Int J Sports Med 2022; 43(10): 875-880
DOI: 10.1055/a-1828-8499
Physiology & Biochemistry

Validity and Reliability of Two Near-infrared Spectroscopy Devices to Measure Resting Hemoglobin in Elite Cyclists

1   Department of Physiology, University of Valencia, Biophysics and Medical Physics Group (GIFIME), Valencia, Spain
,
2   Department of Physical Education and Sport, Faculty of Sport Sciences, University of Granada, Granada, Spain
,
Alejandro Javaloyes
3   Department of Sport Sciences, Sports Research Centre (Department of Sport Sciences), Universidad Miguel Hernández de Elche, Spain
,
Carlos Lorente-Casaus
1   Department of Physiology, University of Valencia, Biophysics and Medical Physics Group (GIFIME), Valencia, Spain
,
2   Department of Physical Education and Sport, Faculty of Sport Sciences, University of Granada, Granada, Spain
,
1   Department of Physiology, University of Valencia, Biophysics and Medical Physics Group (GIFIME), Valencia, Spain
4   Consejo Superior de Deportes (CSD), Red Española de Investigación del Rendimiento Deportivo en Ciclismo y Mujer (REDICYM), Madrid, Spain
5   Department of Physical Education and Sports, University of Valencia, Research Group in Sport Biomechanics (GIBD), Valencia, Spain
,
3   Department of Sport Sciences, Sports Research Centre (Department of Sport Sciences), Universidad Miguel Hernández de Elche, Spain
6   Universidad Europea de Madrid, Sport Sciences, Spain
› Institutsangaben

Abstract

A new method to monitor internal training load from muscle oxygen saturation using near-infrared spectroscopy could be of practical application for research and training purposes. This technology has been validated in different scientific fields, including sports science, and Humon Hex and Moxy are two leading brands. However, its relationship with hemoglobin has not been studied. Forty-eight professional cyclists, 19 men and 29 women, underwent a blood test to measure hemoglobin in the early morning. Immediately afterwards, hemoglobin and muscle oxygenation were monitored at rest by Moxy and Humon Hex on their right quadriceps (where the skinfold was measured). Venous blood hemoglobin was higher than the measurement for both devices (p<0.001). Both hemoglobin (p<0.001) and muscle oxygen saturation measurements (p<0.05) were higher in Humon Hex than for Moxy, and there was a reasonable reproducibility (ICC=0.35 for hemoglobin and 0.26 for muscle oxygen saturation). Skinfold had an inverse relationship with hemoglobin measurement (r=–0.85 p<0.001 for Humon Hex and r=–0.75 p<0.001 for Moxy). These findings suggest that resting hemoglobin data provided by these devices are not coincident with those of blood sample, and skinfold has an inverse relationship with blood hemoglobin measurement.



Publikationsverlauf

Eingereicht: 10. Januar 2022

Angenommen: 19. April 2022

Accepted Manuscript online:
19. April 2022

Artikel online veröffentlicht:
07. Juni 2022

© 2022. Thieme. All rights reserved.

