Effect of Plate Length on Construct Stiffness and Strain in a Synthetic Short-Fragment Fracture Gap Model Stabilized with a 3.5-mm Locking Compression Plate

Fabian N. Trefny
1   Division of Health Sciences, School of Veterinary Medicine, Murdoch University, Perth, Western Australia, Australia
,
1   Division of Health Sciences, School of Veterinary Medicine, Murdoch University, Perth, Western Australia, Australia
,
Giselle Hosgood
1   Division of Health Sciences, School of Veterinary Medicine, Murdoch University, Perth, Western Australia, Australia
,
Alex Hayes
2   Department of Medical Engineering and Physics, Royal Perth Hospital, Perth, Australia
,
Robert Day
2   Department of Medical Engineering and Physics, Royal Perth Hospital, Perth, Australia
› Author Affiliations
Funding The authors received partial financial support for the implants from DePuy Synthes.

Abstract

Objective To evaluate the effect of 3.5-mm locking compression plate (LCP) length on construct stiffness and plate and bone model strain in a synthetic, short-fragment, fracture-gap model.

Study Design Six replicates of 6-hole, 8-hole, 10-hole, and 12-hole LCP constructs on a short-fragment, tubular Delrin fracture gap model underwent four-point compression and tension bending. Construct stiffness and surface strain, calculated using three-dimensional digital image correlation, were compared across plate length and region of interest (ROI) on the construct.

Results The 12-hole plates (80% plate–bone ratio) had significantly higher construct stiffness than 6-hole, 8-hole, and 10-hole plates and significantly lower plate strain than 6-hole plates at all ROIs. Strain on the bone model was significantly lower in constructs with 10-hole and 12-hole plates than 6-hole plates under both compression and tension bending.

Conclusion Incremental increases in construct stiffness and incremental decreases in plate strain were only identified when comparing 6-hole, 8-hole, and 10-hole plates to 12-hole plates, and 6-hole to 12-hole plates, respectively. Strain on the bone model showed an incremental decrease when comparing 6-hole to 10-hole and 12-hole plates. A long plate offered biomechanical advantages of increased construct stiffness and reduced plate and bone model strain, over a short plate in this in vitro model.

Authors' Contribution

All authors contributed to the conception and design of this study. F.T., M.G., R.D., and A.H. were involved with data acquisition and material testing. R.D. and A.H. provided technical support with material testing and use of software. G.H. contributed to acquisition of data and data analysis. All authors contributed to data interpretation and revision of the submitted manuscript.


Note

An abstract of this paper was presented at the annual European College of Veterinary Surgeons Congress, July 6, 2019, Budapest, Hungary. Results from this study were also presented at the AO Vet Symposium, Bridging the Gap: Translating Clinical Research to Clinical Practice, presented November 23, 2020, via Webinar.




Publication History

Received: 27 February 2024

Accepted: 06 August 2024

Article published online:
21 August 2024

© 2024. Thieme. All rights reserved.

