Inter- and intra-individual radiation dose variability in oncologic chest and abdominal computed tomography

Inter- und intraindividuelle Strahlendosisvariabilität bei onkologischen Computertomografien von Thorax und Abdomen

Autor*innen

  • Arian Felix Moradians

    1   Institute of Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Essen, Essen, Germany (Ringgold ID: RIN39081)
  • Daniel Rosok

    1   Institute of Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Essen, Essen, Germany (Ringgold ID: RIN39081)
  • Raya Serger

    1   Institute of Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Essen, Essen, Germany (Ringgold ID: RIN39081)
  • Laura Valentina Klüner

    1   Institute of Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Essen, Essen, Germany (Ringgold ID: RIN39081)
  • Marcel Drews

    1   Institute of Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Essen, Essen, Germany (Ringgold ID: RIN39081)
  • Florian Behr

    2   Department of Diagnostic and Interventional Radiology, University Hospital Bonn, Bonn, Germany (Ringgold ID: RIN39062)
  • Luca Salhöfer

    1   Institute of Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Essen, Essen, Germany (Ringgold ID: RIN39081)
  • Denise Bos

    1   Institute of Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Essen, Essen, Germany (Ringgold ID: RIN39081)
    3   Institute of Diagnostic Radiology, University Hospital Zurich Institute of Diagnostic Radiology, Zürich, Switzerland (Ringgold ID: RIN156265)
  • Michael Forsting

    1   Institute of Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Essen, Essen, Germany (Ringgold ID: RIN39081)
  • Johannes Haubold

    1   Institute of Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Essen, Essen, Germany (Ringgold ID: RIN39081)
  • Marcel Klaus Opitz

    1   Institute of Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Essen, Essen, Germany (Ringgold ID: RIN39081)
  • Sebastian Zensen

    1   Institute of Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Essen, Essen, Germany (Ringgold ID: RIN39081)

Abstract

Purpose

Computed tomography (CT) plays a central role in oncologic imaging, yet repeated examinations contribute substantially to cumulative radiation exposure. This study aimed to evaluate inter- and intra-individual radiation dose variability in chest and abdominal CT and the impact of CT device model and protocol standardization.

Materials and Methods

In this retrospective single-center study, 42441 CT scans from 4986 adult oncologic patients were analyzed. Dose metrics (CTDIvol, DLP, SSDE, effective dose) were extracted using automated dose monitoring. Inter- and intra-individual radiation dose variability was assessed across four CT device models and various protocol subtypes. Intra-individual radiation dose variability was calculated relative to the lowest dose per patient and compared across CT devices and protocol subtypes.

Results

Radiation dose varied substantially between devices, with CTDIvol differences of up to 2.4-fold in chest CT (2.98–7.26 mGy) and 1.7-fold in abdominal CT (5.26–8.77 mGy). The median intra-individual radiation dose variability was 93.7% (IQR 19.8–142.0%) in non-contrast chest CT, 66.3% (31.1–105.4%) in contrast-enhanced chest CT, 19.8% (11.8–32.3%) in non-contrast abdominal CT, and 28.2% (16.6–41.0%) in contrast-enhanced abdominal CT. When consecutive scans were performed on the same scanner, intra-individual radiation dose variability decreased to 14.7% (IQR 8.1–33.1%), 18.1% (9.5–31.3%), 11.7% (7.8–19.8%), and 15.3% (8.3–24.7%), respectively, indicating substantial device-specific effects.

Conclusion

Significant radiation dose variability persists in oncologic CT, both between and within patients, despite the use of standardized protocols. Device-adapted dose management and consistent device use may improve dose consistency, support optimization in oncologic imaging, and reduce radiation exposure.

Key Points

  • Substantial radiation dose variability persists across CT devices despite protocol standardization.

  • Intra-individual radiation dose variability is significant and highest in non-contrast chest CT.

  • Consistent use of the same scanner reduces intra-individual radiation dose variability significantly.