Georg Thieme Verlag
Rüdigerstraße 14, 70469 Stuttgart, Germany

 
  • References

  • 1 Halson SL. Monitoring training load to understand fatigue in athletes. Sports Med 2014; 44: 139-147
  • 2 Faria EW, Parker DL, Faria IE. The science of cycling. Factors affecting performance – Part 2. Sports Med 2005; 35: 313-337
  • 3 Gómez-Carmona CD, Bastida-Castillo A, Rojas-Valverde D. et al. Lower-limb Dynamics of Muscle Oxygen Saturation During the Back-squat Exercise: Effects of Training Load and Effort Level. J Strength Cond Res 2020; 34: 1227-1236
  • 4 Yodh A, Chance B. Spectroscopy and imaging with diffusing light. Phys Today 1995; 48: 34-40
  • 5 Ferrari M, Wilson DA, Hanley DF. et al. Noninvasive determination of hemoglobin saturation in dogs by derivative near-infrared spectroscopy. Am J Physiol 1989; 256: H1493-H1499
  • 6 Mancini DM, Bolinger L, Li H. et al. Validation of near-infrared spectroscopy in humans. J Appl Physiol (1985) 1994; 77: 2740-2747
  • 7 Barstow TJ. Understanding near infrared spectroscopy and its application to skeletal muscle research. J Appl Physiol (1985) 2019; 126: 1360-1376
  • 8 Ferrari M, Quaresima V. Near infrared brain and muscle oximetry: From the discovery to current applications. J Infrared Spectrosc 2012; 20: 1-14
  • 9 Perrey S, Ferrari M. Muscle oximetry in Sports Science: A systematic review. Sports Med 2018; 48: 597-616
  • 10 Salas-Montoro J-A, Mateo-March M, Sánchez-Muñoz C. et al. Determination of second lactate threshold using near-infrared spectroscopy in elite cyclists. Int J Sports Med 2022; Online ahead of print. DOI: 10.1055/a-1738-0252.
  • 11 Faude O, Kindermann W, Meyer T. Lactate threshold concepts. Sports Med 2009; 39: 469-490
  • 12 McMorries RM, Joubert DP, Jones EJ. et al. A validation study of a noninvasive lactate threshold device. Int J Exerc Sci 2019; 12: 221-232
  • 13 Seiler KS, Kjerland GØ. Quantifying training intensity distribution in elite endurance athletes: Is there evidence for an “optimal” distribution?. Scand J Med Sci Sports 2006; 16: 49-56
  • 14 Seiler S. What is best practice for training intensity and duration distribution in endurance athletes?. Int J Sports Physiol Perform 2010; 5: 276-291
  • 15 Wittekind A, Cooper CE, Elwell CE. et al. Warm-up effects on muscle oxygenation, metabolism and sprint cycling performance. Eur J Appl Physiol 2012; 112: 3129-3139
  • 16 Skovereng K, Ettema G, van Beekvelt MCP. Oxygenation, local muscle oxygen consumption and joint specific power in cycling: The effect of cadence at a constant external work rate. Eur J Appl Physiol 2016; 116: 1207-1217
  • 17 Takaishi T, Ishida K, Katayama K. et al. Effect of cycling experience and pedal cadence on the near-infrared spectroscopy parameters. Med Sci Sports Exerc 2002; 34: 2062-2071
  • 18 Faiss R, Léger B, Vesin J-M. et al. Significant molecular and systemic adaptations after repeated sprint training in hypoxia. PLoS One 2013; 8: e56522
  • 19 Girard O, Buchheit M, Goodall S. et al. Oxygen availability affects exercise capacity, but not neuromuscular fatigue characteristics of knee extensors, during exhaustive intermittent cycling. Eur J Appl Physiol 2021; 121: 95-107
  • 20 Crum EM, O’Connor WJ, Loo Lvan. et al. Validity and reliability of the Moxy oxygen monitor during incremental cycling exercise. Eur J Sport Sci 2017; 17: 1037-1043
  • 21 Farzam P, Starkweather Z, Franceschini MA. Validation of a novel wearable, wireless technology to estimate oxygen levels and lactate threshold power in the exercising muscle. Physiol Rep 2018; 6: e13664
  • 22 Feldmann A, Schmitz RW, Erlacher D. Near-infrared spectroscopy-derived muscle oxygen saturation on a 0% to 100% scale: reliability and validity of the Moxy Monitor. J Biomed Opt 2019; 24: 115001
  • 23 Jaén-Carrillo D, Roche-Seruendo LE, Cartón-Llorente A. et al. Agreement between muscle oxygen saturation from two commercially available systems in endurance running: Moxy Monitor versus Humon Hex. Proceedings of the Institution of Mechanical Engineers, Part P: Journal of Sports Engineering and Technology 2021; DOI: 10.1177/17543371211015764.
  • 24 McManus CJ, Collison J, Cooper CE. Performance comparison of the MOXY and PortaMon near-infrared spectroscopy muscle oximeters at rest and during exercise. J Biomed Opt 2018; 23: 1-14
  • 25 Cohen J. Statistical Power Analysis for the Behavioral Sciences. 2nd Edition. New York: Lawrence Erlbaum Associates; 1988
  • 26 Weir JP. Quantifying test-retest reliability using the intraclass correlation coefficient and the SEM. J Strength Cond Res 2005; 19: 231-240
  • 27 O’Rourke N, Hatcher L, Stepanski EJ. A Step-by-step Approach to Using SAS for Univariate & Multivariate Statistics. SAS Institute. 2005
  • 28 Jöbsis FF. Noninvasive, infrared monitoring of cerebral and myocardial oxygen sufficiency and circulatory parameters. Science 1977; 198: 1264-1267
  • 29 Feldmann A, Erlacher D. Critical oxygenation: Can muscle oxygenation inform us about critical power?. Med Hypotheses 2021; 150: 110575
  • 30 Craig JC, Broxterman RM, Wilcox SL. et al. Effect of adipose tissue thickness, muscle site, and sex on near-infrared spectroscopy derived total-[hemoglobin+myoglobin]. J Appl Physiol (1985) 2017; 123: 1571-1578