Simultaneous Measurement of Ventricular Fluid and Fontanelle Pressure in Neonates and Infants with Hydrocephalus

Simultane Registrierung des Ventrikel- und Fontanellendruckes beim Neugeborenen und jungen Säugling mit nativem HydrozephalusG.  Kaiser , H.  Minikus
  • University Children's Hospital Berne, Department of Paediatric Surgery
Further Information

Publication History

Publication Date:
25 March 2008 (online)

Abstract

Simultaneous measurement of fontanelle and ventricular fluid pressure was performed during 1-2 hours in 17 neonates and infants with hydrocephalus at a mean age of 7 weeks on 19 occasions. The fontanelle pressure recorded by a Ladd sensor was on the average 1.1 mmHg lower than the ventricular fluid pressure and differed from it by a standard deviation of ± 3.5 mmHg. The most important problem in continuous fontanometry is the evaluation of the precise application pressure at the beginning of the examination. Fontanometry is for the present appropriate to trend analysis and screening examination, but not to measure the exact intracranial pressure.

Zusammenfassung

Bei 17 Neugeborenen und Säuglingen mit Hydrozephalus wurde im Durchschnittsalter von 7 Wochen 19mal eine simultane Messung des Fontanellen- und Ventrikeldruckes während 1-2 Stunden vorgenommen. Der mit dem Ladd-Monitor registrierte Fontanellendruck ist im Schnitt 1.1 mmHg niedriger als der ventrikuläre Liquordruck und variiert von diesem um ± 3.5 mmHg. Das größte Problem bei der kontinuierlichen Messung des Fontanellendruckes ist die Ermittlung des korrekten Applikationsdruckes zu Beginn der Untersuchung. Die Fontanometrie ist zur Zeit für Trendanalysen und als Screening-Verfahren geeignet, jedoch nicht für eine exakte Bestimmung des intrakraniellen Druckes.

    >