Endoscopy 1990; 22(6): 241-244
DOI: 10.1055/s-2007-1012861
Original Contributions

© Georg Thieme Verlag KG Stuttgart · New York

The Management of Malignant Esophago-Respiratory Fistulas with a Cuffed Prosthesis

M. L. Hordijk, J. Dees, M. van Blankenstein
  • Department of Internal Medicine II, University Hospital Rotterdam “Dijkzigt”, Rotterdam, The Netherlands
Further Information

Publication History

Publication Date:
17 March 2008 (online)

Summary

Peroral intubation with a standard prosthesis fails to occlude inoperable malignant esophago-respiratory fistulas located above the narrowed segment, or when there is little or no stricture at all. Ten patients with malignant esophago-respiratory fistulas were intubated perorally with a prosthesis surrounded by a foam ruber cuff contained in a silicone sheath, in which a vacuum can be created. The type of fistula was esophago-bronchial in 6 patients, esophago-tracheal in 3 patients, and gastro-tracheal in 1 patient. The fistula was located above the stricture in 8 patients and in a malignancy without a stricture in 2 patients. The diameter of the opening of the fistula ranged between 1 and 3 cm. Complications with the cuffed tube were recurrent leakage after a period of 30 days in situ in 1 patient, and compression of the trachea immediately after intubation in 1 patient. Adequate palliation, that is, no symptoms related to the fistula, good transit of food and discharge from hospital within a few days was achieved in 9 patients for an average period of 37.2 days (10-84 days). The cuffed esophageal prosthesis can improve the quality of life in patients with malignant esophago-respiratory fistulas that do not respond to conventional intubation.