CC BY-NC-ND 4.0 · J Neuroanaesth Crit Care 2020; 07(02): 104-106
DOI: 10.1055/s-0039-1679133
Case Report

A Persistently High End-Tidal Carbon Dioxide Value: Can This Be Spurious?

Barkha Bindu
1   Department of Neuroanaesthesiology and Critical Care, Neurosciences Centre, All India Institute of Medical Sciences, New Delhi, India
,
Gyaninder P. Singh
1   Department of Neuroanaesthesiology and Critical Care, Neurosciences Centre, All India Institute of Medical Sciences, New Delhi, India
,
Varun Jain
1   Department of Neuroanaesthesiology and Critical Care, Neurosciences Centre, All India Institute of Medical Sciences, New Delhi, India
,
Arvind Chaturvedi
1   Department of Neuroanaesthesiology and Critical Care, Neurosciences Centre, All India Institute of Medical Sciences, New Delhi, India
› Author Affiliations

Abstract

End-tidal carbon dioxide (EtCO2) monitoring has now become the standard of care not only during anesthesia but also in intensive care units for patients on mechanical ventilation, emergency department, and pre-hospital settings to confirm and monitor the correct placement of endotracheal tube. It is a non-invasive and continuous method of measuring exhaled carbon dioxide (CO2). Continuous waveform capnography measures EtCO2 and monitors ventilation. EtCO2 often correlates with partial pressure of carbon dioxide in arterial blood (PaCO2) and is a reliable indicator of PaCO2. A rise in EtCO2 often implies increased production of CO2 or decreased excretion (rebreathing, decrease ventilation) of CO2. We report an unusual case where the monitor malfunction per se lead to spuriously increased EtCO2 values without any clinical cause and did not correlate with PaCO2, thereby re-emphasizing that various monitors must always be interpreted in correlation with clinical observation.



Publication History

Article published online:
01 May 2019

© .

Thieme Medical and Scientific Publishers Private Ltd.
A-12, Second Floor, Sector -2, NOIDA -201301, India

 
  • References

  • 1 Donald MJ, Paterson B. End tidal carbon dioxide monitoring in prehospital and retrieval medicine: a review. Emerg Med J 2006; 23 (09) 728-730
  • 2 Razi E, Moosavi GA, Omidi K, Khakpour SaebiA, Razi A. Correlation of end-tidal carbon dioxide with arterial carbon dioxide in mechanically ventilated patients. Arch Trauma Res 2012; 1 (02) 58-62
  • 3 Russell GB, Graybeal JM. The arterial to end-tidal carbon dioxide difference in neurosurgical patients during craniotomy. Anesth Analg 1995; 81 (04) 806-810
  • 4 Vinay B, Sriganesh K, Gopala KrishnaKN. An abrupt reduction in end-tidal carbon-dioxide during neurosurgery is not always due to venous air embolism: a capnograph artefact. J Clin Monit Comput 2014; 28 (02) 217-219
  • 5 Breen PH, Mazumdar B, Skinner SC. Capnometer transport delay: measurement and clinical implications. Anesth Analg 1994; 78 (03) 584-586
  • 6 Soumya M, Vimala S, Arulvelan A. Acute reduction in the end-tidal carbon dioxide level during neurosurgery: another cause for capnography artefact. J Neurosurg Anesthesiol 2017; 29 (03) 365-366
  • 7 Vinay B. An abrupt reduction in end tidal carbon di oxide concentration in a mechanically ventilated patient in neurocritical care ward: a capnogram artifact. J Clin Monit Comput 2018; 32 (02) 369-371
  • 8 Tautz TJ, Urwyler A, Antognini JF, Riou B. Case scenario: increased end-tidal carbon dioxide: a diagnostic dilemma. Anesthesiology 2010; 112 (02) 440-446
  • 9 Lee C, Lee KC, Kim HY. et al. Unidirectional valve malfunction by the breakage or malposition of disc - two cases report. Korean J Anesthesiol 2013; 65 (04) 337-340
  • 10 Kumar A, Bithal P, Chouhan RS, Sinha PK. Should one rely on capnometry when a capnogram is not seen?. J Neurosurg Anesthesiol 2002; 14 (02) 153-156
  • 11 McSwain SD, Hamel DS, Smith PB. et al. End-tidal and arterial carbon dioxide measurements correlate across all levels of physiologic dead space. Respir Care 2010; 55 (03) 288-293