CC BY-NC-ND 4.0 · Eur J Dent 2021; 15(04): 618-623
DOI: 10.1055/s-0041-1725575
Original Article

Relation Between Insertion Torque and Implant Stability Quotient: A Clinical Study

João Paulo do Vale Souza
1   Department of Dental Materials and Prosthodontics, Araçatuba Dental School, São Paulo State University, Araçatuba, São Paulo, Brazil
,
Clóvis Lamartine de Moraes Melo Neto
1   Department of Dental Materials and Prosthodontics, Araçatuba Dental School, São Paulo State University, Araçatuba, São Paulo, Brazil
,
Lucas Tavares Piacenza
1   Department of Dental Materials and Prosthodontics, Araçatuba Dental School, São Paulo State University, Araçatuba, São Paulo, Brazil
,
Emily Vivianne Freitas da Silva
1   Department of Dental Materials and Prosthodontics, Araçatuba Dental School, São Paulo State University, Araçatuba, São Paulo, Brazil
,
André Luiz de Melo Moreno
1   Department of Dental Materials and Prosthodontics, Araçatuba Dental School, São Paulo State University, Araçatuba, São Paulo, Brazil
,
Paulo Augusto Penitente
1   Department of Dental Materials and Prosthodontics, Araçatuba Dental School, São Paulo State University, Araçatuba, São Paulo, Brazil
,
Juliana Lujan Brunetto
1   Department of Dental Materials and Prosthodontics, Araçatuba Dental School, São Paulo State University, Araçatuba, São Paulo, Brazil
,
Daniela Micheline dos Santos
1   Department of Dental Materials and Prosthodontics, Araçatuba Dental School, São Paulo State University, Araçatuba, São Paulo, Brazil
2   Oral Oncology Center, Araçatuba Dental School, São Paulo State University, Araçatuba, São Paulo, Brazil
,
Marcelo Coelho Goiato
1   Department of Dental Materials and Prosthodontics, Araçatuba Dental School, São Paulo State University, Araçatuba, São Paulo, Brazil
2   Oral Oncology Center, Araçatuba Dental School, São Paulo State University, Araçatuba, São Paulo, Brazil
› Author Affiliations
Funding None.

Abstract

Objectives This study aimed to assess the relation between the insertion torque and implant stability quotient (ISQ recorded immediately and 6 months after implant placement).

Materials and Methods Twenty-five patients over the age of 18 years were selected for this study. One implant was placed per patient after tooth extraction. All implants had the same size (11.5 × 3.75 mm) and brand (Hexagonal Morse cone, DSP Biomedical). The insertion torque (Ncm) and resonance frequency analysis (ISQ value) (Osstell Mentor) were used to assess the primary stability (on the day of surgery). After 6 months, resonance frequency analysis was used to assess the secondary stability of each implant.

Statistical Analysis The insertion torque data were correlated with ISQ measurements by using Pearson’s correlation. The significance level was 5%.

Results There was a positive correlation between insertion torque and initial ISQ (correlation: 0.457; p = 0.022); however, no correlation was found between insertion torque and final ISQ (p = 0.308).

Conclusion The present study demonstrated that there is a positive correlation between the insertion torque and the initial ISQ. Therefore, the higher the insertion torque, the higher the initial ISQ (and vice versa).



Publication History

Article published online:
07 July 2021

© 2021. European Journal of Dentistry. This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial-License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/).

Thieme Medical and Scientific Publishers Pvt. Ltd.
A-12, 2nd Floor, Sector 2, Noida-201301 UP, India