Georg Thieme Verlag KG
Stuttgart · New York

 
  • References

  • 1 Gautier E, Sommer C. Guidelines for the clinical application of the LCP. Injury 2003; 34 (Suppl. 02) B63-B76
  • 2 Stoffel K, Dieter U, Stachowiak G, Gächter A, Kuster MS. Biomechanical testing of the LCP–how can stability in locked internal fixators be controlled?. Injury 2003; 34 (Suppl. 02) B11-B19
  • 3 Wagner M. General principles for the clinical use of the LCP. Injury 2003; 34 (Suppl. 02) B31-B42
  • 4 Rozbruch SR, Müller U, Gautier E, Ganz R. The evolution of femoral shaft plating technique. Clin Orthop Relat Res 1998; (354) 195-208
  • 5 Beltran MJ, Collinge CA, Gardner MJ. Stress modulation of fracture fixation implants. J Am Acad Orthop Surg 2016; 24 (10) 711-719
  • 6 Miller DL, Goswami T. A review of locking compression plate biomechanics and their advantages as internal fixators in fracture healing. Clin Biomech (Bristol, Avon) 2007; 22 (10) 1049-1062
  • 7 Sanders R, Haidukewych GJ, Milne T, Dennis J, Latta LL. Minimal versus maximal plate fixation techniques of the ulna: the biomechanical effect of number of screws and plate length. J Orthop Trauma 2002; 16 (03) 166-171
  • 8 Gautier E, Perren SM, Cordey J. Strain distribution in plated and unplated sheep tibia an in vivo experiment. Injury 2000; 31 (Suppl. 03) C37-C44
  • 9 Akeson WH, Woo SLY, Rutherford L, Coutts RD, Gonsalves M, Amiel D. The effects of rigidity of internal fixation plates on long bone remodeling. A biomechanical and quantitative histological study. Acta Orthop Scand 1976; 47 (03) 241-249
  • 10 Chan LWM, Gardner AW, Wong MK, Chua K, Kwek EBK. Singapore Orthopaedic Research CollaborativE (SORCE). Non-prosthetic peri-implant fractures: classification, management and outcomes. Arch Orthop Trauma Surg 2018; 138 (06) 791-802
  • 11 Davenport SR, Lindsey RW, Leggon R, Miclau T, Panjabi M. Dynamic compression plate fixation: a biomechanical comparison of unicortical vs bicortical distal screw fixation. J Orthop Trauma 1988; 2 (02) 146-150
  • 12 Park KC, Lim SJ, Song YS, Hwang KT. Factors affecting peri-implant fracture following locking plate for osteoporotic distal femur fractures. Orthop Traumatol Surg Res 2017; 103 (08) 1201-1204
  • 13 Pearson T, Glyde M, Hosgood G, Day R. The effect of intramedullary pin size and monocortical screw configuration on locking compression plate-rod constructs in an in vitro fracture gap model. Vet Comp Orthop Traumatol 2015; 28 (02) 95-103
  • 14 Pearson T, Glyde MR, Day RE, Hosgood GL. The effect of intramedullary pin size and plate working length on plate strain in locking compression plate-rod constructs under axial load. Vet Comp Orthop Traumatol 2016; 29 (06) 451-458
  • 15 Chao P, Lewis DD, Kowaleski MP, Pozzi A. Biomechanical concepts applicable to minimally invasive fracture repair in small animals. Vet Clin North Am Small Anim Pract 2012; 42 (05) 853-872 , v
  • 16 Bird G, Glyde M, Hosgood G, Hayes A, Day R. Effect of plate type and working length on a synthetic compressed juxta-articular fracture model. VCOT Open 2020; 03 (02) e119-e128
  • 17 Bird G, Glyde M, Hosgood G, Hayes A, Day R. Biomechanical comparison of a notched head locking T-plate and a straight locking compression plate in a juxta-articular fracture model. Vet Comp Orthop Traumatol 2021; 34 (03) 161-170
  • 18 de Bruyn BW, Glyde M, Day R, Hosgood G. Effect of an orthogonal locking plate and primary plate working length on construct stiffness and plate strain in an in vitro fracture-gap model. Vet Comp Orthop Traumatol 2024; 37 (04) 173-180
  • 19 Evans A, Glyde M, Day R, Hosgood G. Effect of plate-bone distance and working length on 2.0-mm locking construct stiffness and plate strain in a diaphyseal fracture gap model: a biomechanical study. Vet Comp Orthop Traumatol 2024; 37 (01) 1-7
