Performance of an Automatic Dose Control System for CT: Patient Studies
Praxistauglichkeit eines Systems zur automatischen Dosisregelung für CT: Untersuchungen an PatientenPublication History
16 April 2011
05 September 2012
Publication Date:
29 October 2012 (online)
Abstract
Purpose: To study the effect of an automatic dose control (ADC) system with adequate noise characteristic on the individual perception of image noise and diagnostic acceptance compared to objectively measured image noise and the dose reductions achieved in a representative group of patients.
Materials and Methods: In a retrospective study two matched cohorts of 20 patients each were identified: a manual cohort with exposure settings according to body size (small – regular – large) and an ADC cohort with exposure settings calculated by the ADC system (DoseRight 2.0™, Philips Healthcare). For each patient, 12 images from 6 defined anatomic levels from contrast-enhanced scans of chest and abdomen/pelvis were analyzed by 4 independent readers concerning image noise and diagnostic acceptance on a five-point Likert scale and evaluated for objectively measured image noise. Radiation exposure was calculated from recorded exposure data.
Results: Use of the ADC system reduced the average effective dose for patients by 36 % in chest scans (3.2 vs. 4.9 mSv) and by 17 % in abdomen/pelvis scans (7.6 vs. 8.3 mSv). Average objective noise was slightly lower in the manual cohort (11.1 vs. 12.8 HU), correlating with a slightly better rating in subjective noise score (4.4 vs. 4.2). However, diagnostic acceptance was rated almost equal in both cohorts with excellent image quality (4.6 vs. 4.5).
Conclusion: Use of an ADC system with adequate noise characteristic leads to significant reductions in radiation exposure for patients while maintaining excellent image quality.
Zusammenfassung
Ziel: Evaluation eines automatischen Dosisregelungs-Systems (ADC) mit adäquater Rauschcharakteristik bezüglich des individuellen Eindrucks zum Bildrauschen und zur diagnostischen Akzeptanz im Vergleich zu objektiv gemessenem Bildrauschen und der in einer repräsentativen Patientengruppe erreichbaren Dosisreduktion.
Material und Methoden: In einer retrospektiven Studie wurden 2 vergleichbare Kohorten mit je 20 Patienten identifiziert: eine manuelle Kohorte mit Wahl der Aufnahmeparameter entsprechend der Körperabmessungen (schlank – normal – kräftig) und eine ADC-Kohorte mit Parameterwahl durch das ADC-System (DoseRight 2.0™, Philips Healthcare). Für jeden Patienten wurden insgesamt 12 Bilder von 6 definierten anatomischen Höhen aus kontrastmittelgestützten CT-Untersuchungen von Thorax und Abdomen/Becken von 4 unabhängigen Auswertern bezüglich Bildrauschen und diagnostischer Akzeptanz mit einer 5-Punkte-Likert-Skala analysiert und auf objektiv gemessenes Bildrauschen untersucht. Die Strahlenexposition wurde für jeden Patienten aus den aufgezeichneten Expositionsdaten errechnet.
Ergebnisse: Die Nutzung des ADC-Systems reduzierte die durchschnittliche effektive Strahlendosis für die Patienten beim Thorax-CT um 36 % (3,2 vs. 4,9 mSv) und beim Abdomen/Becken-CT um 17 % (7,6 vs. 8,3 mSv). Das durchschnittliche objektive Bildrauschen war etwas geringer in der manuellen Kohorte (11,1 vs. 12,8 HU). Korrelierend dazu wurden diese Bilder auch subjektiv mit einer etwas besseren Note bewertet (4,4 vs. 4,2). Die diagnostische Akzeptanz wurde jedoch in beiden Kohorten nahezu gleich exzellent eingeschätzt (durchschnittliche Note 4,6 vs. 4,5).
Schlussfolgerung: Die Durchführung von CT-Untersuchungen mit einem ADC-System mit adäquater Rauschcharakteristik führt zu einer signifikanten Reduktion der Strahlenexposition für die Patienten. Dabei wird weiterhin eine exzellente Bildqualität erreicht.
