8 Murat S, Alp G, Alatalı C, Uzun M. In vitro evaluation of adhesion of Candida albicans on CAD/CAM PMMA-based polymers. J Prosthodont 2019; 28 (02) e873-e879
  • 9 de Oliveira Limírio JPJ, Gomes JML, Alves Rezende MCR, Lemos CAA, Rosa CDDRD, Pellizzer EP. Mechanical properties of polymethyl methacrylate as a denture base: conventional versus CAD-CAM resin—a systematic review and meta-analysis of in vitro studies. J Prosthet Dent 2022; 128 (06) 1221-1229
  • 10 Charoenphol K, Peampring C. Fit accuracy of complete denture base fabricated by CAD/CAM milling and 3D-printing methods. Eur J Dent 2023; Jul; 17 (03) 889-894
  • 11 Gad MM, Alalawi H, Akhtar S. et al Strength and wear behavior of three-dimensional printed and prefabricated denture teeth: an in vitro comparative analysis. Eur J Dent 2023; 17 (04) 1248-1256
  • 12 Di Fiore A, Meneghello R, Brun P. et al. Comparison of the flexural and surface properties of milled, 3D-printed, and heat polymerized PMMA resins for denture bases: an in vitro study. J Prosthodont Res 2022; 66 (03) 502-508
  • 13 Gad MM, Fouda SM, Abualsaud R. et al. Strength and surface properties of a 3D-printed denture base polymer. J Prosthodont 2022; 31 (05) 412-418
  • 14 Helal MA, Fadl-Alah A, Baraka YM, Gad MM, Emam AM. In-vitro comparative evaluation for the surface properties and impact strength of CAD/CAM milled, 3D printed, and polyamide denture base resins. J Int Soc Prev Community Dent 2022; 12 (01) 126-131
  • 15 Al-Dwairi ZN, Al Haj Ebrahim AA, Baba NZ. A comparison of the surface and mechanical properties of 3D printable denture-base resin material and conventional polymethylmethacrylate (PMMA). J Prosthodont 2023; 32 (01) 40-48
  • 16 Al-Dwairi ZN, Tahboub KY, Baba NZ, Goodacre CJ, Özcan M. A comparison of the surface properties of CAD/CAM and conventional polymethylmethacrylate (PMMA). J Prosthodont 2019; 28 (04) 452-457
  • 17 Gad MM, Fouda SM, ArRejaie AS, Al-Thobity AM. Comparative effect of different polymerization techniques on the flexural and surface properties of acrylic denture bases. J Prosthodont 2019; 28 (04) 458-465
  • 18 Al-Fouzan AF, Al-Mejrad LA, Albarrag AM. Adherence of Candida to complete denture surfaces in vitro: a comparison of conventional and CAD/CAM complete dentures. J Adv Prosthodont 2017; 9 (05) 402-408
  • 19 Alqanas SS, Alfuhaid RA, Alghamdi SF, Al-Qarni FD, Gad MM. Effect of denture cleansers on the surface properties and color stability of 3D printed denture base materials. J Dent 2022; 120: 104089
  • 20 Alzayyat ST, Almutiri GA, Aljandan JK. et al. Antifungal efficacy and physical properties of poly(methylmethacrylate) denture base material reinforced with SiO2 nanoparticles. J Prosthodont 2021; 30 (06) 500-508
  • 21 Gulati M, Lohse MB, Ennis CL. et al. In vitro culturing and screening of candida albicans biofilms. Curr Protoc Microbiol 2018; 50 (01) e60
  • 22 Mota S, Alves R, Carneiro C. et al. Candida glabrata susceptibility to antifungals and phagocytosis is modulated by acetate. Front Microbiol 2015; 6: 919
  • 23 AlEraky DM, Abuohashish HM, Gad MM. et al. The antifungal and antibiofilm activities of caffeine against Candida albicans on polymethyl methacrylate denture base material. Biomedicines 2022; 10 (09) 2078
  • 24 Fouda SM, Gad MM, Ellakany P. et al. Effect of low nanodiamond concentrations and polymerization techniques on physical properties and antifungal activities of denture base resin. Polymers (Basel) 2021; 13 (24) 4331
  • 25 Brugger SD, Baumberger C, Jost M, Jenni W, Brugger U, Mühlemann K. Automated counting of bacterial colony forming units on agar plates. PLoS One 2012; 7 (03) e33695
  • 26 Kim DJ, Jung MY, Park JH. et al. Moxifloxacin releasing intraocular implant based on a cross-linked hyaluronic acid membrane. Sci Rep 2021; 11 (01) 24115
  • 27 Sanchez M, Xia Z, Rico-Bautista E. et al. Oxidized analogs of Di(1H-indol-3-yl)methyl-4-substituted benzenes are NR4A1-dependent UPR inducers with potent and safe anti-cancer activity. Oncotarget 2018; 9 (38) 25057-25074
  • 28 Alharethi NA. Evaluation of the influence of build orientation on the surface roughness and flexural strength of 3D-printed denture base resin and its comparison with CAD-CAM milled denture base resin. Eur J Dent 2024; 18 (01) 321-328
  • 29 Srinivasan M, Kalberer N, Kamnoedboon P. et al. CAD-CAM complete denture resins: an evaluation of biocompatibility, mechanical properties, and surface characteristics. J Dent 2021; 114: 103785
  • 30 Li P, Fernandez PK, Spintzyk S, Schmidt F, Beuer F, Unkovskiy A. Effect of additive manufacturing method and build angle on surface characteristics and Candida albicans adhesion to 3D printed denture base polymers. J Dent 2022; 116: 103889
  • 31 Shim JS, Kim JE, Jeong SH, Choi YJ, Ryu JJ. Printing accuracy, mechanical properties, surface characteristics, and microbial adhesion of 3D-printed resins with various printing orientations. J Prosthet Dent 2020; 124 (04) 468-475
  • 32 Alzaid M, AlToraibily F, Al-Qarni FD. et al. The effect of salivary pH on the flexural strength and surface properties of CAD/CAM denture base materials. Eur J Dent 2023; 17 (01) 234-241

