Introduction
Transmission of multi-antibiotic-resistant bacteria from contaminated endoscopes has
prompted healthcare facilities to culture their endoscopes in an attempt to detect
contamination of organisms of high concern so that these endoscopes can be removed
from use and reprocessed to eliminate infection transmission [1]
[2]
[3]
[4]
[5]
[6]
[7]. The newly released US Food and Drug Administration/Centers for Disease Control
and Prevention/American Society for Microbiology (FDA/CDC/ASM) duodenoscope sample
collection and culture protocol has been validated by the endoscope manufacturers
and provides a standardized approach to culture duodenoscopes [8].
Unlike endoscope culture protocols from other countries, the FDA/CDC/ASM culture protocol
is similar to that of Ross et al. (2015) [3] and indicates that ideally once a culture sample has been taken, the endoscope should
be quarantined until the results of the culture are known, so that if the endoscope
was contaminated, patient exposure during subsequent procedures could be avoided [8]. However, this culture and quarantine approach severely limits the number of clinical
procedures that can be performed. Instead of being able to use an endoscope multiple
times each day, with the quarantine process, it can only be used for one procedure
every 2 to 3 days (depending on how long the culture results take). Culture of endoscopes
is also fraught with additional standardization issues such as the sample collection
method used (i. e. type of friction, type of extraction fluid, use of neutralizer)
[9]
[10]
[11]
[12]
[13] and the culture protocol used (i. e. concentration of sample, culture media and
duration of incubation) affect the sensitivity of cultures [9]
[10]
[11]
[12]
[13]
[14]
[15]
[16]
[17]. Although culture is recommended in many countries as a monitoring tool for endoscope
reprocessing, [11]
[18]
[19] it is not currently recommended in United States guidelines [20]
[21]
[22] and there is evidence that there can be false-negative culture results [3]
[23]. Many of these issues might be avoided if rapid test methods could be used to reliably
detect endoscope channel contamination post-high-level disinfection (HLD) but prior
to patient use.
A recent review [24] listed a range of commercially available rapid tests that detect organic residuals
(e. g. protein, hemoglobin) and those that detect adenosine triphosphate (ATP) for
assessing adequacy of endoscope cleaning. The rapid ATP test kits have been most widely
studied for monitoring cleaning adequacy of flexible endoscopes [2]
[25]
[26]
[27]. Data suggesting cut-offs for adequate cleaning of flexible endoscopes using rapid
ATP or organic residuals have been published [25]
[26]
[27]
[28]. These cleaning monitoring approaches allow the opportunity to re-clean endoscopes
that have not been properly cleaned before they are disinfected. Some authors have
questioned whether some of these rapid cleaning monitoring test methods can reliably
detect contamination in endoscope channels after HLD [14]
[15]. There have been very limited assessments of organic residuals post-HLD and clinical
studies have reported variable results using the currently available rapid ATP kits
to detect potentially contaminated patient-used endoscopes post-disinfection [10]
[27]
[29]
[30]
[31]. The majority of published studies indicate that the ATP test is too insensitive
to detect low levels of viable bacteria in samples extracted from endoscope channels
after HLD [2]
[10]
[25]
[26]
[27]
[29].
There are few other tests available that reliably detect viable bacteria post-HLD
or sterilization prior to patient use. The only other commercially available test
for evaluating endoscope channels for viable bacteria is the rapid NOW! (RN) test.
The manufacturer’s instruction for use (MIFU) indicates that this test can detect
as low as 10 CFU of Gram-negative bacteria within endoscope channels. The MIFU indicates
the test is not applicable to Gram-positive bacteria as it is based on detection of
a specific enzyme found only in Gram-negative bacteria. The sample collected from
endoscopes needs to be incubated approximately 12 hours so the MIFU suggests that
samples be collected from fully reprocessed endoscopes at the end of the day, incubated
overnight and read the following morning before any patient procedures. If there is
a test failure detected in the morning, the endoscope can be reprocessed prior to
patient use. However (to the best of the authors knowledge), there is only one peer-reviewed
publication on this technology [32] and they did not provide specific data for this test in their publication.
Therefore, the objectives of this study were to determine if after overnight incubation
the RN test satisfies the 10-CFU limit of detection claim in the MIFU and whether
the test can reliably detect low levels of viable Gram-negative bacteria within duodenoscope
instrument channels using simulated-use testing.
Materials and methods
Bacterial strains
The bacterial strains used included Enterococcus faecalis (ATCC 29212), Escherichia coli (ATCC 25922) and Pseudomonas aeruginosa (ATCC 15442). Each isolate was cultured on blood agar (BA) media and incubated aerobically
at 35 oC for 24 hours prior to use in an experimental test.