-
References
- 1 Heyer CM, Peters S, Lemburg S. Die Struktur des Deutschen Röntgenkongresses und die wissenschaftliche Behandlung der Themen Strahlendosis und Dosisreduktion: eine Analyse der Jahre 1998–2008. Fortschr Röntgenstr 2009; 181: 1065-1072
- 2 Cohnen M, Wittsack HJ, Assadi S et al. Radiation exposure of patients in comprehensive computed tomography of the head in acute stroke. Am J Neuroradiol 2006; 27: 1741-1745
- 3 McCollough CH, Guimaraes L, Fletcher JG. In defense of body CT. Am J Roentgenol 2009; 193: 28-39
- 4 McCollough CH, Primak AN, Braun N et al. Strategies for reducing radiation dose in CT. Radiol Clin North Am 2009; 47: 27-40
- 5 Prokop M. Strahlendosis in der Computertomographie Risiko und Herausforderung. Der Radiologe 2008; 48: 229-242
- 6 Brix G, Nagel HD, Stamm G et al. Radiation exposure in multi-slice versus single-slice spiral CT – results of a nationwide survey. Eur Radiol 2003; 13: 1979-1991
- 7 Aldrich JE, Williams J. Change in patient doses from radiological examinations at the Vancouver General Hospital, 1991–2002. Can Assoc Radiol J 2005; 56: 94-99
- 8 Valentin J. Managing patient dose in multi-detector computed tomography (MDCT). ICRP Publication 102. Ann ICRP 2007; 37: 1-79, iii
- 9 Kopka L, Funke M, Breiter N et al. Anatomisch adaptierte Variation des Röhrenstroms bei der CT: Untersuchungen zur Strahlendosisreduktion und Bildqualität. Fortschr Röntgenstr 1995; 163: 383-387
- 10 Poll LW, Cohnen M, Brachten S et al. Reduktion der Strahlenexposition bei der Mehrschicht-Spiral CT des Herzens durch EKG-synchronisierte Modulation des Röhrenstromes („ECG Pulsing“): Phantommessungen. Fortschr Röntgenstr 2002; 174: 1500-1505
- 11 Tack D, De Maertelaer V, Gevenois PA et al. Dose reduction in multidetector CT using attenuation-based online tube current modulation. Am J Roentgenol 2003; 181: 331-334
- 12 Kalra MK, Maher MM, Toth TL et al. Comparison of Z-axis automatic tube current modulation technique with fixed tube current CT scanning of abdomen and pelvis. Radiology 2004; 232: 347-353
- 13 Mulkens TH, Bellinck P, Baeyaert M et al. Use of an automatic exposure control mechanism for dose optimization in multi-detector row CT examinations: clinical evaluation. Radiology 2005; 237: 213-223
- 14 Iball GR, Brettle DS, Moore AC. Assessment of tube current modulation in pelvic CT. Br J Radiol 2006; 79: 62-70
- 15 Keat N. CT Scanner Automatic Exposure Control Systems. Report 05016. London: Medicine and Healthcare Products Regulatory Agency; 2005
- 16 Söderberg M. Automatic Exposure Control in CT: An Investigation between Different Manufacturers Considering Radiation Dose and Image Quality. Master thesis Lund University; 2008
- 17 Nagel HD, Stumpp P, Kahn T et al. Performance of an automatic dose control system for CT: specifications and basic phantom tests. Fortschr Röntgenstr 2011; 183: 60-67
- 18 Gosch D, Stumpp P, Kahn T et al. Performance of an Automatic Dose Control System for CT: Anthropomorphic Phantom Studies. Fortschr Röntgenstr 2011; 183: 154-162
- 19 Bundesamt für Strahlenschutz. Bekanntmachung der aktualisierten diagnostischen Referenzwerte für diagnostische und interventionelle Röntgenuntersuchungen vom 22. Juni 2010. 2010
- 20 Rizzo S, Kalra M, Schmidt B et al. Comparison of angular and combined automatic tube current modulation techniques with constant tube current CT of the abdomen and pelvis. Am J Roentgenol 2006; 186: 673-679
- 21 European Commission. European guidelines on quality criteria for computed tomography. Report EUR 16262 EN. Luxembourg: Office for Official Publications of the European Communities; 1999
- 22 Kalra MK, Rizzo S, Maher MM et al. Chest CT performed with z-axis modulation: scanning protocol and radiation dose. Radiology 2005; 237: 303-308
- 23 Papadakis AE, Perisinakis K, Damilakis J. Automatic exposure control in pediatric and adult multidetector CT examinations: a phantom study on dose reduction and image quality. Med Phys 2008; 35: 4567-4576
- 24 Wilting JE, Zwartkruis A, van Leeuwen MS et al. A rational approach to dose reduction in CT: individualized scan protocols. Eur Radiol 2001; 11: 2627-2632
- 25 GE Healthcare. FeatherLight Imaging – Technology for dose optimization – CT procedure-based protocols (DOC0509389 Rev1 Ver1). Supplement to CT clarity – the Magazine of CT, spring 2009. GE Healthcare; 2008
- 26 Peng Y, Li J, Ma D et al. Use of automatic tube current modulation with a standardized noise index in young children undergoing chest computed tomography scans with 64-slice multidetector computed tomography. Acta Radiol 2009; 50: 1175-1181
- 27 Aldrich JE, Chang SD, Bilawich AM et al. Radiation dose in abdominal computed tomography: the role of patient size and the selection of tube current. Can Assoc Radiol J 2006; 57: 152-158