  • 20 Amin Yavari S, van der Stok J, Weinans H, Zadpoor AA. Full-field strain measurement and fracture analysis of rat femora in compression test. J Biomech 2013; 46 (07) 1282-1292
  • 21 Palanca M, Tozzi G, Cristofolini L. The use of digital image correlation in the biomechanical area: a review. Int Biomech 2016; 3 (01) 1-21
  • 22 Sztefek P, Vanleene M, Olsson R, Collinson R, Pitsillides AA, Shefelbine S. Using digital image correlation to determine bone surface strains during loading and after adaptation of the mouse tibia. J Biomech 2010; 43 (04) 599-605
  • 23 Väänänen SP, Amin Yavari S, Weinans H, Zadpoor AA, Jurvelin JS, Isaksson H. Repeatability of digital image correlation for measurement of surface strains in composite long bones. J Biomech 2013; 46 (11) 1928-1932
  • 24 Withers PJ. Strain measurement by digital image correlation. Strain 2008; 44 (06) 421-422
  • 25 Weiss DB, Kaar SG, Frankenburg EP, Karunakar MA. Locked versus unlocked plating with respect to plate length in an ulna fracture model. Bull NYU Hosp Jt Dis 2008; 66 (01) 5-8
  • 26 Törnkvist H, Hearn TC, Schatzker J. The strength of plate fixation in relation to the number and spacing of bone screws. J Orthop Trauma 1996; 10 (03) 204-208
  • 27 Golish SR, Mihalko WM. Principles of biomechanics and biomaterials in orthopaedic surgery. J Bone Joint Surg Am 2011; 93 (02) 207-212
  • 28 Liu X, Zhang S, Bao Y, Zhang Z, Yue Z. Strain-controlled fatigue behavior and microevolution of 316L stainless steel under cyclic shear path. Materials (Basel) 2022; 15 (15) 5362-5374
  • 29 Sistaninia M, Niffenegger M. Fatigue crack initiation and crystallographic growth in 316L stainless steel. Int J Fatigue 2015; 70: 163-170
  • 30 Johnson AL, Houlton JEF, Vannini R. AO Principles of Fracture Management in the Dog and Cat. 1st ed. Vol 148. Davos Platz, Switzerland: AO Publishing; 2005
  • 31 Veterinary Product Catalog Implants. Instruments, Sets. DePuy Synth Vet; 2017 . Accessed August 13, 2021 at: https://catalog.synthes.com/getprg.ch?catalogId=VET&chapterId=VET_02&systemId=S_VET_0202.004&programId=S_VET_0200.004.03
  • 32 MacLeod AR, Serrancoli G, Fregly BJ, Toms AD, Gill HS. The effect of plate design, bridging span, and fracture healing on the performance of high tibial osteotomy plates: an experimental and finite element study. Bone Joint Res 2019; 7 (12) 639-649
  • 33 Niemeyer P, Südkamp NP. Principles and clinical application of the locking compression plate (LCP). Acta Chir Orthop Traumatol Cech 2006; 73 (04) 221-228
  • 34 Perren SM. Evolution of the internal fixation of long bone fractures. The scientific basis of biological internal fixation: choosing a new balance between stability and biology. J Bone Joint Surg Br 2002; 84 (08) 1093-1110
  • 35 Garofolo S, Pozzi A. Effect of plating technique on periosteal vasculature of the radius in dogs: a cadaveric study. Vet Surg 2013; 42 (03) 255-261
  • 36 Pozzi A, Lewis D. Surgical approaches for minimally invasive plate osteosynthesis in dogs. Vet Comp Orthop Traumatol 2009; 22 (04) 316-320
  • 37 Pozzi A, Hudson CC, Gauthier CM, Lewis DD. Retrospective comparison of minimally invasive plate osteosynthesis and open reduction and internal fixation of radius-ulna fractures in dogs. Vet Surg 2013; 42 (01) 19-27
  • 38 Sharir A, Barak MM, Shahar R. Whole bone mechanics and mechanical testing. Vet J 2008; 177 (01) 8-17
  • 39 Maxwell M, Horstman CL, Crawford RL, Vaughn T, Elder S, McLaughlin R. The effects of screw placement on plate strain in 3.5 mm dynamic compression plates and limited-contact dynamic compression plates. Vet Comp Orthop Traumatol 2009; 22 (02) 125-131
  • 40 Zahn K, Frei R, Wunderle D. et al. Mechanical properties of 18 different AO bone plates and the clamp-rod internal fixation system tested on a gap model construct. Vet Comp Orthop Traumatol 2008; 21 (03) 185-194