Thromb Haemost 2002; 88(03): 459-461
DOI: 10.1055/s-0037-1613238
Review Article
Schattauer GmbH

A Reassessment of the Relationship between International Reference Preparations for Human and Rabbit Thromboplastins

M. H. P. A. van den Besselaar
1   Haemostasis and Thrombosis Research Centre, Leiden University Medical Centre, Leiden, The Netherlands
,
A. Tripodi
2   A. Bianchi Bonomi Hemophilia and Thrombosis Centre, University and IRCCS Maggiore Hospital, Milan, Italy
› Author Affiliations
Further Information

Publication History

Received 05 April 2002

Accepted 07 May 2002

Publication Date:
08 December 2017 (online)

Summary

Two established international reference preparations (IRP) for thromboplastins, i.e. RBT/90 (rabbit, plain) and rTF/95 (recombinant human, plain) have been calibrated against each other in a 7-centre exercise performed in 2000. The purpose of the study was to compare the calibration results with those of the original calibration study performed in 1995. The international sensitivity index (ISI) for rTF/95 was calculated relative to the ISI for RBT/90. The mean ISI for rTF/95 was 0.961 (between-lab SD 0.028) which was 2% greater than the historical value determined in 1995. There is no indication that the two IRP have deteriorated, but further monitoring of the calibration relationship is recommended.

 
  • References

  • 1 Tripodi A, Chantarangkul V, Negri B, Clerici M, Mannucci PM. International collaborative study for the calibration of a proposed reference preparation for thromboplastin, human recombinant, plain. Thromb Haemost 1998; 79: 439-43.
  • 2 WHO Expert Committee on Biological Standardization. Guidelines for thromboplastins and plasma used to control oral anticoagulant therapy. WHO Technical Report Series. no. 889. 1999: 64-93.
  • 3 Tomenson JA. A statistician’s independent evaluation. In: Thromboplastin calibration and oral anticoagulation. Van den Besselaar AMHP, Gralnick HR, Lewis SM. eds. Martinus Nijhoff Publishers: Boston; 1984: 87-108