Subscribe to RSS
DOI: 10.3414/ME09-02-0046
Analysis of Fluorescence Lifetime Imaging Microscopy (FLIM) Data
Based on a Fully Automated Laguerre Deconvolution MethodPublication History
received:
10 October 2009
accepted:
23 January 2010
Publication Date:
17 January 2018 (online)
Summary
Objectives: A novel Fluorescence Lifetime Imaging Microscopy (FLIM) deconvolution method based on the linear expansion of fluorescence decays on a set of orthonormal Laguerre functions was recently proposed. The Laguerre deconvolution method applies linear least-square estimation to estimate the expansion coefficients of all pixel decays simultaneously, performing at least two orders of magnitude faster than the other algorithms. In the original Laguerre FLIM deconvolution implementation, however, the Laguerre parameter α is selected using a heuristic approach, making it unsuitable for online applications.
Methods: In this study, we present a fully automated implementation of the Laguerre FLIM deconvolution, whereby the Laguerre parameter α is treated as a free parameter within a nonlinear least-squares optimization scheme.
Results: The performance of this method has been successfully validated on simulated data, and experimental FLIM images of standard fluorescent dyes and endogenous tissue fluorescence.
Conclusions: The main advantage of the proposed method is that it does not require any user intervention for tuning up the deconvolution process. Thus, we believe this method will facilitate the translation of FLIM to online applications, including real-time clinical diagnosis.
-
References
- 1 Ramanujan VK, Jo JA, Cantu G, Herman BA. Spatially resolved fluorescence lifetime mapping of enzyme kinetics in living cells. J Microsc 2008; 230 Pt 3 329-338.
- 2 Marmarelis VZ. Identification of nonlinear biological systems using Laguerre expansions of kernels. Ann Biomed Eng 1993; 21 (06) 573-589.
- 3 Dabir AS, Trivedi CA, Ryu Y, Pande P, Jo JA. Fully automated deconvolution method for on-line analysis of time-resolved fluorescence spectroscopy data based on an iterative Laguerre expansion technique. J Biomed Opt 2009; 14 (02) 024030.
- 4 Lakowicz J. Principles of Fluorescence Spectroscopy. Springer; 2006
- 5 Fang QY, Papaioannou T, Jo JA, Vaitha R, Shastry K, Marcu L. Time-domain laser-induced fluorescence spectroscopy apparatus for clinical diagnostics. Review of Scientific Instruments 2004; 75 (01) 151-162.
- 6 Arakawa K, Isoda K, Ito T, Nakajima K, Shibuya T, Ohsuzu F. Fluorescence analysis of biochemical constituents identifies atherosclerotic plaque with a thin fibrous cap. Arterioscler Thromb Vasc Biol 2002; 22 (06) 1002-1007.
- 7 Marcu L, Jo JA, Fang Q, Papaioannou T, Reil T, Qiao JH. et al. Detection of rupture-prone athero-sclerotic plaques by time-resolved laser-induced fluorescence spectroscopy. Atherosclerosis 2009; 204 (01) 156-164.