Accuracy and Feasibility Analysis of SEEG Electrode Implantation using the VarioGuide Frameless Navigation System in Patients with Drug-Resistant Epilepsy
Shiwei Song
1
Department of Neurosurgery, Fujian Medical University Union Hospital, Fuzhou, Fujian, China
,
Yihai Dai
1
Department of Neurosurgery, Fujian Medical University Union Hospital, Fuzhou, Fujian, China
,
Zhen Chen
1
Department of Neurosurgery, Fujian Medical University Union Hospital, Fuzhou, Fujian, China
,
Songsheng Shi
1
Department of Neurosurgery, Fujian Medical University Union Hospital, Fuzhou, Fujian, China
› InstitutsangabenFunding This study received funds from the Joint Funds for the Innovation of Science and Technology, Fujian province (2018Y9059).
Objective The objective of the study was to evaluate the feasibility and accuracy of frameless stereoelectroencephalography (SEEG) electrode implantation in patients with drug-resistant epilepsy using the VarioGuide system.
Methods The VarioGuide frameless navigation system was used to implant SEEG electrodes in patients with medically drug-resistant epilepsy. Demographic data, surgery duration, number of electrodes, and complications were retrospectively analyzed. Accuracy was compared by measuring the distance between the planned and actual electrode positions as determined by postoperative computed tomography images.
Results A total of 141 SEEG electrodes were implanted in 19 patients from May 2015 to December 2018 with an average of 7.42 (range: 4–10) leads per patient. The average entry point localization error (EPLE) was 1.96 ± 0.47 mm (range: 0.32–3.29) and average target point localization error (TPLE) was 2.47 ± 0.79 mm (range: 0.72–4.83). The average operating time per lead (OTPL) was 14.16 ± 2.68 minutes (range: 8.64–21.58). No complications occurred.
Conclusion The VarioGuide frameless navigation system can be an effective method for SEEG electrode implantation in patients with drug-resistant epilepsy, particularly when the electrodes are concentrated in a relatively small region and the number of implanted electrodes is small.
Georg Thieme Verlag KG Rüdigerstraße 14, 70469 Stuttgart, Germany
References
1
Katz JS,
Abel TJ.
Stereoelectroencephalography versus subdural electrodes for localization of the epileptogenic zone: what is the evidence?. Neurotherapeutics 2019; 16 (01) 59-66
4
Cardinale F.
Talairach methodology in the era of 3D multimodal imaging: “The song remains the same,” but catchier, and therefore more helpful for clinical decision making and surgical planning in epilepsy surgery. Epilepsia 2015; 56 (06) 976-977
6
Budke M,
Avecillas-Chasin JM,
Villarejo F.
Implantation of depth electrodes in children using VarioGuide® frameless navigation system: technical note. Oper Neurosurg (Hagerstown) 2018; 15 (03) 302-309
8
van der Loo LE,
Schijns OEMG,
Hoogland G.
et al.
Methodology, outcome, safety and in vivo accuracy in traditional frame-based stereoelectroencephalography. Acta Neurochir (Wien) 2017; 159 (09) 1733-1746
9
Yu H,
Pistol C,
Franklin R,
Barborica A.
Clinical accuracy of customized stereotactic fixtures for stereoelectroencephalography. World Neurosurg 2018; 109: 82-88
10
Dewan MC,
Shults R,
Hale AT.
et al.
Stereotactic EEG via multiple single-path omnidirectional trajectories within a single platform: institutional experience with a novel technique. J Neurosurg 2018; 129 (05) 1173-1181
11
Ollivier I,
Behr C,
Cebula H.
et al.
Efficacy and safety in frameless robot-assisted stereo-electroencephalography (SEEG) for drug-resistant epilepsy. Neurochirurgie 2017; 63 (04) 286-290
13
Spyrantis A,
Cattani A,
Strzelczyk A,
Rosenow F,
Seifert V,
Freiman TM.
Robot-guided stereoelectroencephalography without a computed tomography scan for referencing: analysis of accuracy, Int J Med Robot. 2018 14. 02
16
Lefranc M,
Capel C,
Pruvot AS.
et al.
The impact of the reference imaging modality, registration method and intraoperative flat-panel computed tomography on the accuracy of the ROSA® stereotactic robot. Stereotact Funct Neurosurg 2014; 92 (04) 242-250