Simulated-use study
Duodenoscopes and automated endoscope reprocessors (AERs)
A TJF-Q180V model (sealed elevator guidewire) Olympus duodenoscope (Olympus Corporation
of the Americas, Center Valley, Pennsylvania, United States) was used for simulated-use
testing. After each test, the duodenoscope was cleaned as per Olympus MIFU and then
processed using the STERIS SYSTEM 1E (SS1E) (STERIS Inc, Mentor, Ohio, United States)
and stored in a channel purge storage cabinet (Torvan Inc, Toronto, Ontario, Canada).
The duodenoscope instrument channel was inoculated by injecting 1 mL of 10 % blood
containing 104 CFU/mL or 102 CFU/mL of E. coli into the distal opening of the suction channel and then raising the distal end allowing
the inoculum to flow towards the biopsy port. The inoculated duodenoscope was allowed
to dry for 2 hours prior to sample collection.
Rapid NOW! (RN) test
Duodenoscope testing
Samples from the inoculated duodenoscopes were collected by flushing 5 mL of extraction
fluid (sterile water) from the RN test kit through the biopsy port of the instrument
channel and the extracted sample was collected from the distal end into a sterile
container. A small volume of extraction fluid was used to ensure that the most concentrated
channel sample was available for testing. A 0.5-mL portion of the extracted channel
sample was inoculated into the test cuvette and incubated overnight as per MIFU. The
following day, test reagent was added to the cuvette, which was then placed into the
fluorometer and the numeric fluorometer value (NFV) recorded. Baseline testing (i. e.
negative endoscope testing) of the fully reprocessed, uninoculated duodenoscope was
also done.
The MIFU indicates that an NFV < 200 indicates no viable Gram-negative bacteria, an
NFV 200 to 300 indicates the likely presence of viable Gram-negative bacteria and
an NFV > 300 indicates presence of viable Gram-negative bacteria. The MIFU indicates
that the RN test detects the activity of a specific enzyme found only in viable Gram-negative
bacteria and therefore does not detect other types of microorganisms that are not
Gram-negative bacteria. The MIFU indicates that the limit of detection for the RN
test is 10 CFU in the sample tested.
Suspension testing
The RN test (Healthmark Industries, Fraser, MI) was evaluated following the MIFU.
The limit of detection was assessed using varying separate concentrations of E. faecalis, E. coli and P. aeruginosa prepared in sterile reverse osmosis (sRO) water. A 0.5-mL portion of each concentration
of bacteria was inoculated into the test cuvette and incubated overnight as per the
MIFU. The following day test reagent was added to the cuvette, which was then placed
into the fluorometer and the NFV was recorded.
Viable count for cultures
The inoculum for suspension testing and also for simulated-use testing were each serially
diluted 1:10 using sterile phosphate buffered saline (PBS), and then 0.1 mL of the
direct, as well as each dilution, was spread over the surface of BA plates. Plates
were incubated at 35 °C and colony counts were performed at 24 hours. All tests were
performed in triplicate.
Statistical analysis
The t test was used to assess the area under the Receiver Operator Characteristics (ROC)
curve (AUROCC) to determine if AUROCC was significantly different from 0.5. The greater
AUROCC, the more accurate the test. AUROCC of 0.5 suggests that categorization by
an evaluated test is similar to random categorization and that the evaluated test
is not a useful test. Detection cutoffs of 1 CFU as well as 10 CFU were evaluated.
Sensitivity, specificity, positive and negative predictive values (NPV) were calculated.
Results
The data in [Table 1] are based on simulated-use testing to assess how well the RN test can detect bacteria
inoculated into a duodenoscope instrument channel. The RN test MIFU endoscope sample
collection consists of a squeezable plastic bottle containing 5 mL fluid that is flushed
through the instrument channel (i. e. does not incorporate friction in the sample
extraction protocol). The estimated maximum CFU recoverable in the 0.5-mL samples
that were tested by the RN test ranged from 32 to 39 CFU and all three replicates
gave a positive test result ([Table 1]).
Table 1
Simulated-use evaluation of the RN test for detection of E. coli in duodenoscope instrument channel.
