Successful closure of a complicated endoscopic ultrasound-related duodenal perforation

José Celso Ardengh
Center for Gastrointestinal Endoscopy, Department of Surgery and Anatomy, Ribeirão Preto Medical School, University of São Paulo, Ribeirão Preto, São Paulo, Brazil
,
Roberto Mundim Porto Filho
Center for Gastrointestinal Endoscopy, Department of Surgery and Anatomy, Ribeirão Preto Medical School, University of São Paulo, Ribeirão Preto, São Paulo, Brazil
,
Vitor Ottoboni Brunaldi
Center for Gastrointestinal Endoscopy, Department of Surgery and Anatomy, Ribeirão Preto Medical School, University of São Paulo, Ribeirão Preto, São Paulo, Brazil
,
José Eduardo Brunaldi
Center for Gastrointestinal Endoscopy, Department of Surgery and Anatomy, Ribeirão Preto Medical School, University of São Paulo, Ribeirão Preto, São Paulo, Brazil
,
João Almiro Ferreira Filho
Center for Gastrointestinal Endoscopy, Department of Surgery and Anatomy, Ribeirão Preto Medical School, University of São Paulo, Ribeirão Preto, São Paulo, Brazil
,
Rafael Kemp
Center for Gastrointestinal Endoscopy, Department of Surgery and Anatomy, Ribeirão Preto Medical School, University of São Paulo, Ribeirão Preto, São Paulo, Brazil
,
José Sebastião dos Santos
Center for Gastrointestinal Endoscopy, Department of Surgery and Anatomy, Ribeirão Preto Medical School, University of São Paulo, Ribeirão Preto, São Paulo, Brazil
› Author Affiliations
Further Information

Corresponding author

Vitor Ottoboni Brunaldi, MD, MSc
Bandeirantes Avenue 3900
Monte Alegre 14049-900
Ribeirão Preto
São Paulo
Brazil   
Fax: +55-11-30617000   
Email: vbrunaldi@gmail.com   

Publication History

Publication Date:
09 September 2019 (online)

 

The incidence of duodenal perforation during endosonography (EUS) ranges from 0 % to 0.04 % [1]. The tip of the linear endoscope may perforate the duodenum as a consequence of angulation and the presence of stenosis or diverticula [1] [2]. Unfortunately, in general, surgical repair is needed [3].

We report here an extensive perforation of the second duodenal portion after linear EUS, treated in an original and successful manner by the use of an endoloop and metallic endoclips [4]. The procedure was performed in a 72-year-old woman with abdominal and lumbar pain, a mass in the left hypochondrium, a 3-kg weight loss, and an expansive injury to the body/tail of the pancreas measuring about 10.7 × 8.9 × 7.2 cm, suggesting the presence of a serous cystadenoma upon computed tomography ([Fig. 1]).

Zoom Image
Fig. 1 Tomographic image of an abdominal mass with numerous confluent cysts in a central lobulated calcified area of large proportions.

EUS was indicated for the assessment of vascular invasion. During the procedure, retraction and deformity were observed in the duodenal bulb with reduced distensibility, and an attempted transposition resulted in a large perforation of the duodenal wall ([Fig. 2]). The perforation was closed with endoclips, which were applied to the entire circumference of the lesion, involving the margin of the lesion and an endoloop ([Video 1]). The endoloop was then closed in order to obtain coaptation of the edges of the perforation ([Fig. 3]). A nasoenteral tube and nasogastric tube were positioned for feeding and drainage, respectively.

Zoom Image
Fig. 2 Endoscopic image of a perforation caused by the passage of a linear transducer from the echoendoscope to the second duodenal portion.

Video 1 Closure of an endoscopic ultrasound-related large duodenal perforation using a combination of an endoloop and endoclips.


Quality:
Zoom Image
Fig. 3 Endoscopic image after endoloop closure with all clips interposed one on top of the other, with coaptation of the margins of the perforation.

The patient recovered well and was discharged from the hospital on the 4th postoperative day. The nasogastric and nasoenteral tubes were removed on the 3rd and 20th postoperative days, respectively.

Endoscopic repair using only clips could be technically difficult due to the tangential angle between the bulb and the second portion, the diameter of the lesion, and the accommodation of the margins of the perforation between the rods of the clip. The use of clips for the fixation of an endoloop to the margins of the lesion provided complete obliteration of the perforated area and good results after their activation. This technique, used in cases of perforation after endoscopic retrograde cholangiopancreatography [4] [5], appears to be an alternative for the repair of perforations caused by endoscopic procedures.

Endoscopy_UCTN_Code_CPL_1AL_2AB

Endoscopy E-Videos
https://eref.thieme.de/e-videos

Endoscopy E-Videos is a free access online section, reporting on interesting cases and new techniques in gastroenterological endoscopy. All papers include a high quality video and all contributions are freely accessible online.
This section has its own submission website at https://mc.manuscriptcentral.com/e-videos


#

Competing interests

None

  • References

  • 1 Lachter J. Fatal complications of endoscopic ultrasonography: a look at 18 cases. Endoscopy 2007; 39: 747-750
  • 2 Liu Y, Wang D, Li Z. Endoscopic closure for EUS and ERCP related duodenal perforation by endoclips. Gastroenterol Res Pract 2016; 2016: 1051597
  • 3 Ince AT, Kocaman O, Yildiz K. et al. Endoscopic repair of duodenal perforation with over-the-scope clipping system and endoclips: a case report. Turk J Gastroenterol 2014; 25: 103-105
  • 4 Lee TH, Han JH, Park SH. Endoscopic treatments of endoscopic retrograde cholangiopancreatography-related duodenal perforations. Clin Endosc 2013; 46: 522-528
  • 5 Nakagawa Y, Nagai T, Soma W. et al. Endoscopic closure of a large ERCP-related lateral duodenal perforation by using endoloops and endoclips. Gastrointest Endosc 2010; 72: 216-217

Corresponding author

Vitor Ottoboni Brunaldi, MD, MSc
Bandeirantes Avenue 3900
Monte Alegre 14049-900
Ribeirão Preto
São Paulo
Brazil   
Fax: +55-11-30617000   
Email: vbrunaldi@gmail.com   

  • References

  • 1 Lachter J. Fatal complications of endoscopic ultrasonography: a look at 18 cases. Endoscopy 2007; 39: 747-750
  • 2 Liu Y, Wang D, Li Z. Endoscopic closure for EUS and ERCP related duodenal perforation by endoclips. Gastroenterol Res Pract 2016; 2016: 1051597
  • 3 Ince AT, Kocaman O, Yildiz K. et al. Endoscopic repair of duodenal perforation with over-the-scope clipping system and endoclips: a case report. Turk J Gastroenterol 2014; 25: 103-105
  • 4 Lee TH, Han JH, Park SH. Endoscopic treatments of endoscopic retrograde cholangiopancreatography-related duodenal perforations. Clin Endosc 2013; 46: 522-528
  • 5 Nakagawa Y, Nagai T, Soma W. et al. Endoscopic closure of a large ERCP-related lateral duodenal perforation by using endoloops and endoclips. Gastrointest Endosc 2010; 72: 216-217

Zoom Image
Fig. 1 Tomographic image of an abdominal mass with numerous confluent cysts in a central lobulated calcified area of large proportions.
Zoom Image
Fig. 2 Endoscopic image of a perforation caused by the passage of a linear transducer from the echoendoscope to the second duodenal portion.
Zoom Image
Fig. 3 Endoscopic image after endoloop closure with all clips interposed one on top of the other, with coaptation of the margins of the perforation.