CC BY-NC-ND 4.0 · Rev Bras Ortop (Sao Paulo) 2019; 54(03): 295-302
DOI: 10.1055/s-0039-1691762
Artigo Original | Original Article
Sociedade Brasileira de Ortopedia e Traumatologia. Published by Thieme Revnter Publicações Ltda Rio de Janeiro, Brazil

Comparison of Wear Rate between Ceramic-on-Ceramic, Metal on Highly Cross-linked Polyethylene, and Metal-on-Metal Bearings[]

Article in several languages: português | English
1   Departamento de Cirurgia Ortopédica, Nagoya University Graduate School of Medicine, Tsurumai-cho, Showa-ku, Nagoya, Japão
,
Taisuke Seki
1   Departamento de Cirurgia Ortopédica, Nagoya University Graduate School of Medicine, Tsurumai-cho, Showa-ku, Nagoya, Japão
,
Daigo Morita
1   Departamento de Cirurgia Ortopédica, Nagoya University Graduate School of Medicine, Tsurumai-cho, Showa-ku, Nagoya, Japão
,
Daigo Komatsu
1   Departamento de Cirurgia Ortopédica, Nagoya University Graduate School of Medicine, Tsurumai-cho, Showa-ku, Nagoya, Japão
,
Yasuhiko Takegami
1   Departamento de Cirurgia Ortopédica, Nagoya University Graduate School of Medicine, Tsurumai-cho, Showa-ku, Nagoya, Japão
,
Naoki Ishiguro
1   Departamento de Cirurgia Ortopédica, Nagoya University Graduate School of Medicine, Tsurumai-cho, Showa-ku, Nagoya, Japão
› Author Affiliations
Further Information

Publication History

19 January 2018

10 July 2018

Publication Date:
27 June 2019 (online)

Abstract

Objective Currently, there is a lack of evidence on the medium-term follow-up of cementless total hip arthroplasty (THA) using metal on highly cross-linked polyethylene (MoP), ceramic-on-ceramic (CoC), and metal-on-metal (MoM) bearings. Our aim was to calculate the 5- to 10-year wear rate and the incidence rate of osteolysis for 3 types of bearings.

Methods A total of 77 patients underwent MoP, 105 underwent ceramic CoC, and 55 underwent MoM THAs. The average patient age at the time of surgery was 64.7, 55.9, and 59.9 years old in the MoP, CoC, and MoM bearings, respectively. Clinical and radiologic measurements at a mean follow-up of 7.6 years were analyzed.

Results The mean postoperative Harris hip scores showed no difference among the groups. The mean annual liner wear rates were 0.0160, 0.0040 and 0.0054 mm/year in MoP, CoC bearings, and MoM bearings, with that of CoC bearings being significantly lower than the others. Osteolysis (14.5%) among MoM bearings was significantly more frequently observed compared with the others. Kaplan-Meier survival at 10 years with implant loosening, or revision THA as the endpoint, was 96.1% (95% confidence interval [CI]: 88.4–98.7) for MoP, 98.6% (95% CI: 90.3–98.6) for CoC bearing, and 98.2% (95% CI: 88.0–99.7) for MoM bearings (p = 0.360).

Conclusion Excellent clinical and radiological outcomes were obtained for MoP and CoC bearings.

The present study was conducted at the Department of Orthopedic Surgery of the Nagoya University Graduate School of Medicine, Tsurumai-cho, Showa-ku Nagoya, Japan.