Address for correspondence

Shaimaa Mohamed Fouda, BDS, MSc, PhD
College of Dentistry, Imam Abdulrahman Bin Faisal University
P.O. Box 1982, Dammam 31441
Saudi Arabia   

Publication History

Article published online:
12 December 2023

© 2023. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution License, permitting unrestricted use, distribution, and reproduction so long as the original work is properly cited. (https://creativecommons.org/licenses/by/4.0/)

Thieme Medical and Scientific Publishers Pvt. Ltd.
A-12, 2nd Floor, Sector 2, Noida-201301 UP, India

  • References

  • 1 da Silva WJ, Leal CMB, Viu FC, Gonçalves LM, Barbosa CMR, Del Bel Cury AA. Influence of surface free energy of denture base and liner materials on Candida albicans biofilms. J Investig Clin Dent 2015; 6 (02) 141-146
  • 2 Queiroz JRC, Fissmer SF, Koga-Ito CY. et al. Effect of diamond-like carbon thin film coated acrylic resin on Candida albicans biofilm formation. J Prosthodont 2013; 22 (06) 451-455
  • 3 Koch C, Bürgers R, Hahnel S. Candida albicans adherence and proliferation on the surface of denture base materials. Gerodontology 2013; 30 (04) 309-313
  • 4 Gad MM, Abualsaud R, Khan SQ. Hydrophobicity of denture base resins: a systematic review and meta-analysis. J Int Soc Prev Community Dent 2022; 12 (02) 139-159
  • 5 Mayahara M, Kataoka R, Arimoto T. et al. Effects of surface roughness and dimorphism on the adhesion of Candida albicans to the surface of resins: scanning electron microscope analyses of mode and number of adhesions. J Investig Clin Dent 2014; 5 (04) 307-312
  • 6 Al-Harbi FA, Abdel-Halim MS, Gad MM. et al. Effect of nanodiamond addition on flexural strength, impact strength, and surface roughness of PMMA denture base. J Prosthodont 2019; 28 (01) e417-e425
  • 7 Schubert A, Wassmann T, Holtappels M, Kurbad O, Krohn S, Bürgers R. Predictability of microbial adhesion to dental materials by roughness parameters. Coatings 2019; 9 (07) 456-467
  • 8 Murat S, Alp G, Alatalı C, Uzun M. In vitro evaluation of adhesion of Candida albicans on CAD/CAM PMMA-based polymers. J Prosthodont 2019; 28 (02) e873-e879
  • 9 de Oliveira Limírio JPJ, Gomes JML, Alves Rezende MCR, Lemos CAA, Rosa CDDRD, Pellizzer EP. Mechanical properties of polymethyl methacrylate as a denture base: conventional versus CAD-CAM resin—a systematic review and meta-analysis of in vitro studies. J Prosthet Dent 2022; 128 (06) 1221-1229
  • 10 Charoenphol K, Peampring C. Fit accuracy of complete denture base fabricated by CAD/CAM milling and 3D-printing methods. Eur J Dent 2023; Jul; 17 (03) 889-894
  • 11 Gad MM, Alalawi H, Akhtar S. et al Strength and wear behavior of three-dimensional printed and prefabricated denture teeth: an in vitro comparative analysis. Eur J Dent 2023; 17 (04) 1248-1256
  • 12 Di Fiore A, Meneghello R, Brun P. et al. Comparison of the flexural and surface properties of milled, 3D-printed, and heat polymerized PMMA resins for denture bases: an in vitro study. J Prosthodont Res 2022; 66 (03) 502-508
  • 13 Gad MM, Fouda SM, Abualsaud R. et al. Strength and surface properties of a 3D-printed denture base polymer. J Prosthodont 2022; 31 (05) 412-418
  • 14 Helal MA, Fadl-Alah A, Baraka YM, Gad MM, Emam AM. In-vitro comparative evaluation for the surface properties and impact strength of CAD/CAM milled, 3D printed, and polyamide denture base resins. J Int Soc Prev Community Dent 2022; 12 (01) 126-131
  • 15 Al-Dwairi ZN, Al Haj Ebrahim AA, Baba NZ. A comparison of the surface and mechanical properties of 3D printable denture-base resin material and conventional polymethylmethacrylate (PMMA). J Prosthodont 2023; 32 (01) 40-48
