Rofo 2002; 174(6): 696-699
DOI: 10.1055/s-2002-32221
Rapid Communication
Originalarbeit
© Georg Thieme Verlag Stuttgart · New York

Average glandular dose with amorphous silicon
full-field digital mammography - Clinical results

Mittlere Parenchymdosis bei der digitalen Vollfeldmammographie mit einem Detektor aus amorphem Silizium - Klinische ErgebnisseK.-P.  Hermann, S.  Obenauer, K.  Marten, S.  Kehbel, U.  Fischer, E.  Grabbe
  • 1Department of Diagnostic Radiology, Georg-August-University of Göttingen
Further Information

Publication History

Publication Date:
13 June 2002 (online)

Zusammenfassung

Ziel: Bestimmung der mittleren Parenchymdosis für ein großes Patientenkollektiv bei der digitalen Vollfeldmammographie mit einem Flachdetektor auf der Basis von amorphem Silizium. Material und Methoden: Das Patientenkollektiv umfasste Frauen, die in einem Zeitraum von 4 Monaten mit dem digitalen Mammographiesystem Senographe 2000D untersucht wurden. Die Zahl der Frauen betrug 591, die Zahl der Aufnahmen 1116. Es wurden nur kranio-kaudale Projektionen berücksichtigt. Während jeder Aufnahme wurden verschiedene Kenngrößen, darunter die Einfalldosis, das mAs-Produkt und die Kompressionsschichtdicke ermittelt. Auf dieser Basis wurde die mittlere Parenchymdosis unter Verwendung tabellierter Konversionsfaktoren g für eine angenommene Standardzusammensetzung der Brust berechnet. Ergebnisse: Die durchschnittliche mittlere Parenchymdosis betrug 1,51 mGy (0,66 - 4,05 mGy) für eine Einzelaufnahme. Die mittlere Kompressionsschichtdicke lag bei 55,7 mm. Das mittlere Alter der Patientinnen war 55 Jahre (34 - 81 Jahre). Schlussfolgerung: Die Ergebnisse zeigen, dass die digitale Vollfeldmammographie mit einem Flachdetektor auf der Basis von amorphem Silizium im Vergleich zur konventionellen Film-Folien-Mammographie ca. 25% weniger Dosis benötigt.

Abstract

Purpose: Determination of average glandular dose with a full-field digital mammography system using a flat-panel X-ray detector based on amorphous silicon technology for a large group of patients. Material and Methods: The patient group includes women who were examined in a 4-month period with the digital mammographic system Senographe 2000D. The number of women was 591 and the number of exposures was 1116; only cranio-caudal projections were considered. Various quantities, including entrance surface air kerma, tube loading, and compressed breast thickness, were determined during actual mammography. Average glandular dose was determined using conversion factors g for standard breast composition. Results: The mean average glandular dose was 1.51 mGy (0.66 - 4.05 mGy) for a single view. The mean compressed breast thickness was 55.7 mm. The mean age of patients was 55 years (34 - 81 years). Conclusion: The results demonstrate that full-field digital mammography with a flat-panel detector based on amorphous silicon needs about 25 % less dose in comparison with conventional screen-film mammography.


Literatur

  • 1 Hermann K-P, Obenauer S, Grabbe E. Radiation exposure in full-field digital mammography with a flat-panel X-ray detector based on amorphous silicon in comparison with conventional screen-film mammography.  Fortschr Röntgenstr. 2000;  172 940-945
  • 2 Hendrick R E, Berns E A. Optimization of technique factors for a silicon diode array full-field digital mammography system and comparison to screen-film mammography with matched mean glandular dose.  Radiology. 2001;  221(P) 338-339
  • 3 Hermann K-P, Obenauer S, Funke M, Grabbe E. Full-field digital mammography: Effect of target-filtration and peak voltage settings on lesion detection and breast dose.  Radiology. 2001;  221(P) 339
  • 4 European Protocol on Dosimetry in Mammography. Lucembourg; European Commission 1996
  • 5 Kruger R L, Schueler B A. A survey of clinical factors and patient dose in mammography.  Med Phys. 2001;  28 1449-1454
  • 6 Gennaro G, di Maggio C, Bellan E. Full-field digital mammography and patient dose reduction.  Eur Radiol. 2002;  12(S) 163
  • 7 Diekmann F, Diekmann S, Bick U, Hamm B. Reduced-dose digital mammography of skin calcifications.  Am J Roentgenol. 2002;  178 473-474

Dr. rer. nat. Klaus-Peter Hermann

Department of Diagnostic Radiology, Georg-August-University
of Göttingen

Robert-Koch-Straße 40

37075 Göttingen

Germany

Phone: + 49-551-397964

Fax: + 49-551-399606

Email: kherman@gwdg.de

    >