 
  • References

  • 1 da Cunha HA, Francischone CE, Filho HN, de Oliveira RC. A comparison between cutting torque and resonance frequency in the assessment of primary stability and final torque capacity of standard and TiUnite single-tooth implants under immediate loading. Int J Oral Maxillofac Implants 2004; 19 (04) 578-585
  • 2 Esposito M, Siormpas K, Mitsias M, Bechara S, Trullenque-Eriksson A, Pistilli R. Immediate, early (6 weeks) and delayed loading (3months) of single implants: 4-month post-loading from a multicenter pragmatic randomised controlled trial. Eur J Oral Implantology 2016; 9 (03) 249-260
  • 3 Pigozzo MN, Rebelo da Costa T, Sesma N, Laganá DC. Immediate versus early loading of single dental implants: A systematic review and meta-analysis. J Prosthet Dent 2018; 120 (01) 25-34
  • 4 Tettamanti L, Andrisani C, Bassi MA, Vinci R, Silvestre-Rangil J, Tagliabue A. Immediate loading implants: review of the critical aspects. Oral Implantol (Rome) 2017; 10 (02) 129-139
  • 5 Grognard N, Verleye G, Mavreas D, Vande-Vannet B. Updated secondary implant stability data of two dental implant systems. A retrospective cohort study. J Clin Exp Dent 2017; 9 (09) e1121-e1128
  • 6 Mathieu V, Vayron R, Richard G. et al Biomechanical determinants of the stability of dental implants: influence of the bone-implant interface properties. J Biomech 2014; 47 (01) 3-13
  • 7 Baldi D, Lombardi T, Colombo J. et al Correlation between insertion torque and implant stability quotient in tapered implants with knife-edge thread design. BioMed Res Int 2018; 2018: 7201093
  • 8 Degidi M, Daprile G, Piattelli A. Primary stability determination by means of insertion torque and RFA in a sample of 4,135 implants. Clin Implant Dent Relat Res 2012; 14 (04) 501-507
  • 9 Kahraman S, Bal BT, Asar NV, Turkyilmaz I, Tözüm TF. Clinical study on the insertion torque and wireless resonance frequency analysis in the assessment of torque capacity and stability of self-tapping dental implants. J Oral Rehabil 2009; 36 (10) 755-761
  • 10 Park KJ, Kwon JY, Kim SK. et al The relationship between implant stability quotient values and implant insertion variables: a clinical study. J Oral Rehabil 2012; 39 (02) 151-159
  • 11 Sarfaraz H, Johri S, Sucheta P, Rao S. Study to assess the relationship between insertion torque value and implant stability quotient and its influence on timing of functional implant loading. J Indian Prosthodont Soc 2018; 18 (02) 139-146
  • 12 Johansson P, Strid K. Assessment of bone quality from cutting resistance during implant surgery. Int J Oral Maxillofac Implants 1994; 9: 279-288
  • 13 Friberg B, Sennerby L, Meredith N, Lekholm U. A comparison between cutting torque and resonance frequency measurements of maxillary implants. A 20-month clinical study. Int J Oral Maxillofac Surg 1999; 28 (04) 297-303
  • 14 Friberg B, Sennerby L, Roos J, Johansson P, Strid CG, Lekholm U. Evaluation of bone density using cutting resistance measurements and microradiography: an in vitro study in pig ribs. Clin Oral Implants Res 1995; 6 (03) 164-171
  • 15 Kuchler U, Chappuis V, Bornstein MM. et al Development of Implant Stability Quotient values of implants placed with simultaneous sinus floor elevation - results of a prospective study with 109 implants. Clin Oral Implants Res 2017; 28 (01) 109-115
  • 16 Meredith N, Alleyne D, Cawley P. Quantitative determination of the stability of the implant-tissue interface using resonance frequency analysis. Clin Oral Implants Res 1996; 7 (03) 261-267
  • 17 Herrero-Climent M, Santos-García R, Jaramillo-Santos R. et al Assessment of Osstell ISQ’s reliability for implant stability measurement: a cross-sectional clinical study. Med Oral Patol Oral Cir Bucal 2013; 18 (06) e877-e882
  • 18 Santamaría-Arrieta G, Brizuela-Velasco A, Fernández-González FJ. et al Biomechanical evaluation of oversized drilling technique on primary implant stability measured by insertion torque and resonance frequency analysis. J Clin Exp Dent 2016; 8 (03) e307-e311
  • 19 Açil Y, Sievers J, Gülses A, Ayna M, Wiltfang J, Terheyden H. Correlation between resonance frequency, insertion torque and bone-implant contact in self-cutting threaded implants. Odontology 2017; 105 (03) 347-353
  • 20 Melo SL, Bortoluzzi EA, Abreu Jr M, Corrêa LR, Corrêa M. Diagnostic ability of a cone-beam computed tomography scan to assess longitudinal root fractures in prosthetically treated teeth. J Endod 2010; 36 (11) 1879-82
  • 21 Ataullah K, Chee LF, Peng LL, Tho CY, Wei WCS, Baig MR. Implant placement in extraction sockets: a short review of the literature and presentation of a series of three cases. J Oral Implantol 2008; 34 (02) 97-106
  • 22 Daabiss M. American Society of Anaesthesiologists physical status classification. Indian J Anaesth 2011; 55 (02) 111-115
  • 23 Natale Júnior V, Souza FA, Vedovatto E, Nishioka RS, Poli PP, Carvalho PS. Preservation of dental sockets filled with composite bovine bone. a single-blind randomized clinical trial. Braz Dent J 2018; 29 (06) 583-591
  • 24 Lozano-Carrascal N, Salomó-Coll O, Gilabert-Cerdà M, Farré-Pagés N, Gargallo-Albiol J, Hernández-Alfaro F. Effect of implant macro-design on primary stability: a prospective clinical study. Med Oral Patol Oral Cir Bucal 2016; 21 (02) e214-e221
  • 25 Becker W, Hujoel P, Becker BE. Resonance frequency analysis: comparing two clinical instruments. Clin Implant Dent Relat Res 2018; 20 (03) 308-312
  • 26 Chugh T, Jain AK, Jaiswal RK, Mehrotra P, Mehrotra R. Bone density and its importance in orthodontics. J Oral Biol Craniofac Res 2013; 3 (02) 92-97
  • 27 Turkyilmaz I, McGlumphy EA. Influence of bone density on implant stability parameters and implant success: a retrospective clinical study. BMC Oral Health 2008; 8: 32
  • 28 Misch CE. Dental Implant Prosthetics. 2nd Ed. St. Louis, Mosby: Elsevier 2014: 237-252
  • 29 Albrektsson T, Zarb GA. Determinants of correct clinical reporting. Int J Prosthodont 1998; 11 (05) 517-521
  • 30 The evidence-based Osstell ISQ Scale - Clinical Guidelines. Available at: https://www.osstell.com/clinical-guidelines/. Ac- cessed January 10, 2021
  • 31 Trisi P, Berardi D, Paolantonio M, Spoto G, D’Addona A, Perfetti G. Primary stability, insertion torque, and bone density of conical implants with internal hexagon: is there a relationship?. J Craniofac Surg 2013; 24 (03) 841-844
  • 32 Trisi P, Berardi D, Paolantonio M, Spoto G, D’Addona A, Perfetti G.. Primary stability, insertion torque, and bone density of conical implants with internal hexagon: is there a relationship?. J Craniofac Surg 2013; 24 (03) 841-844