Subscribe to RSS
DOI: 10.1055/s-0038-1624329
Three-Dimensional ROI-Based Quantification of Stress/Rest 201TI Myocardial SPECT: Presentation of Method
Publication History
Received:
02 July 1986
Publication Date:
12 January 2018 (online)
A semiautomated method for three-dimensional quantification of stress/rest 201TI myocardial SPECT of the left ventricle (LV) is introduced. A total of 33 ringsegment-shaped regions of interest (ROIs) are distributed over 6 oblique (short axis cuts) double-slices (thickness 1.3 cm) from LV apex to base, commencing with 1 ROI for the apex up to 8 ROIs for the base (1-4-4-8-8-8). Within the last two double slices an additional adjacent ROI is automatically placed in the upper right quadrant to observe lung Tl-uptake. Correspondence of slices for exercise and rest is achieved by adjusting the long axis outlines rebuilt from summed profiles of 15 oblique slices of each examination. Calculated numeric arrays of the extracted parameters (vitality, washout and redistribution) are “displayed in a target-like two-dimensional polar fashion.
Zusammenfassung
Vorgestellt wird ein neues, halbautomatisches Verfahren zur quantitativen Auswertung der myokardialen 201T1 SPECT mit Belastung. In je 6 aufeinanderfolgenden, zur Herzlängsachse senkrechten Doppelschichten (Dicke ca. 1,3 cm) von Belastungsund Ruheuntersuchung, wird die linksventrikuläre Tl-Speicherung durch interaktiv positionierte kreisund ringsegmentförmige »regions of interest« (ROIs) erfaßt. Von der Herzspitze zur Basis erhöht sich die Anzahl der ROIs in der Folge 1-4-4-8- 8-8. Die Quantifizierung der Tl-Speicherung der Lunge erfolgt mit Hilfe einer automatisch erzeugten viertelsegmentförmigen ROI, die im rechten oberen Bildquadranten der beiden basisnächsten Schnitte zu liegen kommt. Die örtliche Entsprechung der regionalen Tl-Speicherung (Belastung gegenüber Ruhe) wird erreicht, indem die myokardialen »Summationskonturen « (erzeugt aus den winkelsummierten Profilen von jeweils 15 Schnitten) von Belastungs- und Ruheaufnahme zur Übereinstimmung gebracht werden. Die Darstellung der errechneten regionalen Parameter (Vitalität, Washout und Redistributionsindex) erfolgt als zweidimensionale winkelgetreue Projektion der räumlichen ROI-Anordnung in Form einer »Zielscheibe«.
-
REFERENCES
- 1 Büll U, Kleinhans E, Seiderer M. et al. Quantitative assessment of thallium-201 images. Cardiovasc. Radiol 1979; 02: 183-93.
- 2 Büll U, Kirsch C. M., Höfling B. et al. Ergebnisse der 201-Thallium SPECT des Myokards zur Erfassung der koronaren Herzkrankheit und zur Verlaufskontrolle bei perkutaner transluminaler koronarer AngioplastiK. Nucl. Med 1986; 25: 71-7.
- 3 Eisner R, Gullberg G, Malko J. et al. Effects of 180 acquisition on tomographic image quality. J. nucl. Med 1984; 25: 30.
- 4 Garcia E. V., van Train K, Maddahi J. et al. Quantification of rotational thallium- 201 myocardial tomography. J. nucl. Med 1985; 26: 17-26.
- 5 Go R. T., Cook S. A., Maclntyre W. J.. et al. Comparative accuracy of stress and redistribution thallium-201 cardiac single photon emission transaxial tomography and planar imaging in the diagnosis of myocardial ischemia. J. nucl. Med 1982; 23: p25.
- 6 Go R. T., Maclntyre W. J., Houser T. S.. et al. Clinical evaluation of 360 and 180 data sampling techniques for transaxial SPECT thallium-201 myocardial perfusion imaging. J. nucl. Med 1985; 26: 695-706.
- 7 Hirzel H. O., Nüsch K, Grüntzig A. R.. et al. Short- and long-term changes in myocardial perfusion after percutaneous transluminal coronary angioplasty assessed by thallium-201 exercise scintigraphy. Circulation 1981; 63: 1001-7.
- 8 Hör G, Maul F. D.. Beitrag der Myokardszintigraphie in der Therapiekontrolle (gegenwärtiger Stand und Ausblikke). Z. Kardiol 1985; 74: 65-75.
- 9 Iskandrian A. S., Haaz W, Segal B. L.. et al. Exercise thallium-201 scintigraphy in evaluation aortocoronary bypass surgery. Chest 1981; 80: 11.
- 10 Kirsch C. M., Doliwa R, Büll U. et al. Sensitivität und Spezifität von Belastungsuntersuchungen des Herzens in SPECT Technik. Der Nuklearmediziner 1985; 03: 241-9.
- 11 Kirsch C. M., Moser E, Büll U. Quantitative Auswertung von 201T1-Myokardszintigrammen in SPECT Technik. NUC-Comp 1985; 16: 230-4.
- 12 Knapp W. H., Zimmermann R, Tillmanns H. et al. Processing of 201TI-serial images for the assessment of myocardial clearance and redistribution. Eur. J. nucl. Med 1982; 07: 248.
- 13 Laird E. E., Rajathurai A, Williams E. D.. et al. Quantitative analysis of rotating gamma camera thallium-201 scintigrams of myocardium. Nucl. Med. Comm 1984; 05: 577-86.
- 14 Meindl S, Adam W, Pfannenstiel P. Weiterentwicklung der quantitativen Auswertung der Myokardszintigraphie mit 201T1. Nucl. Med 1983; 22: 63-7.
- 15 Standke R, Hör G. Sektoranalyse der myokardialen 201-Thallium-Restributionskinetik. NUC-Comp 1980; 11: 249-55.
- 16 Tamaki N, Yonekura Y, Mukai T. et al. Values and limitations of segmental analysis of stress and redistribution T1 ECT for location of coronary artery disease. J. nucl. Med 1982; 24: 18.