 
  • Referências

  • 1 D'Antonio JA, Capello WN, Naughton M. Ceramic bearings for total hip arthroplasty have high survivorship at 10 years. Clin Orthop Relat Res 2012; 470 (02) 373-381
  • 2 Shetty V, Shitole B, Shetty G, Thakur H, Bhandari M. Optimal bearing surfaces for total hip replacement in the young patient: a meta-analysis. Int Orthop 2011; 35 (09) 1281-1287
  • 3 Callary SA, Solomon LB, Holubowycz OT, Campbell DG, Munn Z, Howie DW. Wear of highly crosslinked polyethylene acetabular components. Acta Orthop 2015; 86 (02) 159-168
  • 4 D'Antonio J, Capello W, Manley M, Bierbaum B. New experience with alumina-on-alumina ceramic bearings for total hip arthroplasty. J Arthroplasty 2002; 17 (04) 390-397
  • 5 Engh Jr CA, Stepniewski AS, Ginn SD, Beykirch SE, Sychterz-Terefenko CJ, Hopper RH. , et al. A randomized prospective evaluation of outcomes after total hip arthroplasty using cross-linked marathon and non-cross-linked Enduron polyethylene liners. J Arthroplasty 2006; 21 (06) (Suppl. 02) 17-25
  • 6 Nikolaou VS, Edwards MR, Bogoch E, Schemitsch EH, Waddell JP. A prospective randomised controlled trial comparing three alternative bearing surfaces in primary total hip replacement. J Bone Joint Surg Br 2012; 94 (04) 459-465
  • 7 Takao M, Ohzono K, Nishii T, Miki H, Nakamura N, Sugano N. Cementless modular total hip arthroplasty with subtrochanteric shortening osteotomy for hips with developmental dysplasia. J Bone Joint Surg Am 2011; 93 (06) 548-555
  • 8 Milošev I, Kovač S, Trebše R, Levašič V, Pišot V. Comparison of ten-year survivorship of hip prostheses with use of conventional polyethylene, metal-on-metal, or ceramic-on-ceramic bearings. J Bone Joint Surg Am 2012; 94 (19) 1756-1763
  • 9 Wyles CC, Jimenez-Almonte JH, Murad MH, Norambuena-Morales GA, Cabanela ME, Sierra RJ. , et al. There Are No Differences in Short- to Mid-term Survivorship Among Total Hip-bearing Surface Options: A Network Meta-analysis. Clin Orthop Relat Res 2015; 473 (06) 2031-2041
  • 10 Bascarevic Z, Vukasinovic Z, Slavkovic N, Dulic B, Trajkovic G, Bascarevic V. , et al. Alumina-on-alumina ceramic versus metal-on-highly cross-linked polyethylene bearings in total hip arthroplasty: a comparative study. Int Orthop 2010; 34 (08) 1129-1135
  • 11 Higuchi Y, Hasegawa Y, Seki T, Komatsu D, Ishiguro N. Significantly Lower Wear of Ceramic-on-Ceramic Bearings Than Metal-on-Highly Cross-Linked Polyethylene Bearings: A 10- to 14-Year Follow-Up Study. J Arthroplasty 2016; 31 (06) 1246-1250
  • 12 Kanoh T, Hasegawa Y, Masui T, Yamaguchi J, Kawabe K, Ishiguro N. Accurate acetabular component orientation after total hip arthroplasty using an acetabular alignment guide. J Arthroplasty 2010; 25 (01) 81-86
  • 13 Mestriner MB, Verquietini CM, Waisberg G, Akkari M, Fukunaga ET, Santili C. Radiographic evaluation in epiphysiolysis: possible predictors of bilaterality?. Acta Ortop Bras 2012; 20 (04) 203-206
  • 14 Kim YH, Kim JS, Oh SH, Kim JM. Comparison of porous-coated titanium femoral stems with and without hydroxyapatite coating. J Bone Joint Surg Am 2003; 85-A (09) 1682-1688
  • 15 Sutherland CJ, Wilde AH, Borden LS, Marks KE. A ten-year follow-up of one hundred consecutive Müller curved-stem total hip-replacement arthroplasties. J Bone Joint Surg Am 1982; 64 (07) 970-982
  • 16 Lee JH, Lee BW, Lee BJ, Kim SY. Midterm results of primary total hip arthroplasty using highly cross-linked polyethylene: minimum 7-year follow-up study. J Arthroplasty 2011; 26 (07) 1014-1019
  • 17 Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR. Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am 1978; 60 (02) 217-220
  • 18 Barrack RL, Lavernia C, Szuszczewicz ES, Sawhney J. Radiographic wear measurements in a cementless metal-backed modular cobalt-chromium acetabular component. J Arthroplasty 2001; 16 (07) 820-828
  • 19 García-Rey E, García-Cimbrelo E, Cruz-Pardos A, Ortega-Chamarro J. New polyethylenes in total hip replacement: a prospective, comparative clinical study of two types of liner. J Bone Joint Surg Br 2008; 90 (02) 149-153
  • 20 Brooker AF, Bowerman JW, Robinson RA, Riley Jr LH. Ectopic ossification following total hip replacement. Incidence and a method of classification. J Bone Joint Surg Am 1973; 55 (08) 1629-1632
  • 21 Cai P, Hu Y, Xie J. Large-diameter Delta ceramic-on-ceramic versus common-sized ceramic-on-polyethylene bearings in THA. Orthopedics 2012; 35 (09) e1307-e1313
  • 22 Gallo J, Goodman SB, Lostak J, Janout M. Advantages and disadvantages of ceramic on ceramic total hip arthroplasty: a review. Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub 2012; 156 (03) 204-212
  • 23 Lombardi Jr AV, Mallory TH, Cuckler JM, Williams J, Berend KR, Smith TM. Mid-term results of a polyethylene-free metal-on-metal articulation. J Arthroplasty 2004; 19 (07) (Suppl. 02) 42-47
  • 24 Yin S, Zhang D, Du H, Du H, Yin Z, Qiu Y. Is there any difference in survivorship of total hip arthroplasty with different bearing surfaces? A systematic review and network meta-analysis. Int J Clin Exp Med 2015; 8 (11) 21871-21885
  • 25 Greiner JJ, Callaghan JJ, Bedard NA, Liu SS, Goetz DD, Mahoney CR. Metal-on-Metal Total Hip Arthroplasty at Five to Twelve Years Follow-Up: A Concise Follow-Up of a Previous Report. J Arthroplasty 2016; 31 (08) 1773-1778
  • 26 Capello WN, D'Antonio JA, Ramakrishnan R, Naughton M. Continued improved wear with an annealed highly cross-linked polyethylene. Clin Orthop Relat Res 2011; 469 (03) 825-830
  • 27 Reynolds SE, Malkani AL, Ramakrishnan R, Yakkanti MR. Wear analysis of first-generation highly cross-linked polyethylene in primary total hip arthroplasty: an average 9-year follow-up. J Arthroplasty 2012; 27 (06) 1064-1068
  • 28 Epinette J-A, Manley MT. No differences found in bearing related hip survivorship at 10-12 years follow-up between patients with ceramic on highly cross-linked polyethylene bearings compared to patients with ceramic on ceramic bearings. J Arthroplasty 2014; 29 (07) 1369-1372
  • 29 Al-Hajjar M, Fisher J, Tipper JL, Williams S, Jennings LM. Wear of 36-mm BIOLOX(R) delta ceramic-on-ceramic bearing in total hip replacements under edge loading conditions. Proc Inst Mech Eng H 2013; 227 (05) 535-542