J Neurol Surg A Cent Eur Neurosurg 2021; 82(05): 430-436
DOI: 10.1055/s-0040-1721002
Original Article

Accuracy and Feasibility Analysis of SEEG Electrode Implantation using the VarioGuide Frameless Navigation System in Patients with Drug-Resistant Epilepsy

Shiwei Song
1   Department of Neurosurgery, Fujian Medical University Union Hospital, Fuzhou, Fujian, China
,
Yihai Dai
1   Department of Neurosurgery, Fujian Medical University Union Hospital, Fuzhou, Fujian, China
,
Zhen Chen
1   Department of Neurosurgery, Fujian Medical University Union Hospital, Fuzhou, Fujian, China
,
Songsheng Shi
1   Department of Neurosurgery, Fujian Medical University Union Hospital, Fuzhou, Fujian, China
› Author Affiliations
Funding This study received funds from the Joint Funds for the Innovation of Science and Technology, Fujian province (2018Y9059).

Abstract

Objective The objective of the study was to evaluate the feasibility and accuracy of frameless stereoelectroencephalography (SEEG) electrode implantation in patients with drug-resistant epilepsy using the VarioGuide system.

Methods The VarioGuide frameless navigation system was used to implant SEEG electrodes in patients with medically drug-resistant epilepsy. Demographic data, surgery duration, number of electrodes, and complications were retrospectively analyzed. Accuracy was compared by measuring the distance between the planned and actual electrode positions as determined by postoperative computed tomography images.

Results A total of 141 SEEG electrodes were implanted in 19 patients from May 2015 to December 2018 with an average of 7.42 (range: 4–10) leads per patient. The average entry point localization error (EPLE) was 1.96 ± 0.47 mm (range: 0.32–3.29) and average target point localization error (TPLE) was 2.47 ± 0.79 mm (range: 0.72–4.83). The average operating time per lead (OTPL) was 14.16 ± 2.68 minutes (range: 8.64–21.58). No complications occurred.

Conclusion The VarioGuide frameless navigation system can be an effective method for SEEG electrode implantation in patients with drug-resistant epilepsy, particularly when the electrodes are concentrated in a relatively small region and the number of implanted electrodes is small.



Publication History

Received: 15 January 2020

Accepted: 22 June 2020

Article published online:
22 February 2021

© 2021. Thieme. All rights reserved.

Georg Thieme Verlag KG
Rüdigerstraße 14, 70469 Stuttgart, Germany

 
  • References

  • 1 Katz JS, Abel TJ. Stereoelectroencephalography versus subdural electrodes for localization of the epileptogenic zone: what is the evidence?. Neurotherapeutics 2019; 16 (01) 59-66
  • 2 Abou-Al-Shaar H, Brock AA, Kundu B, Englot DJ, Rolston JD. Increased nationwide use of stereoencephalography for intracranial epilepsy electroencephalography recordings. J Clin Neurosci 2018; 53: 132-134
  • 3 Cardinale F, Casaceli G, Raneri F, Miller J, Lo Russo G. Implantation of stereoelectroencephalography electrodes: a systematic review. J Clin Neurophysiol 2016; 33 (06) 490-502
  • 4 Cardinale F. Talairach methodology in the era of 3D multimodal imaging: “The song remains the same,” but catchier, and therefore more helpful for clinical decision making and surgical planning in epilepsy surgery. Epilepsia 2015; 56 (06) 976-977
  • 5 Gonzalez-Martinez J, Mullin J, Vadera S. et al. Stereotactic placement of depth electrodes in medically intractable epilepsy. J Neurosurg 2014; 120 (03) 639-644
  • 6 Budke M, Avecillas-Chasin JM, Villarejo F. Implantation of depth electrodes in children using VarioGuide® frameless navigation system: technical note. Oper Neurosurg (Hagerstown) 2018; 15 (03) 302-309
  • 7 Vakharia VN, Sparks R, O'Keeffe AG. et al. Accuracy of intracranial electrode placement for stereoencephalography: A systematic review and meta-analysis. Epilepsia 2017; 58 (06) 921-932
  • 8 van der Loo LE, Schijns OEMG, Hoogland G. et al. Methodology, outcome, safety and in vivo accuracy in traditional frame-based stereoelectroencephalography. Acta Neurochir (Wien) 2017; 159 (09) 1733-1746
  • 9 Yu H, Pistol C, Franklin R, Barborica A. Clinical accuracy of customized stereotactic fixtures for stereoelectroencephalography. World Neurosurg 2018; 109: 82-88
  • 10 Dewan MC, Shults R, Hale AT. et al. Stereotactic EEG via multiple single-path omnidirectional trajectories within a single platform: institutional experience with a novel technique. J Neurosurg 2018; 129 (05) 1173-1181
  • 11 Ollivier I, Behr C, Cebula H. et al. Efficacy and safety in frameless robot-assisted stereo-electroencephalography (SEEG) for drug-resistant epilepsy. Neurochirurgie 2017; 63 (04) 286-290
  • 12 Ho AL, Muftuoglu Y, Pendharkar AV. et al. Robot-guided pediatric stereoelectroencephalography: single-institution experience. J Neurosurg Pediatr 2018; 22 (05) 1-8
  • 13 Spyrantis A, Cattani A, Strzelczyk A, Rosenow F, Seifert V, Freiman TM. Robot-guided stereoelectroencephalography without a computed tomography scan for referencing: analysis of accuracy, Int J Med Robot. 2018 14. 02
  • 14 Iordanou JC, Camara D, Ghatan S, Panov F. approach angle affects accuracy in robotic stereoelectroencephalography lead placement. World Neurosurg 2019; 128: e322-e328
  • 15 Candela-Cantó S, Aparicio J, López JM. et al. Frameless robot-assisted stereoelectroencephalography for refractory epilepsy in pediatric patients: accuracy, usefulness, and technical issues. Acta Neurochir (Wien) 2018; 160 (12) 2489-2500
  • 16 Lefranc M, Capel C, Pruvot AS. et al. The impact of the reference imaging modality, registration method and intraoperative flat-panel computed tomography on the accuracy of the ROSA® stereotactic robot. Stereotact Funct Neurosurg 2014; 92 (04) 242-250
  • 17 Brainlab AG. Germany, Software User Guide Revision 1.3 Copyright. 2016: 55