Subscribe to RSS
DOI: 10.1055/s-0037-1615647
Prothrombin Time Ratio Is Reduced by Magnesium Contamination in Evacuated Blood Collection Tubes
Publication History
Received
11 September 2000
Accepted after revision
12 December 2000
Publication Date:
08 December 2017 (online)
Summary
Magnesium ions were detected in sodium citrate solutions in several lots of evacuated blood collection tubes. The mean concentrations ranged between 1.3 and 1.6 mmol/L. Magnesium was also present in the rubber stoppers of the blood collection tubes and could be leached into the citrate solution. It was shown that magnesium added to citrated plasma shortened the prothrombin time of both coumarin and normal plasma. The effect of magnesium was relatively greater on coumarin than on normal plasma resulting in reduced prothrombin time ratio. Shortening of the prothrombin time was also observed when magnesium chloride was added to dialysed plasma, i.e., in the absence of citrate. These results indicate that magnesium contamination can interfere with accurate INR determination in the control of oral anticoagulant therapy.
-
References
- 1 Duncan EM, Casey CR, Duncan BM, Lloyd JV. Effect of concentration of trisodium citrate anticoagulant on calculation of the international normalised ratio and the international sensitivity index of thromboplastin. Thromb Haemost 1994; 72: 84-88.
- 2 Adcock DM, Kressin DC, Marlar RA. Effect of 3.2% vs 3.8% sodium citrate concentration on routine coagulation testing. Am J Clin Pathol 1997; 107: 105-110.
- 3 Danielson CFM, Davis K, Jones G, Benson J, Arney K, Martin J. Effect of citrate concentration in specimen collection tubes on the international normalized ratio. Arch Pathol Lab Med 1997; 121: 956-959.
- 4 Chantarangkul V, Tripodi A, Clerici M, Negri B, Mannucci PM. Assessment of the influence of citrate concentration on the international normalized ratio (INR) determined with twelve reagent-instrument combinations. Thromb Haemost 1998; 80: 258-262.
- 5 Van den Besselaar AMHP, Meeuwisse-Braun J, Witteveen E, van Meegen E. Effect of evacuated blood collection tubes on thromboplastin calibration. Thromb Haemost 1998; 79: 1062-3.
- 6 Van den Besselaar AMHP, Bertina RM, van der Meer FJM, den Hartigh J. Different sensitivities of various thromboplastins to two blood collection systems for monitoring oral anticoagulant therapy. Thromb Haemost 1999; 82: 153-4.
- 7 D’Angelo G, Villa C. Measurement of prothrombin time in patients on oral anticoagulant therapy: effect of two different evacuated tubes. Haematologica 1999; 84: 656-657.
- 8 Van den Besselaar AMHP, Chantarangkul V, Tripodi A. A comparison of two sodium citrate concentrations in two evacuated blood collection systems for prothrombin time and ISI determination. Thromb Haemost 2000; 84: 664-667.
- 9 Palmer RN, Kessler CM, Gralnick HR. Warfarin anticoagulation: difficulties in interpretation of the prothrombin time. Thromb Res 1982; 25: 125-130.
- 10 Sekiya F, Yoshida M, Yamashita T, Morita T. Magnesium(II) is a crucial constituent of the blood coagulation cascade. J Biol Chem 1996; 271: 8541-8544.
- 11 W. H. O. Expert Committee on Biological Standardization.. Guidelines for thromboplastins and plasma used to control oral anticoagulant therapy. WHO techn rep series, no. 889 1999
- 12 Kennedy BW, Beal TS. Minerals leached into drinking water from rubber stoppers. Lab Anim Sci 1991; 41: 233-236.
- 13 Abou Shady EA, Farrag HE, el-Damarawy NA, Mohammed FA, Kamel AM, Massoud AA. In vitro effects of trace elements on blood clotting and platelet function. B—Zinc and magnesium. J Egypt Public Health Assoc 1991; 66: 49-72.
- 14 Butenas S, Lawson JH, Kalafatis M, Mann KG. Cooperative interaction of divalent metal ions, substrate, and tissue factor with factor VIIa. Biochemistry 1994; 33: 3449-56.
- 15 DuP Heyns A, van den Berg DJ, Kleynhans PHT, du Toit PW. Unsuitability of evacuated tubes for monitoring heparin therapy by activated partial thromboplastin time. J Clin Pathol 1981; 34: 63-68.