Replicate number
|
CFU Inoculum per channel
|
Maximum Total CFU Tested[1]
|
NOW! test Incubation time (Hrs)
|
Numeric fluorometer value (NFV)
|
Test 1: Low inoculum
|
1
|
320.00
|
32.00
|
21.75
|
2376.00
|
2
|
360.00
|
36.00
|
21.67
|
2810.00
|
3
|
390.00
|
39.00
|
22.25
|
1832.00
|
Average:
|
356.70
|
35.67
|
21.89
|
2339.33
|
SD:
|
35.12
|
3.51
|
0.31
|
490.03
|
Test 2: Moderate inoculum
|
1
|
15600.00
|
1560.00
|
20.00
|
2178.00
|
2
|
30000.00
|
3000.00
|
20.25
|
4534.00
|
3
|
22000.00
|
2200.00
|
21.92
|
4405.00
|
Average:
|
22500.00
|
2253.33
|
21.39
|
3705.67
|
SD:
|
7214.80
|
721.48
|
1.04
|
1324.57
|
Bacteria were suspended in (10 % sheep blood) and the instrument channel of a duodenoscope
was inoculated and then allowed to dry for 2 hours before extraction as per MIFU (see
Materials and Methods for details). The negative control from the test kit showed
an NFV of 0 and 10 % blood without any bacteria had an NFV of 3.
1 The inoculum per channel was extracted in 5 mL extraction fluid so the maximum CFU
tested is calculated as (CFU inoculated/5 mL) × 0.5 mL (volume of extracted channel
sample that is tested).
Detection results for the RN test in suspension testing are shown in [Table 2] for E. coli, P. aeruginosa and E. faecalis. As indicated in the MIFU, the Gram-positive bacteria E. faecalis did not generate consistent NFVs over 300 whereas the Gram-negative bacteria tested
did have high NFVs even at very low CFU. All nine suspensions with E. coli and/or P. aeruginosa ≥ 10 CFU were correctly detected by the RN test. For the cut-off of 1 CFU there was
only one of the 24 tests that was incorrect (i. e. P. aeruginosa at 2.35 CFU that was a false-negative). In contrast, for E. faecalis, only one of three suspensions with CFU ≥ 10 tested positive with the RN Test (NFV > 300)
and none of the three suspensions with CFU between ≥ 1 and 10 tested positive.
Table 2
Performance of the RN test for various bacterial concentrations suspended in sRO water.
Numeric fluorometer value (NFV) for various CFU in suspension
|
Test 1[1]
|
Test 2[1]
|
Test 3[1]
|
Total CFU tested
|
NFV
|
Total CFU tested
|
NFV
|
Total CFU tested
|
NFV
|
Enterococcus faecalis
|
13.50
|
178.00
|
20.00
|
331.00
|
15.50
|
43.00
|
1.35
|
84.00
|
2.00
|
145.00
|
1.55
|
255.00
|
0.14
|
240.00
|
0.20
|
320.00
|
0.16
|
270.00
|
0.01
|
193.00
|
0.02
|
121.00
|
0.02
|
246.00
|
Escherichia coli
|
150.00
|
14373.00
|
225.00
|
15966.00
|
105.00
|
15348.00
|
15.00
|
10825.00
|
22.50
|
10052.00
|
10.50
|
11891.00
|
1.50
|
8389.00
|
2.25
|
9753.00
|
1.05
|
2109.00
|
0.15
|
117.00
|
0.23
|
1.00
|
0.11
|
124.00
|
Pseudomonas aeruginosa
|
16.00
|
9551.00
|
23.50
|
24196.00
|
22.00
|
10896.00
|
1.60
|
13333.00
|
2.35
|
51.00
|
2.20
|
12036.00
|
0.16
|
13.00
|
0.24
|
38.00
|
0.22
|
27.00
|
0.02
|
70.00
|
0.02
|
0.00
|
0.02
|
5.00
|
1 Tests 1, 2 and 3 were independent bacterial suspensions prepared on three different
days and each of the three organisms was evaluated separately (i. e. each microbe
tested separately). Any Total CFU tested for a Gram-negative bacterium that is < 1
CFU would be considered to be a negative sample.
Negative control test using sRO water showed an NFV of 92, 40 and 0 on three separate
test days.
Therefore, sensitivity for detection of CFU ≥ 10 was 100 % for both E. coli and P. aeruginosa and for CFU ≥ 1 was 100 % for E coli, 83 % for P. aeruginosa and 93 % for either. The AUROCC for E. coli was 1 for both CFU ≥ 1 as well as CFU ≥ 10 and for P. aeruginosa was 0.85 for CFU ≥ 10 and 0.97 for CFU ≥ 1 (P < 0.0001). Importantly, when the E. faecalis suspensions were included as negative controls in the analysis, the sensitivity of
NFV > 300 for detection of CFU ≥ 1 Gram-negative bacteria was 93 % and specificity
was 90 %, NPV was 95.0 % and positive predictive value was 88 %; the AUROCC was 0.95
(P < 0.0001).
