Cool Mist Irrigation Improves Heat Dissipation during Surgical Bone Drilling

Breana R. Siljander
1   Medical School, University of Michigan, Ann Arbor, Michigan, United States
,
Anthony C. Wang
2   Department of Neurosurgery, University of Michigan, Ann Arbor, Michigan, United States
,
Lihui Zhang
3   Department of Mechanical Engineering, University of Michigan, Ann Arbor, Michigan, United States
,
Albert J. Shih
3   Department of Mechanical Engineering, University of Michigan, Ann Arbor, Michigan, United States
4   Department of Biomedical Engineering, University of Michigan, Ann Arbor, Michigan, United States
,
Stephen E. Sullivan
2   Department of Neurosurgery, University of Michigan, Ann Arbor, Michigan, United States
,
Bruce L. Tai
2   Department of Neurosurgery, University of Michigan, Ann Arbor, Michigan, United States
3   Department of Mechanical Engineering, University of Michigan, Ann Arbor, Michigan, United States
› Author Affiliations
Further Information

Publication History

26 November 2013

08 December 2013

Publication Date:
04 April 2014 (online)

Abstract

Objective High-speed drilling generates heat in small cavities and may pose a risk for neurovascular tissues. We hypothesize that a continuous pressurized cold mist could be an alternative approach for better cooling during drilling of bone to access cranial lesions. This study aims to examine this idea experimentally.

Design Ex-vivo drilling tests with controlled speed, feed, and depth were performed on cortical bone samples. Thermocouples were embedded underneath the drilling path to compare the temperature rises under mist cooling (at 3°C, < 300 mL/h) and flood irrigation (at 22°C, > 800 mL/h).

Results A significant difference exists between these two systems (p value < 0.05). The measured temperature was ∼ 4°C lower for mist cooling than for flood irrigation, even with less than a third of the flow rate.

Conclusion Experimental data indicate the capability of mist cooling to reduce heat generation while simultaneously enabling flow reduction and targeted cooling. An improved field of view in an extremely narrow access corridor may be achieved with this technology.

 
  • References

  • 1 Bhatki AM, Pant H, Snyderman CH , et al. The expanded endonasal approach for the treatment of anterior skull base tumors. Oper Tech Otolaryngol 2010; 21: 66-73
  • 2 Kondo S, Okada Y, Iseki H , et al. Thermological study of drilling bone tissue with a high-speed drill. Neurosurgery 2000; 46 (5) 1162-1168
  • 3 Matsumi N, Matsumoto K, Mishima N , et al. Thermal damage threshold of brain tissue—histological study of heated normal monkey brains. Neurol Med Chir (Tokyo) 1994; 34 (4) 209-215
  • 4 Shaw ML, Kelley B, Camarata P, Sokol JA. Collateral damage: heat transfer as a possible mechanism of optic nerve injury during neurosurgical intervention. Ophthal Plast Reconstr Surg 2012; 28 (5) 328-330
  • 5 Abbas GM, Jones RO. Measurements of drill-induced temperature change in the facial nerve during mastoid surgery: a cadaveric model using diamond burs. Ann Otol Rhinol Laryngol 2001; 110 (9) 867-870
  • 6 Sasaki M, Morris S, Goto T, Iwatsuki K, Yoshimine T. Spray-irrigation system attached to high-speed drills for simultaneous prevention of local heating and preservation of a clear operative field in spinal surgery. Neurol Med Chir (Tokyo) 2010; 50 (10) 900-904