Rofo 2016; 188(06): 566-573
DOI: 10.1055/s-0042-104204
Interventional Radiology
© Georg Thieme Verlag KG Stuttgart · New York

Interventional Removal of Intravascular Medical Devices: Methods and Technical Success

Interventionelle Bergung intravasaler Fremdkörper: Methoden und technischer Erfolg
I. Ayx
1   Radiology, KH Barmherzige Brüder, Regensburg, Germany
,
H. Goessmann
2   Radiology, University Hospital Regensburg, Germany
,
H. Hubauer
1   Radiology, KH Barmherzige Brüder, Regensburg, Germany
,
W. Uller
2   Radiology, University Hospital Regensburg, Germany
,
I. Wiesinger
2   Radiology, University Hospital Regensburg, Germany
,
C. Uhl
3   Vascular Surgery, KH Barmherzige Brüder, Regensburg, Germany
,
I. Töpel
3   Vascular Surgery, KH Barmherzige Brüder, Regensburg, Germany
,
N. Zorger
3   Vascular Surgery, KH Barmherzige Brüder, Regensburg, Germany
› Author Affiliations
Further Information

Publication History

29 October 2015

17 February 2016

Publication Date:
19 April 2016 (online)

Abstract

Purpose: Evaluation of the technical success rate and complications when retrieving dislocated intravascular foreign bodies.

Material and Methods: Between 1999 and 2015 38 patients (21 female; 17 male; Age: 17 – 92; Average 54.3 years) underwent an extraction of intravascular dislocated foreign bodies, which were not lost during a radiological intervention. The extracted material included 29 port catheters, 3 tips of tunneled dialysis catheters, 2 stents, 2 guide wires, 1 CVC tip and 1 AS occluder device. Various catheters for repositioning and extraction were used. The access was transarterial as well as transvenous. Technical success was defined as complete removal of the foreign body.

Results: The technical success rate was 92.1 % (35 of 38). In 17 patients an additional catheter was necessary to reposition the foreign body in order to make it accessible for the extraction catheter. In one case a stent was relocated and remodeled within the patient and was not extracted. In another case we experienced a dislocation of a small fragment of the port catheter into the distal parts of the pulmonary artery, which couldn’t be extracted. A guide wire could not be extracted as it was already adhered with the vessel wall. Peri-interventional complications were not documented.

Conclusion: The percutaneous extraction of dislocated intravascular foreign bodies is technically successful and poor of complications. Interventional therapy can avoid surgical removal.

Key points:

• The percutaneous extraction of dislocated intravascular foreign bodies is technically successful and safe.

• In most cases surgical removal can be avoided.

• The gooseneck-snare catheter was mainly used for the extraction of intravascular foreign bodies.

Citation Format:

• Ayx I, Goessmann H, Hubauer H et al. Interventional Removal of Intravascular Medical Devices: Methods and Technical Success. Fortschr Röntgenstr 2016; 188: 566 – 573

Zusammenfassung

Ziel: Evaluation des technischen Erfolgs und der Komplikationsrate bei der Bergung von intravasal disloziertem Fremdmaterial.

Material und Methoden: Zwischen 1999 und 2015 wurde bei 38 Patienten (21 Frauen, 17 Männer; Alter: 17 – 92 Jahre; Durchschnitt: 54,3 Jahre) eine Bergung von intravasal disloziertem Fremdmaterial durchgeführt, welches nicht im Rahmen einer radiologischen Intervention verloren wurde. Bei 29 Portkathetern, 3 getunnelten Dialysekatheter, 2 Stents, 2 Führungsdrähten, einer ZVK-Spitze und einem Schirmchen zur Okklusion eines Vorhofseptumdefekts wurden verschiedene Repositionskatheter, Extraktionskatheter und Zugänge (arteriell, venös) verwendet.

Ergebnisse: Die vollständige Bergung des Fremdmaterials war in 92,1 % erfolgreich (35 von 38 Fällen). In 17 Fällen war eine Reposition mittels eines Hilfskatheters nötig, um das Fremdmaterial dem Bergekatheter zugänglich zu machen. Ein Stent wurde intrakorporal anmodelliert und nicht geborgen. Bei der Bergung eines Portkatheterschlauches kam es zu einer Dislokation eines kleinen Schlauchanteiles in die distalen Anteile der A. pulmonalis. Dieser konnte nicht mehr geborgen werden. Ein Führungsdraht entlang der Aorta thoracalis/abdominalis imponierte schon mit der Gefäßwand verwachsen, eine Bergung war nicht mehr möglich. Periinterventionelle Komplikationen wurden nicht verzeichnet.

Schlussfolgerung: Die Bergung von intravasal disloziertem Fremdmaterial ist technisch erfolgreich und komplikationsarm durchzuführen. Eine operative Entfernung kann meist vermieden werden.

