Periprothetische Zementverteilung bei femoralen Oberflächenersatzprothesen im Vergleich: Kunstknochen versus humaner Ex-vivo-Knochen. Eine Pilotstudie

Periprosthetic Bone Cement Allocation at the Femoral Durom Hip Prosthesis. A Pilot Study: Artificial Bone Models versus ex-vivo FemuraA. G. Barbe1 , J. Spennes1 , A. Follmann2 , 3 , C. Plumhans3 , M. de la Fuente2 , T. Mumme1
  • 1Klinik für Orthopädie und Unfallchirurgie, Schwerpunkt Orthopädie, Uniklinik der RWTH Aachen
  • 2Helmholtz Institut für Biomedizinische Technik der RWTH Aachen, Lehrstuhl für Medizintechnik
  • 3Klinik für Radiologie, Uniklinik der RWTH Aachen
Further Information

Publication History

Publication Date:
16 August 2010 (online)

Zusammenfassung

Einleitung: Trotz kontroverser Diskussionen wird der künstliche Oberflächenersatz als knochenschonendstes Operationsverfahren des Femurknochens im Bereich der Hüftendoprothetik angesehen. Hauptrisiken dieses Verfahrens stellen die septischen und aseptischen Lockerungen sowie eine Nekrose des koxalen Femurendes dar. Beeinflusst wird dies durch die periprothetische Zementverteilung, welche aufgrund der mangelnden Strahlentransparenz der Oberflächenersatzprothesen nativradiologisch nicht zu beurteilen ist. Durchgeführt wurde eine Pilotstudie zur Validierung der makromechanischen Verzahnung (Interdigitation) des Knochenzements mit spongiösem Knochen am anatomischen Ex-vivo-Präparat, verglichen mit Kunstknochen. Mit der Studie sollte gezeigt werden, dass Untersuchungen der Zementverteilung an Kunstknochen ähnliche Ergebnisse bringen wie die an Ex-vivo-Femura, um so im Verlauf weitere Untersuchungen einfacher durchführen zu können. Methodik: Hierzu wurden an Ex-vivo-Femura (n = 14) versus Kunstknochen (n = 24, Sawbones) dreier verschiedener spongiöser Dichten (0,16; 0,20; 0,32 g/cm³, je n = 8) nach klinisch standardisiertem operativen Verfahren Plastikmodelle aus Polyoxymethylen (POM) entsprechend der Innengeometrie einer Durom Hip (Metasul Durom Hip Prosthesis, Zimmer Germany GmbH, Freiburg, Deutschland) implantiert und die bestückten Probekörper mittels CT-Technik reproduziert und computergestützt ausgewertet. Ergebnisse: Im Vergleich der Sawbones unterschiedlicher Dichte zeigte sich, dass die Sawbones der niedrigeren Dichten (0,16 und 0,20 g/cm³) dem Eindringverhalten von Knochenzement im Vergleich zu den Ex-vivo-Femura am ähnlichsten waren. Zwischen den Sawbones dieser Dichte und den untersuchten Ex-vivo-Femura zeigten sich in Bezug auf die Interdigitation keine signifikanten Unterschiede. Da auch die Ex-vivo-Femura naturgemäß eine natürliche Variationsbreite aufweisen, sind beide Sawbones der o. g. Dichten für weitere Studien zu verwenden.

Abstract

Introduction: Despite controversial discussions, hip resurfacing at the moment seems to be the most bone-sparing surgical procedure of the femoral bone when implanting hip endoprostheses. Main risks are septic and aseptic loosening and necroses of the coxal end of the femoral bone mainly influenced by the periprosthetic allocation of bone cement. Because of a lack of radiation transparency of the hip resurfacing implant, this cement allocation cannot be sufficiently evaluated by common radiological procedures. A pilot study was conducted to describe and validate the macromechanical interdigitation of bone cement with spongy bone of anatomic specimens compared to artificial bone models and to evaluate whether artificial bones may be used for further interdigitation studies of different implantation techniques. Methods: Plastic models of polyoxymethylene (POM) according to the inside geometry of the Metasul Durom hip prosthesis were implanted on ex vivo femora (n = 14) versus artificial bone models (n = 24) (Sawbones) of three different spongy densities (0.16; 0.20; 0.32 g/cm³) (each n = 8) in a clinically standardised surgical procedure and reproduced by highly resolving computed tomography. Afterwards a computer-based analysis of the cement allocation was accomplished. Results: It could be shown that the Sawbones of the lower spongy densities (0.16 and 0.20 g/cm³) were similar to the ex vivo femora regarding the bone penetration of cement. No significant differences could be shown regarding interdigitation. According to our data, both groups of Sawbones may be used for further studies.

Literatur

  • 1 McMinn D, Treacy R, Lin K et al. Metal on metal surface replacement of the hip. Experience of the McMinn prosthesis.  Clin Orthop Relat Res. 1996;  329 (Supp.) 89-98
  • 2 Amstutz H C, Grigoris P, Dorey F J. Evolution and future of surface replacement of the hip.  J Orthop Sci. 1998;  3 169-186
  • 3 Vall T P, Mina C A, Yergler J D et al. Metal-on-metal hip resurfacing compares favourably with THA at 2 years followup.  Clin Orthop Relat Res. 2006;  453 123-131
  • 4 Steffen R T, Smith S R, Urban J P et al. The effect of hip resurfacing on oxygen concentration in the femoral head.  J Bone Joint Surg [Br]. 2005;  87 1468-1474
  • 5 Shimmin A J, Back D. Femoral neck fractures following Birmingham hip resurfacing: a national review of 50 cases.  J Bone Joint Surg [Br]. 2005;  87 463-464
  • 6 Morlock M M, Bishop N, Ruther W et al. Biomechanical, morphological, and histological analysis of early failures of hip resurfacing arthroplasty.  Proc Inst Mech Eng (H). 2006;  220 333-344
  • 7 Campbell P, Beaule P, Ebramzadeh E et al. A study of implant failure in metal-on-metal surface arthroplasties.  Clin Orthop Relat Res. 2006;  453 35-46
  • 8 Le Duff M J, Amstutz H C, Dorey F J. Metal-on-metal hip resurfacing for obese patients.  J Bone Joint Surg [Am]. 2007;  89 2705-2711
  • 9 Beaule P E, Dorey F J, Le Duff M J et al. Risk factors affecting outcome of metal-on-metal surface arthroplasty of the hip.  Clin Orthop Relat Res. 2004;  418 87-93
  • 10 Gill H S, Campbell P A, Murray D W et al. Reduction of the potential for thermal damage during hip resurfacing.  J Bone Joint Surg [Br]. 2007;  89 16-20
  • 11 Bitsch R, Loidolt T, Heisel C et al. Cementing techniques for hip resurfacing arthroplasty: an In-vitro-Study.  J Arthroplasty. 2008;  23 329

Dr. med. Anna Greta Barbe

Klinik für Orthopädie und Unfallchirurgie, Schwerpunkt Orthopädie
Uniklinik der RWTH Aachen

Pauwelsstraße 30

52074 Aachen

Phone: 02 41/8 03 50 68

Fax: 02 41/8 08 07 88

Email: annagretabarbe@gmx.de

    >