Methods Inf Med 1990; 29(04): 413-423
DOI: 10.1055/s-0038-1634810
ECG Interpretation Systems
Schattauer GmbH

Value of Scatter-Graphs for the Assessment of ECG Computer Measurement Results

D. Morlet
1   INSERM U 121, Hôpital Cardiovas-culaire, Lyon, France
,
P. Rubel
2   INSA, Dept. Informatique, Villeur-banne, France
,
J. L. Willems
3   CSE Coordinating Center, University Hospital Gasthuisberg, Leuven, Belgium
› Institutsangaben
The CSE Concerted Action Project was supported by the Commission of the European Communities (E.C.) within the framework of its Medical and Public Health Research Program and by different funding agencies in various E.C. Member States. We are grateful to Dr. J. Carew for reading the manuscript, and the skil-full assistance of Miss M.T. Rameaux in typing is also greatly appreciated.
Weitere Informationen

Publikationsverlauf

Publikationsdatum:
06. Februar 2018 (online)

Abstract

A new method is presented, different from usual methods, for the discussion of the results of computer ECG measurement programs, based on a new graphical evaluation method. The proposed “scatter-graphs” both highlight the main program results and facilitate the comparison between various wave-recognition algorithms. They allow the distinction between the reliability of an algorithm, which is its capacity to provide a maximum of measurements with a minimum amount of errors, and its precision, i.e., the standard deviation of the differences between its point estimates and the reference.

The method proves to be a powerful tool for the discussion of individual, as well as the median program results in CSE Measurement Study. It allows to highlight limitations induced in performance assessment by the variability of the reference itself, and to state that the median program is at least as close to the “Gold Standard” as the median of the referees.

 
  • REFERENCES

  • 1 Willems JL, Arnaud P, Degani R, MacFar-lane PW, van Bemmel JH, Zywietz C. Protocol for the Concerted Action Project “Common Standards for Quantitative Electrocardiography”. 2nd. R. & D. Program in the Field of Medical and Public Health Research of the E.E.C. (80/344/ EEC), C.S.E. Ref. 80-06-00. Leuven: Acco Publ; 1980
  • 2 Willems JL, Arnaud P, van Bemmel JH. et al. Establishment of a reference library for evaluating computer ECG measurement programs. Comp Biomed Res 1985; 18: 439-57.
  • 3 The CSE. Working Party. Recommendations for measurement standards in quantitative electrocardiography. Europ Heart J 1985; 06: 815-25.
  • 4 Morlet D, Rubel P, Arnaud P, Willems JL. An improved method to evaluate the precision of computer ECG measurement programs. Int J Biomed Comp 1988; 22: 199-216.
  • 5 Willems JL, Arnaud P, Van Bemmel JH. et al. Assessment of the performance of electrocardiographic computer programs with the use of a reference data base. Circulation 1985; 71: 523-34.
  • 6 Willems JL. Common Standards for Quantitative Electrocardiography. 2nd. Progress Report. Leuven: Acco Publ; 1982
  • 7 Willems JL. Common Standards for Quantitative Electrocardiography. 3rd. Progress Report. Leuven: Acco Publ; 1983
  • 8 The CSE. Working Party. A reference database for multilead electrocardiographic computer measurement programs. JACC 1987; 10: 1313-21.
  • 9 Rubel P, Ayad B. The true boundary recognition power of multidimensional detection functions. An optimal comparison. In: Willems JL, Van Bemmel JH, Zywietz C. eds. Computer ECG Analysis: Towards Standardization. Amsterdam: North-Holland Publ Comp; 1986: 97-103.
  • 10 Ayad B. Optimisation du Prétraitement des Electrocardiogrammes. Algorithmes et Methodes (These de Doctorat d’Ingénieur en Informatique et Automatique). No IDI 1 8716 INSA de Lyon. 1987