Discussion
Because it was not feasible to form biofilm within a flexible endoscope, we chose
to perform simulated-use testing whereby an organic matrix containing 10 % blood containing
low to moderate levels of E. coli was used to inoculate a duodenoscope instrument channel that was then allowed to
dry for 2 hours. Collecting and testing samples extracted from the inoculated instrument
channel confirmed that RN test provided a very strong NFV of over 2300. It should
be noted that this signal was generated by a maximum of 32 to 39 CFU in the 0.5 mL
of extracted sample that was used for the RN test. Furthermore, we were able to demonstrate
that up to 10 % blood did not interfere with the ability of the test to generate a
high NFV.
Our study is the first to demonstrate that in suspension testing, the RN Test does
detect very low levels of Gram-negative bacteria but (as stated in the MIFU) cannot
reliably detect Gram-positive bacteria. Our data support the manufacturer’s claim
that the RN test can detect as low as 10 CFU of Gram-negative bacteria in the sample
tested. In addition, most of the samples with Gram-negative bacteria CFU between 1
and 10 were also detected as positive by the RN test.
Our previous data showed that for 6639 CFU/mL of P. aeruginosa extracted from biofilm, there was only 1 RLU by the ATP test [9]. Furthermore, it took 7,424,795 CFU/mL extracted from biofilm to generate 1 μg/mL
protein [9]. These data demonstrated that although rapid ATP and rapid protein testing are very
useful for monitoring manual cleaning compliance, they are very insensitive and cannot
act as a substitute for culture post-HLD [2]
[10]
[27]
[29] Unlike rapid ATP and protein testing that do not detect low levels of viable bacteria
[9]
[23]
[25]
[26]
[27]
[29] the RN test was sensitive enough to reliably detect nine of nine suspension tests
when there were ≥ 10 CFU of viable E. coli and P. aeruginosa in the 0.5-mL sample tested. This is an excellent limit of detection but it must
be borne in mind that all the current guidelines for interpretation of endoscope cultures
[8]
[18]
[19] indicate that even 1 CFU per channel of a Gram-negative bacteria is considered unacceptable.
Of interest, the RN test correctly detected 14 of 15 of the tests for both of these
Gram-negative bacteria when there was ≥ 1 CFU in the 0.5-mL sample tested. All nine
suspensions with Gram-negative bacteria < 1 CFU (i. e. considered as a negative sample)
had NFV readings < 300. Although there are a number of Gram-positive bacteria that
are organisms of concern [8]
[18]
[19], there have been no infectious outbreaks attributed to such bacteria arising from
contaminated endoscopes, so utilizing a test that only detects Gram-negative organisms
of concern maybe a reasonable alternative option to routine universal culturing. Potential
advantages of RN test compared to cultures are a more rapid test (overnight compared
with 48- to 72-hour hold for cultures), lack of need for addition of a neutralizer
and concentration of the samples (neutralizer and sample concentration increase sensitivity
of cultures but have no role with RN test) and ability to process in the endoscopy
unit itself (compared with cultures which require transportation to a Microbiology
laboratory).
Although 10 % blood did not generate a NFV ([Table 1]), it may affect the optical NFV reading as the signal generated from samples with
10 % blood ([Table 1]; 2,253.33 CFU of E. coli had NFV of 3,705) were not as high as those generated when the same organism was
in sRO water ([Table 2]; 105 CFU E. coli had NFV of 15,348). This aspect did not affect the ability of the RN test to detect
low levels of viable Gram-negative bacteria, however, further assessment of the effect
of organic residuals on the RN test is needed.
A limitation of this study is that we only evaluated two Gram-negative bacteria so
further studies to show the applicability of the RN test for a wider range of Gram-negative
bacteria would be valuable. The MIFU does not indicate the extraction efficacy of
the 5-mL flush with extraction fluid. In addition, we did not evaluate the extraction
efficacy of flushing 5 mL through an instrument channel containing dried organic debris,
biofilm or build-up biofilm. Future clinical studies are needed to evaluate the feasibility
of implementation of the RN test for patient-used endoscopes in a busy endoscopy clinic.
The RN test cannot be used for testing endoscopes used more than once during the day
as this test requires overnight incubation. In addition, it is unclear if testing
endoscopes used on a Friday would require reading of RN test results on Saturday or
whether the test could still provide reliable results if the completion of testing
was done Monday morning. Despite these limitations, the RN test is the only currently
available test that can provide reliable detection of 1 to 10 CFU with an overnight
turnaround time. Our data support the recent study by Washburn et al. who reported
that the RN test had similar results compared to culture [32].
Conclusion
In conclusion we have demonstrated that the RN test can reliably detect ≥ 10 CFU (and
most of ≥ 1 CFU) of Gram-negative bacteria which is superior to what current rapid
ATP and protein test kits can detect. Our encouraging preliminary data suggest that
the RN test may be a useful alternative to culture for assessing contamination with
Gram-negative bacteria in patient-ready endoscopes but further evaluation of extraction
efficiency, effect of organic residuals and clinical testing is needed.