  • Internal diagnostic reference levels may improve radiation dose consistency and minimize exposure.

Citation Format

  • Moradians AF, Rosok D, Serger R et al. Inter- and intra-individual radiation dose variability in oncologic chest and abdominal computed tomography. Rofo 2026; DOI 10.1055/a-2786-2534

Zusammenfassung

Ziel

Die Computertomografie (CT) spielt eine zentrale Rolle in der onkologischen Bildgebung, jedoch tragen wiederholte Untersuchungen wesentlich zu einer kumulativen Strahlenexposition bei. Ziel dieser Studie war es, die inter- und intraindividuelle Variabilität der Strahlendosis bei Thorax- und Abdomen-CTs zu untersuchen und den Einfluss des CT-Gerätemodells sowie der Standardisierung der Protokolle zu analysieren.

Material und Methoden

In dieser retrospektiven Einzelzentrumsstudie wurden 42441 CT-Scans von 4986 erwachsenen onkologischen Patient:innen analysiert. Der volumengewichtete CT-Dosisindex (CTDIvol), das Dosis-Längen-Produkt (DLP), das Size-Specific Dose Estimate (SSDE) und die effektive Dosis (ED) wurden mittels eines automatisierten Dosisüberwachungssystems extrahiert. Die inter- und intraindividuelle Strahlendosisvariabilität wurde zwischen vier CT-Gerätemodellen und verschiedenen Protokollsubtypen evaluiert. Die intraindividuelle Dosisvariabilität wurde jeweils im Verhältnis zur niedrigsten individuellen Dosis berechnet und geräte- sowie protokollübergreifend verglichen.

Ergebnisse

Die Strahlendosis variierte deutlich zwischen den Geräten, mit bis zu 2,4-fachen Unterschieden des CTDIvol bei Thorax-CTs (2,98–7,26 mGy) und bis zu 1,7-fachen Unterschieden bei Abdomen-CTs (5,26–8,77 mGy). Die mediane intraindividuelle Dosisvariabilität betrug 93,7% (IQR 19,8–142,0%) für native Thorax-CTs, 66,3% (31,1–105,4%) für Thorax-CTs mit Kontrastmittel, 19,8% (11,8–32,3%) für native Abdomen-CTs und 28,2% (16,6–41,0%) für Abdomen-CTs mit Kontrastmittel. Wurden die Folgeuntersuchungen jeweils am gleichen Scanner durchgeführt, verringerte sich die intraindividuelle Variabilität auf 14,7% (IQR 8,1–33,1%), 18,1% (9,5–31,3%), 11,7% (7,8–19,8%) bzw. 15,3% (8,3–24,7%), was auf substanzielle gerätespezifische Faktoren hindeutet.

Schlussfolgerung

Trotz standardisierter Untersuchungsprotokolle besteht bei der onkologischen Computertomografie eine signifikante Variabilität der Strahlendosis, sowohl zwischen als auch innerhalb einzelner Patient:innen. Ein gerätespezifisches Dosismanagement und die konsequente Nutzung gleicher CT-Geräte könnten die Dosiskonstanz verbessern, zur Optimierung der onkologischen Bildgebung beitragen und die Strahlenexposition reduzieren.

Kernaussagen

  • Trotz standardisierter Protokolle besteht eine substanzielle Strahlendosisvariabilität zwischen unterschiedlichen CT-Geräten.

  • Die intraindividuelle Strahlendosisvariabilität ist signifikant und am höchsten bei nativen Thorax-CTs.

  • Eine konsequente Nutzung des gleichen CT-Geräts verringert die intraindividuelle Strahlendosisvariabilität deutlich.

  • Institutionsinterne Referenzwerte können zur Verbesserung der Dosiskonstanz und Reduktion der Strahlenexposition beitragen.



Publikationsverlauf

Eingereicht: 04. August 2025

Angenommen nach Revision: 04. Januar 2026

Artikel online veröffentlicht:
30. Januar 2026