CC BY-NC-ND 4.0 · J Lab Physicians 2019; 11(02): 103-106
DOI: 10.4103/JLP.JLP_141_18
Original Article

M2G1G2 white blood cell flag by three-part automated hematology analyzer: A hint to dengue infection in appropriate clinical context

Maitrayee Roy
Department of Pathology, M.M. Institute of Medical Sciences and Research, Ambala, Haryana, India
,
Akshay Bali
Department of Hematology, Maitri Diagnostic Lab, Ambala, Haryana, India
› Author Affiliations
Financial support and sponsorship: Nil.

Abstract

BACKGROUND: Automated hematology analyzers often generate many flags which can provide important clues to the underlying hematological abnormality. Although pathologists are generally well versed in recognizing the importance of flags indicating potential leukemic blasts, their utility in hinting toward infectious etiology, especially during epidemics, is less well known. We analyzed any abnormal flags generated by a three-part automated hematology analyzer in serologically confirmed cases of dengue.

MATERIALS AND METHODS: The study included 28 patients diagnosed with serologically confirmed dengue infection. The venous samples were run on ABX Miros-60 three-part hematology analyzer. The complete blood count data and any abnormal flags were noted and correlated with peripheral blood film findings in all patients.

RESULTS: An abnormal white blood cell (WBC) flag was noted in all 28 patients, including two in whom all other hematological parameters were within normal limits. In 26 (93%) patients, M2G1G2 WBC flag was noted while the samples of the remaining two patients generated M2G1 and L1M2G1G2 WBC flags, respectively.

CONCLUSION: An abnormal WBC flag, when correlated in appropriate clinical context, especially during a dengue outbreak, can aid in targeting the cohort of patients who will require immediate referral for serological confirmation.



Publication History

Received: 20 October 2018

Accepted: 22 March 2019

Article published online:
06 April 2020

© 2019.

Thieme Medical and Scientific Publishers Private Ltd.
A-12, Second Floor, Sector -2, NOIDA -201301, India

 
  • References

  • 1 Green R, Wachsmann-Hogiu S. Development, history, and future of automated cell counters. Clin Lab Med 2015;35:1-10.
  • 2 Asad S, Ahmed I, Ali N. Utility of peripheral film findings and its correlation with automated analyzer – An audit from tertiary care hospital. J Lab Physicians 2017;9:1-4.
  • 3 Kline A, Bird A, Adams L, Wale C, Edwards F, Perreira E. Identification of blast cells in the peripheral blood of patients with acute leukaemia using the technicon H-1. Clin Lab Haematol 1989;11:111-6.
  • 4 Hoyer JD, Fisher CP, Soppa VM, Lantis KL, Hanson CA. Detection and classification of acute leukemia by the coulter STKS hematology analyzer. Am J Clin Pathol 1996;106:352-8.
  • 5 Aulesa C, Pastor I, Naranjo D, Galimany R. Application of receiver operating characteristics curve (ROC) analysis when definitive and suspect morphologic flags appear in the new coulter LH 750 analyzer. Lab Hematol 2004;10:14-23.
  • 6 Kakkar N, Kaur R. Utility of white blood cell suspect flags and histogram pattern in the detection of acute leukemia by advia-60 automated hematology analyzer. Indian J Pathol Microbiol 2004;47:322-6.
  • 7 Guarner J, Atuan MA, Nix B, Mishak C, Vejjajiva C, Curtis C, et al. Process to evaluate hematological parameters that reflex to manual differential cell counts in a pediatric institution. Clin Lab 2010;56:21-7.
  • 8 Kunal O, Anne W. White blood cell suspect flags and white blood cell histogram pattern in acute leukemias. Panacea J Med Sci 2012;2:27-30.
  • 9 Yoon J, Yun SG, Kim J, Jung YN, Koh YE, Nam JH, et al. Evaluation of the LH780 hematology analyzer for detection and therapeutic monitoring of malaria: Cross-reactivity with nucleated RBCs. Acta Trop 2016;164:425-30.
  • 10 Sharma P, Bhargava M, Sukhachev D, Datta S, Wattal C. LH750 hematology analyzers to identify malaria and dengue and distinguish them from other febrile illnesses. Int J Lab Hematol 2014;36:45-55.
  • 11 Jain M, Gupta S, Jain J, Grover RK. Usefulness of automated cell counter in detection of malaria in a cancer set up – Our experience. Indian J Pathol Microbiol 2012;55:467-73.
  • 12 Sharma S, Sethi N, Pujani M, Kushwaha S, Sehgal S. Abnormal WBC scattergram: A clue to the diagnosis of malaria. Hematology 2013;18:101-5.
  • 13 Mehta RC, Goswami HM, Katara RK, Patel PS, Parikh UV, Vegad MM, et al. Importance of complete blood count and peripheral smear examination in early diagnosis of dengue patients. J Infect Dis Lett 2013;2:22-4.
  • 14 Khatri K, Rajani A, Khalla AR. Plasmacytoid lymphocytes: A diagnostic clue to dengue infection. Int J Sci Res 2016;5:1002-5.
  • 15 Oehadian A, Michels M, de Mast Q, Prihatni D, Puspita M, Hartantri Y, et al. New parameters available on sysmex XE-5000 hematology analyzers contribute to differentiating dengue from leptospirosis and enteric fever. Int J Lab Hematol 2015;37:861-8.