Simultaneous Frame-assisted Stereotactic Placement of Subdural Grid Electrodes and Intracerebral Depth Electrodes

Peter C. Reinacher
1   Department of Stereotactic and Functional Neurosurgery, Freiburg University Medical Center, Freiburg, Germany
2   Medical Faculty, Freiburg University, Freiburg, Germany
,
Dirk-Matthias Altenmüller
2   Medical Faculty, Freiburg University, Freiburg, Germany
3   Department of Epileptology, Freiburg University Medical Center, Freiburg, Germany
,
Marie T. Krüger
2   Medical Faculty, Freiburg University, Freiburg, Germany
4   Department of Neurosurgery, Freiburg University Medical Center, Freiburg, Germany
,
Andreas Schulze-Bonhage
2   Medical Faculty, Freiburg University, Freiburg, Germany
3   Department of Epileptology, Freiburg University Medical Center, Freiburg, Germany
,
Horst Urbach
2   Medical Faculty, Freiburg University, Freiburg, Germany
5   Department of Neuroradiology, Freiburg University Medical Center, Freiburg, Germany
,
Michael Trippel
1   Department of Stereotactic and Functional Neurosurgery, Freiburg University Medical Center, Freiburg, Germany
2   Medical Faculty, Freiburg University, Freiburg, Germany
,
Volker A. Coenen
1   Department of Stereotactic and Functional Neurosurgery, Freiburg University Medical Center, Freiburg, Germany
2   Medical Faculty, Freiburg University, Freiburg, Germany
,
Daniel Delev
2   Medical Faculty, Freiburg University, Freiburg, Germany
4   Department of Neurosurgery, Freiburg University Medical Center, Freiburg, Germany
› Author Affiliations
Further Information

Publication History

19 July 2018

17 October 2018

Publication Date:
13 May 2019 (online)

Abstract

Background and Study Aims In complex cases of drug-resistant focal epilepsy, the precise localization of the epileptogenic zone requires simultaneous implantation of depth and subdural grid electrodes. This study describes a new simple frame-assisted method that facilitates the simultaneous placement of both types of intracranial electrodes.

Material and Methods Ten consecutive patients were evaluated and divided into two groups. Group A included patients with simultaneous frame-assisted placement of depth and subdural grid electrodes. In group B, depth electrodes were implanted stereotactically; grid electrodes were implanted in a separate surgery.

Results The placement of the subdural grid was accurate as individually designed by the epileptologists in all five patients from group A. In group B, one patient showed a slight and another one a significant deviation of the subdural grid position postoperatively. The mean surgical time in group A was shorter (280±62 minutes) compared with the mean duration of the surgical procedures in group B (336±51 minutes).

Conclusion The frame-assisted placement of subdural grid electrodes facilitates the surgical procedure for invasive video-electroencephalography monitoring in complex cases of drug-resistant focal epilepsy in which a combination of depth electrodes and subdural grid electrodes is needed, by reducing the surgical time and guaranteeing highly accurate electrode localizations.

 
  • References

  • 1 Wiebe S, Blume WT, Girvin JP, Eliasziw M. ; Effectiveness and Efficiency of Surgery for Temporal Lobe Epilepsy Study Group. A randomized, controlled trial of surgery for temporal-lobe epilepsy. N Engl J Med 2001; 345 (05) 311-318
  • 2 Kral T, Clusmann H, Urbach J. , et al. Preoperative evaluation for epilepsy surgery (Bonn algorithm). Zentralbl Neurochir 2002; 63 (03) 106-110
  • 3 Rosenow F, Lüders H. Presurgical evaluation of epilepsy. Brain 2001; 124 (Pt 9): 1683-1700
  • 4 Pardoe H, Kuzniecky R. Advanced imaging techniques in the diagnosis of nonlesional epilepsy: MRI, MRS, PET, and SPECT. Epilepsy Curr 2014; 14 (03) 121-124
  • 5 Kim S, Mountz JM. SPECT imaging of epilepsy: an overview and comparison with F-18 FDG PET. Int J Mol Imaging 2011; 2011 (10) 813028
  • 6 Desai A, Bekelis K, Thadani VM. , et al. Interictal PET and ictal subtraction SPECT: sensitivity in the detection of seizure foci in patients with medically intractable epilepsy. Epilepsia 2013; 54 (02) 341-350
  • 7 Vézina L-G. MRI-negative epilepsy: protocols to optimize lesion detection. Epilepsia 2011; 52 (Suppl. 04) 25-27
  • 8 Surbeck W, Bouthillier A, Weil AG. , et al. The combination of subdural and depth electrodes for intracranial EEG investigation of suspected insular (perisylvian) epilepsy. Epilepsia 2011; 52 (03) 458-466
  • 9 Ortler M, Sohm F, Eisner W. , et al. Frame-based vs frameless placement of intrahippocampal depth electrodes in patients with refractory epilepsy: a comparative in vivo (application) study. Neurosurgery 2011; 68 (04) 881-887 ; discussion 887
  • 10 González-Martínez J, Bulacio J, Thompson S. , et al. Technique, results, and complications related to robot-assisted stereoelectroencephalography. Neurosurgery 2016; 78 (02) 169-180
  • 11 Enatsu R, Bulacio J, Najm I. , et al. Combining stereo-electroencephalography and subdural electrodes in the diagnosis and treatment of medically intractable epilepsy. J Clin Neurosci 2014; 21 (08) 1441-1445
  • 12 Mehta AD, Labar D, Dean A. , et al. Frameless stereotactic placement of depth electrodes in epilepsy surgery. J Neurosurg 2005; 102 (06) 1040-1045
  • 13 Bjartmarz H, Rehncrona S. Comparison of accuracy and precision between frame-based and frameless stereotactic navigation for deep brain stimulation electrode implantation. Stereotact Funct Neurosurg 2007; 85 (05) 235-242
  • 14 Chamoun RB, Nayar VV, Yoshor D. Neuronavigation applied to epilepsy monitoring with subdural electrodes. Neurosurg Focus 2008; 25 (03) E21
  • 15 Van Gompel JJ, Meyer FB, Marsh WR, Lee KH, Worrell GA. Stereotactic electroencephalography with temporal grid and mesial temporal depth electrode coverage: does technique of depth electrode placement affect outcome?. J Neurosurg 2010; 113 (01) 32-38
  • 16 Munyon CN, Koubeissi MZ, Syed TU, Lüders HO, Miller JP. Accuracy of frame-based stereotactic depth electrode implantation during craniotomy for subdural grid placement. Stereotact Funct Neurosurg 2013; 91 (06) 399-403
  • 17 Shah AK, Mittal S. Invasive electroencephalography monitoring: indications and presurgical planning. Ann Indian Acad Neurol 2014; 17 (05) (Suppl. 01) S89-S94
  • 18 Dorfer C, Stefanits H, Pataraia E. , et al. Frameless stereotactic drilling for placement of depth electrodes in refractory epilepsy: operative technique and initial experience. Neurosurgery 2014; 10 (Suppl. 04) 582-590 ; discussion 590–591
  • 19 Nowell M, Rodionov R, Diehl B. , et al. A novel method for implementation of frameless StereoEEG in epilepsy surgery. Neurosurgery 2014; 10 (Suppl. 04) 525-533 ; discussion 533–534