Placental Abruption Following Placement of Disposable Intrauterine Pressure Transducer System

Anthony C. Sciscione, James S. Manley, Marie E. Pinizzotto, Garrett H.C. Colmorgen
  • Section of Maternal-Fetal Medicine, Department of Obstetrics and Gynecology, Medical Center of Delaware, Christiana Hospital, Newark, Delaware
Further Information

Publication History

Publication Date:
04 March 2008 (online)

ABSTRACT

There have been few reported complications of intrauterine pressure monitoring. We present four cases of placental abruption following insertion of a disposable intrauterine pressure transducer system (INTRAN). All four patients had placental abruptions, which were evident soon after insertion of the disposable intrauterine pressure transducer system (all of our cases had predisposing factors or signs of placental abruption at presentation). Whether there was a preexisting placental abruption, which was caused, revealed, or worsened by insertion of the disposable intrauterine pressure transducer system is unclear. We encourage careful patient selection with gentle insertion of the system opposite the placental site.