Abstract
Introduction: After completing a craniotomy, whenever possible, it is crucial to replace and fix the removed bone flap to the cranium; this in order to keep the brain's protection as well as for cosmetic purposes. Visible skull defects might cause patients psychosocial problems and, most importantly, expose the brain to accidental damage. A fixation device should not only provide optimal attachment of the flap to the skull but also allow fast bony healing to avoid possible pseudoarthrosis and/or osteolytic changes.
Methods: After performing 12 different craniotomies on 4 human cadaver heads the skull flaps were replaced using traditional sutures and a new skull fixation device; for each fixation technique a load-bearing test was performed and the results compared.
Results: Bone flaps fixed with the “Skull Grip” showed a strong fixation with optimal plastic deformation when compared to flaps held by sutures that showed less resistance to pressure and could be easily dislocated.
Conclusion: The “Skull Grip” has shown to be a reliable, effective, and stronger bone flap fixation superior to suturing technique.
Key words
cranial bone flap - skull flap fixation - titanium implants
References
- 1 Frenzel D, Lerch K-D. CranioFix titanium clamp for refixation of cranial bone flaps: Clinical long-term results in comparison with suture material. Zentralbl Neurochir Suppl. 1998; 24 ((abstr))
- 2 Spetzler RF. Bone flap fixation. A new technique. J Neurosurg. 1997; 87 475-476
- 3 Lerch K-D. Stereotactically guided microsurgical resection of deep-seated brain lesions: Minimization of established and development of new unconventional approaches. Minim Invas Neurosurg. 1995; 38 60-78
- 4 Lerch K-D, Schäfer D. Stereotactic microresection of deep-seated small cerebral vascular malformations. Acta Neurochir. 1994; 130 28-34
- 5 New PFL, Rosen BR, Brady TJ. Potential hazards and artifacts of ferromagnetic and nonferromagnetic surgical and dental materials and devices in nuclear magnetic resonance imaging. Radiology. 1983; 147 139-148
- 6 Shellock FG, Shellock VJ. Cranial bone flap fixation clamps: Compatibility at MR imaging. Radiology. 1998; 207 822-825
- 7 Shellock FG, Morisoli S, Kanal E. MR procedures and biomedical implants, materials and devices: 1993 update. Radiology. 1983; 189 587-599
Correspondence
Dr. S. Chibbaro
Service de Neurochirurgie
Hopital Lariboisière
2 rue Ambroise-Paré
75475 Paris Cedex 10
France
Telefon: +33/1/4995 81 46
Fax: +33/1/4995 81 55
eMail: schibbaro@hotmail.com