Evaluation of Low-kVp Low-Volume Iodinated Contrast Protocol for Coronary CT Angiography Using Retrospective ECG Gating

Abstract
Objective The aim of this study was to evaluate the use of low peak kilovoltage (kVp) low-volume iodinated contrast protocol for performing coronary computed tomography (CT) angiography (CCTA) in patients using retrospective electrocardiogram (ECG) gating.
Materials and Methods Hundred prospective patients undergoing CCTA were studied in two groups, A and B, using 70 kilovoltage (kV) and 120 kV protocols with half and standard intravenous volumes of injected iodinated contrast, respectively. All patients had heart rates less than 100 beats/min and body mass index (BMI) less than 31 kg/m2. Both the groups were evaluated for signal-to-noise (S/N) and contrast-to-noise (C/N) ratios along with radiation dose delivered in millisievert (mSv), and for image quality (IQ), on per patient and per segment basis.
Results Patients with group A showed statistically reduced radiation dose of 1.86 mSv compared with 6.86 mSv in group B patients. Marked reduction in image noise with statistically improved S/N and C/N ratios in all coronary vessels was seen in group A. S/N ratios in group A were 20.25, 18.68, 19.04, 17.41, and 18.69 for aorta, left main, left anterior descending, right coronary, and left circumflex arteries while they were 13.34, 11.12, 10.96, 9.74, and 8.67 in group B patients. C/N ratios were also higher in all vessels in group A patients, that is, 19.48, 19.48, 19.04, 19.48, and 17.68, compared with group B patients, who had 12.43, 10.03, 9.23, 9.57, and 8.23 ratios (p < 0.0001). No significant difference in IQ per patient and per vessel was seen between both the groups.
Discussion Retrospective ECG-gated low-kVp low-volume iodinated contrast protocol provides good diagnostic quality angiograms in patients with BMI up to 31 kg/m2 and with heart rates of less than 100 beats/min with three times reduced radiation dose. The reduced volume of contrast reduces the cost as well as the chance of contrast-induced nephropathy.
Declaration of Patient Consent
The authors certify that they have obtained all appropriate patient consent forms. In the consent forms, the patients have given their consent for their images and other clinical information to be reported in the journal. The patients understand that their names and initials will not be published and due efforts will be made to conceal their identity, but anonymity cannot be guaranteed.
Publication History
Article published online:
10 January 2022
© 2022. Indian Radiological Association. This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/)
Thieme Medical and Scientific Publishers Pvt. Ltd.
A-12, 2nd Floor, Sector 2, Noida-201301 UP, India
-
References
- 1 Maron DJ, Hochman JS, Reynolds HR. et al; ISCHEMIA Research Group. Initial invasive or conservative strategy for stable coronary disease. N Engl J Med 2020; 382 (15) 1395-1407
- 2 Hausleiter J, Meyer T, Hadamitzky M. et al. Radiation dose estimates from cardiac multislice computed tomography in daily practice: impact of different scanning protocols on effective dose estimates. Circulation 2006; 113 (10) 1305-1310
- 3 von Ballmoos MW, Haring B, Juillerat P, Alkadhi H. Meta-analysis: diagnostic performance of low-radiation-dose coronary computed tomography angiography. Ann Intern Med 2011; 154 (06) 413-420
- 4 Mitchell AM, Jones AE, Tumlin JA, Kline JA. Prospective study of the incidence of contrast-induced nephropathy among patients evaluated for pulmonary embolism by contrast-enhanced computed tomography. Acad Emerg Med 2012; 19 (06) 618-625
- 5 Morsbach F, Desbiolles L, Plass A. et al. Stenosis quantification in coronary CT angiography: impact of an integrated circuit detector with iterative reconstruction. Invest Radiol 2013; 48 (01) 32-40
- 6 Komatsu S, Kamata T, Imai A. et al. Coronary computed tomography angiography using ultra-low-dose contrast media: radiation dose and image quality. Int J Cardiovasc Imaging 2013; 29 (06) 1335-1340
- 7 Zhang LJ, Qi L, Wang J. et al. Feasibility of prospectively ECG-triggered high-pitch coronary CT angiography with 30 mL iodinated contrast agent at 70 kVp: initial experience. Eur Radiol 2014; 24 (07) 1537-1546
- 8 Den Harder AM, Willemink MJ, De Ruiter QM. et al. Dose reduction with iterative reconstruction for coronary CT angiography: a systematic review and meta-analysis. Br J Radiol 2016; 89 (1058): 20150068
- 9 McCollough CH, Primak AN, Braun N, Kofler J, Yu L, Christner J. Strategies for reducing radiation dose in CT. Radiol Clin North Am 2009; 47 (01) 27-40
- 10 Murakami Y, Kakeda S, Kamada K. et al. Effect of tube voltage on image quality in 64-section multidetector 3D CT angiography: evaluation with a vascular phantom with superimposed bone skull structures. AJNR Am J Neuroradiol 2010; 31 (04) 620-625
- 11 Kok M, Mihl C, Hendriks BM. et al. Optimizing contrast media application in coronary CT angiography at lower tube voltage: evaluation in a circulation phantom and sixty patients. Eur J Radiol 2016; 85 (06) 1068-1074
- 12 Lell MM, Jost G, Korporaal JG. et al. Optimizing contrast media injection protocols in state-of-the art computed tomographic angiography. Invest Radiol 2015; 50 (03) 161-167
- 13 Husmann L, Valenta I, Gaemperli O. et al. Feasibility of low-dose coronary CT angiography: first experience with prospective ECG-gating. Eur Heart J 2008; 29 (02) 191-197
- 14 Zhang LJ, Qi L, De Cecco CN. et al. High-pitch coronary CT angiography at 70 kVp with low contrast medium volume: comparison of 80 and 100 kVp high-pitch protocols. Medicine (Baltimore) 2014; 93 (22) e92
- 15 Cao JX, Wang YM, Lu JG, Zhang Y, Wang P, Yang C. Radiation and contrast agent doses reductions by using 80-kV tube voltage in coronary computed tomographic angiography: a comparative study. Eur J Radiol 2014; 83 (02) 309-314
- 16 Gordic S, Husarik DB, Desbiolles L, Leschka S, Frauenfelder T, Alkadhi H. High-pitch coronary CT angiography with third generation dual-source CT: limits of heart rate. Int J Cardiovasc Imaging 2014; 30 (06) 1173-1179