Thromb Haemost 1981; 45(02): 162-168
DOI: 10.1055/s-0038-1650156
Original Article
Schattauer GmbH Stuttgart

The Diagnosis of Mild Haemophilia by the Partial Thromboplastin Time Test. WFH/ICTH Study of the Manchester Method [*]

P F O'Brien
The Haematology Dept., St. Thomas' Hospital and Medical School, London
The Epidemiology and Medical Care Unit, Northwick Park Hospital, Harrow, U. K.
,
W R S North
The Haematology Dept., St. Thomas' Hospital and Medical School, London
The Epidemiology and Medical Care Unit, Northwick Park Hospital, Harrow, U. K.
,
G I C Ingram
The Haematology Dept., St. Thomas' Hospital and Medical School, London
The Epidemiology and Medical Care Unit, Northwick Park Hospital, Harrow, U. K.
› Author Affiliations
Further Information

Publication History

Received 28 August 1980

Accepted after revision 09 February 1981

Publication Date:
05 July 2018 (online)

Summary

In two trials, various methods of performing the partial thromboplastin time test (PTT) were compared with the method described from Manchester by Poller and Thomson (1) and Poller et al. (2), for their ability to diagnose mild haemophilia. In the first trial, 64 laboratories in various countries compared their own methods and reagents with the Manchester method on three coded haemophilic plasmas distributed freeze-dried and on four local fresh normals collected in each laboratory. In the second trial, seven well-standardized commercial methods were tested in pairs against the Manchester method in seven experienced laboratories on 22 mild haemophilic samples and six normal samples each coded and distributed frozen. PTTs were related to factor VIII: C level.

In the first trial, the better standardized methods gave more consistent results than others; but even so there were large differences between laboratories even on normal plasmas. The second trial established a reasonably linear relationship between log PTT and log VIII :C in the range 15-150 iu/dl; but different methods did not greatly vary in their ability to detect a mild haemophilic defect; this was confirmed by calculating the index Slope/(Residual S. D.) which was used to provide a rank order. Scatter diagrams suggested that a proportion of mild haemophilic samples would be likely to have PTTs within the normal range by any method. It is concluded that the Manchester method does not detect mild haemophilia with greater discrimination than the other methods studied; that each laboratory must establish its own normal range even if using a well-standardized method; and that in screening patients a PTT must be supplemented by a discriminating history.

* A fuller analysis of the data is archived with the Secretariat of the World Federation of Haemophilia, 1170 Peel Street, Room 1126, Montreal, Que., Canada H3B 2T4, to whom enquiries should be adressed.


 
  • References

  • 1 Poller L, Thomson JM. The partial thromboplastin (Cephalin) time test. J Clin Pathol 1972; 25: 1038-1044
  • 2 Poller L, Thomson JM, Palmer MK. Measuring partial thromboplastin-time: an international collaborative study. Lancet. 1976; II 842-846
  • 3 Lehovich M. Comparative assessment of four commercial activated thromboplastins. Technical Report for A. R. T. Haematology Certification. Ontario; Canada: 1973
  • 4 Koepke JA. The partial thromboplastin time in the CAP survey program. Am J Clin Pathol 1975; 63: 990-995
  • 5 Goulian M, Beck WS. The partial thromboplastin time test. Modifications of the procedure, and study of the sensitivity and optimal conditions. Am J Clin Pathol 1965; 44: 97-103
  • 6 Matchett MO, Ingram GIC. Partial thromboplastin time with kaolin. J Clin Pathol 1965; 18: 465-471
  • 7 Sibley C, Singer JW, Wood RJ. Comparison of activated partial thromboplastin reagents. Am J Clin Pathol 1973; 59: 581-586
  • 8 Cox DR. Some systematic experimental designs. Biometrika 1951; 38: 312-327
  • 9 Ingram GIC, Hills M. The prothrombin time test. Effect of varying citrate concentration. Thromb Haemostas 1976; 36: 230-236
  • 10 Cawkwell RD. Patient’s age and the activated partial thromboplastin time. Thromb Haemostas 1978; 39: 780-781
  • 11 Nilsson IM, Mannucci PM, Refsum N, Gilhuus-Moe C. “Cephotest” in detection of haemophiliacs. Evaluation of a standardized Activated Partial Thromboplastin Time (APTT) test. Farmakoterapi 1976; 01 02: 18-26
  • 12 Refsum N, Abildgaard U. “Cephotest” in monitoring heparin therapy. Evaluation of a standardized Activated Partial Thromboplastin Time (APTT) test. Farmakoterapi 1976; 01 02: 27-35
  • 13 Babson AL, Babson SR. Comparative evaluation of a partial thromboplastin reagent containing a non-settling, particulate activator. Am J Clin Pathol 1974; 62: 856-860
  • 14 Hoffman J JML, Meulendijk PN. Comparison of reagents for determining the activated partial thromboplastin time. Thromb Haemostas 1978; 39: 640-645
  • 15 Koepke JA. College of American Pathologists partial thromboplastin time survey. Am J Clin Pathol 1976; 65: 126-127
  • 16 Janson T, Grimmer Ø. Test conditions and sensitivity of the Activated Partial Thromboplastin Time (APTT) test “Cephotest”. Farmakoterapi 1976; 01 02: 36-44
  • 17 Joist JH, Cowan JF, Khan M. Rapid loss of factor XII and XI activity in ellagic acid-activated normal plasma; role of plasma inhibitors and implications for automated activated partial thromboplastin time recording. J Lab Clin Med 1977; 90: 1054-1065
  • 18 Healy MJR, Ingram G IC. The “normal range” in tests of haemostasis. Thromb Haemostas 1978; 39: 504-509