  • 16 Al-Dwairi ZN, Tahboub KY, Baba NZ, Goodacre CJ, Özcan M. A comparison of the surface properties of CAD/CAM and conventional polymethylmethacrylate (PMMA). J Prosthodont 2019; 28 (04) 452-457
  • 17 Gad MM, Fouda SM, ArRejaie AS, Al-Thobity AM. Comparative effect of different polymerization techniques on the flexural and surface properties of acrylic denture bases. J Prosthodont 2019; 28 (04) 458-465
  • 18 Al-Fouzan AF, Al-Mejrad LA, Albarrag AM. Adherence of Candida to complete denture surfaces in vitro: a comparison of conventional and CAD/CAM complete dentures. J Adv Prosthodont 2017; 9 (05) 402-408
  • 19 Alqanas SS, Alfuhaid RA, Alghamdi SF, Al-Qarni FD, Gad MM. Effect of denture cleansers on the surface properties and color stability of 3D printed denture base materials. J Dent 2022; 120: 104089
  • 20 Alzayyat ST, Almutiri GA, Aljandan JK. et al. Antifungal efficacy and physical properties of poly(methylmethacrylate) denture base material reinforced with SiO2 nanoparticles. J Prosthodont 2021; 30 (06) 500-508
  • 21 Gulati M, Lohse MB, Ennis CL. et al. In vitro culturing and screening of candida albicans biofilms. Curr Protoc Microbiol 2018; 50 (01) e60
  • 22 Mota S, Alves R, Carneiro C. et al. Candida glabrata susceptibility to antifungals and phagocytosis is modulated by acetate. Front Microbiol 2015; 6: 919
  • 23 AlEraky DM, Abuohashish HM, Gad MM. et al. The antifungal and antibiofilm activities of caffeine against Candida albicans on polymethyl methacrylate denture base material. Biomedicines 2022; 10 (09) 2078
  • 24 Fouda SM, Gad MM, Ellakany P. et al. Effect of low nanodiamond concentrations and polymerization techniques on physical properties and antifungal activities of denture base resin. Polymers (Basel) 2021; 13 (24) 4331
  • 25 Brugger SD, Baumberger C, Jost M, Jenni W, Brugger U, Mühlemann K. Automated counting of bacterial colony forming units on agar plates. PLoS One 2012; 7 (03) e33695
  • 26 Kim DJ, Jung MY, Park JH. et al. Moxifloxacin releasing intraocular implant based on a cross-linked hyaluronic acid membrane. Sci Rep 2021; 11 (01) 24115
  • 27 Sanchez M, Xia Z, Rico-Bautista E. et al. Oxidized analogs of Di(1H-indol-3-yl)methyl-4-substituted benzenes are NR4A1-dependent UPR inducers with potent and safe anti-cancer activity. Oncotarget 2018; 9 (38) 25057-25074
  • 28 Alharethi NA. Evaluation of the influence of build orientation on the surface roughness and flexural strength of 3D-printed denture base resin and its comparison with CAD-CAM milled denture base resin. Eur J Dent 2024; 18 (01) 321-328
  • 29 Srinivasan M, Kalberer N, Kamnoedboon P. et al. CAD-CAM complete denture resins: an evaluation of biocompatibility, mechanical properties, and surface characteristics. J Dent 2021; 114: 103785
  • 30 Li P, Fernandez PK, Spintzyk S, Schmidt F, Beuer F, Unkovskiy A. Effect of additive manufacturing method and build angle on surface characteristics and Candida albicans adhesion to 3D printed denture base polymers. J Dent 2022; 116: 103889
  • 31 Shim JS, Kim JE, Jeong SH, Choi YJ, Ryu JJ. Printing accuracy, mechanical properties, surface characteristics, and microbial adhesion of 3D-printed resins with various printing orientations. J Prosthet Dent 2020; 124 (04) 468-475
  • 32 Alzaid M, AlToraibily F, Al-Qarni FD. et al. The effect of salivary pH on the flexural strength and surface properties of CAD/CAM denture base materials. Eur J Dent 2023; 17 (01) 234-241

Zoom Image
Fig. 1 Representative images of contact angle of all tested resins. (A) heat-polymerized (HP), (B) AvaDent, (C) IvoCad, (D) ASIGA, (E) FormLabs, and (F) NextDent.
Zoom Image
Fig. 2 The colony forming unit assay of Candida albicans biofilm recovered from each tested group: (A) heat-polymerized (HP), (B) AvaDent, (C) IvoCad, (D) ASIGA, (E) FormLabs, and (F) NextDent.