Deutscher Artikel/German Article

 
  • References

  • 1 McGee DC, Gould MK. Preventing complications of central venous catheterization. N Engl J Med 2003; 348: 1123-1133
  • 2 Merrer J, De Jonghe B, Golliot F et al. Complications of femoral and subclavian venous catheterization in critically ill patients: a randomized controlled trial. JAMA 2001; 286: 700-707
  • 3 Bernhardt LC, Wegner GP, Mendenhall JT. Intravenous catheter embolization to the pulmonary artery. Chest 1970; 57: 329-332
  • 4 Fisher RG, Ferreyro R. Evaluation of current techniques for nonsurgical removal of intravascular iatrogenic foreign bodies. Am J Roentgenol 1978; 130: 541-548
  • 5 Önal B, Coşkun B, Karabulut R et al. Interventional radiological retrieval of embolized vascular access device fragments. Diagn Interv Radiol 2012; 18: 87-91
  • 6 Bessoud B, de Baere T, Kuoch V et al. Experience at a single institution with endovascular treatment of mechanical complications caused by implanted central venous access devices in pediatric and adult patients. Am J Roentgenol 2003; 180: 527-532
  • 7 Turner DD, Sommers SC. Accidental passage of a polyethylene catheter from cubital vein to right atrium; report of a fatal case. N Engl J Med 1954; 251: 744-745
  • 8 Ballarini C, Intra M, Pisani Ceretti A et al. Complications of subcutaneous infusion port in the general oncology population. Oncology 1999; 56: 97-102
  • 9 Biffi R, de Braud F, Orsi F et al. Totally implantable central venous access ports for long-term chemotherapy. A prospective study analyzing complications and costs of 333 devices with a minimum follow-up of 180 days. Ann Oncol 1998; 9: 767-773
  • 10 Kock HJ, Pietsch M, Krause U et al. Implantable vascular access systems: experience in 1500 patients with totally implanted central venous port systems. World J Surg 1998; 22: 12-16
  • 11 Cheng CC, Tsai TN, Yang CC et al. Percutaneous retrieval of dislodged totally implantable central venous access system in 92 cases: experience in a single hospital. Eur J Radiol 2009; 69: 346-350
  • 12 Wang PC, Liang HL, Wu TH et al. Percutaneous retrieval of dislodged central venous port catheter: experience of 25 patients in a single institute. Acta Radiol 2009; 50: 15-20
  • 13 Vandoni RE, Guerra A, Sanna P et al. Randomised comparison of complications from three different permanent central venous access systems. Swiss Med Wkly 2009; 139: 313-316
  • 14 Gabelmann A, Kramer S, Gorich J. Percutaneous retrieval of lost or misplaced intravascular objects. Am J Roentgenol 2001; 176: 1509-1513
  • 15 Boysan E, Cicek OF, Cicek MC et al. Surgical removal of an atrial septal occluder device embolized to the main pulmonary artery. Tex Heart Inst J 2014; 41: 91-93
  • 16 Motta Leal Filho JM, Carnevale FC, Nasser F et al. Endovascular techniques and procedures, methods for removal of intravascular foreign bodies. Rev Bras Cir Cardiovasc 2010; 25: 202-208
  • 17 Egglin TK, Dickey KW, Rosenblatt M et al. Retrieval of intravascular foreign bodies: experience in 32 cases. Am J Roentgenol 1995; 164: 1259-1264
  • 18 Yang FS, Ohta I, Chiang HJ et al. Non-surgical retrieval of intravascular foreign body: experience of 12 cases. Eur J Radiol 1994; 18: 1-5
  • 19 Liu JC, Tseng HS, Chen CY et al. Percutaneous retrieval of intravascular foreign bodies: experience with 19 cases. Kaohsiung J Med Sci 2002; 18: 492-499
  • 20 Mallmann CV, Wolf KJ, Wacker FK. Retrieval of vascular foreign bodies using a self-made wire snare. Acta Radiol 2008; 49: 1124-1128
  • 21 Guimaraes M, Denton CE, Uflacker R et al. Percutaneous retrieval of an Amplatzer septal occluder device that had migrated to the aortic arch. Cardiovasc Intervent Radiol 2012; 35: 430-433
  • 22 Ferrero E, Ferri M, Viazzo A et al. Migration of an AMPLATZER atrial septal occluder to the abdominal aorta. Am J Cardiol 2013; 112: 612-613
  • 23 Slonim SM, Dake MD, Razavi MK et al. Management of misplaced or migrated endovascular stents. J Vasc Interv Radiol 1999; 10: 851-859
  • 24 Massmann A, Stroeder J, Schuerholz H et al. Percutaneous complication management for iliaco-aortal stent displacement. Vasa 2014; 43: 293-297