Endoscopy 2020; 52(04): E148-E149
DOI: 10.1055/a-1025-1773
E-Videos
© Georg Thieme Verlag KG Stuttgart · New York

Surgical removal of a hepaticogastrostomy stent because of its recurrent dislocation into the esophagus

Hoonsub So
1   Division of Gastroenterology, Department of Internal Medicine, University of Ulsan College of Medicine, Asan Medical Center, Seoul, South Korea
,
Jae Hoon Lee
2   Division of Hepato-Biliary and Pancreatic Surgery, Department of Surgery, University of Ulsan College of Medicine, Asan Medical Center, Seoul, South Korea
,
Jin Hee Noh
1   Division of Gastroenterology, Department of Internal Medicine, University of Ulsan College of Medicine, Asan Medical Center, Seoul, South Korea
,
Sung Woo Ko
1   Division of Gastroenterology, Department of Internal Medicine, University of Ulsan College of Medicine, Asan Medical Center, Seoul, South Korea
,
Jun Seong Hwang
1   Division of Gastroenterology, Department of Internal Medicine, University of Ulsan College of Medicine, Asan Medical Center, Seoul, South Korea
,
Dongwook Oh
1   Division of Gastroenterology, Department of Internal Medicine, University of Ulsan College of Medicine, Asan Medical Center, Seoul, South Korea
,
Tae Jun Song
1   Division of Gastroenterology, Department of Internal Medicine, University of Ulsan College of Medicine, Asan Medical Center, Seoul, South Korea
› Author Affiliations
Further Information

Publication History

Publication Date:
15 November 2019 (online)

For endoscopic ultrasound-guided hepaticogastrostomy (EUS-HGS), an echoendoscope is advanced to the gastric cardia to puncture the ducts in segment 2 or 3 (B2 or B3) [1]. As proximal migration of a stent can be fatal, endoscopists usually prefer to use a long stent [2] [3]. We describe a case in which an HGS metal stent was surgically removed because of its recurrent dislocation into the esophagus, with accompanying vomiting.

A 68-year-old woman with hilar cholangiocarcinoma presented with cholangitis. An endoscopic retrograde cholangiopancreatography (ERCP) was performed, but after insertion of bilateral metal stents (8 mm × 8 cm; LCD Biliary, Taewoong, Korea) into the right anterior and B3 ducts, the B2 duct still showed isolated dilatation and the patient continued to have fever and jaundice. Endoscopic ultrasound (EUS)-guided antegrade stenting (8 mm × 6 cm; LCD Biliary, Taewoong, Korea) of the B2 duct across the previously inserted stent was therefore performed. An EUS-HGS was also performed using a fully covered metal stent (6 mm × 10 cm; Bonastent Biliary, Standard Sci Tech Inc., Korea) ([Fig. 1]).

Zoom Image
Fig. 1 Radiographic image showing the endoscopic ultrasound-guided hepaticogastrostomy (arrowhead) and the antegrade stent (arrow).

Following the stenting and HGS procedure, the patient’s symptoms improved, but the HGS stent migrated into the esophagus, which was accompanied by vomiting ([Fig. 2]). At first, the stent was pushed back into the gastric cavity using forceps; however, the patient experienced recurrent vomiting and radiographic imaging showed repeat stent dislocation into the esophagus. In an attempt to prevent further stent dislocation, another fully covered metal stent was inserted through the previous stent; however, dislocation still occurred. We then tried to anchor the stent to the stomach by placing a double-pigtail plastic stent (5 Fr × 7 cm, Zimmon; Cook Medical, USA) at the distal tip of the HGS stent ([Fig. 3]). Instead of providing anchorage however, the metal stent migrated into the stomach ([Fig. 4]). Therefore, we removed the plastic stent. Finally, the HGS stent was removed during a laparoscopy ([Video 1]). After stent removal, the patient no longer experienced vomiting.

Zoom Image
Fig. 2 Endoscopic view showing the dislocated hepaticogastrostomy stent in the esophagus.
Zoom Image
Fig. 3 Appearance after placement of a double-pigtail stent across the previously inserted metal stent: a on endoscopic view b on fluoroscopic view.
Zoom Image
Fig. 4 Abdominal radiograph showing proximal migration of hepaticogastrostomy stent owing to the anchoring plastic stent (arrow).

Video 1 Laparoscopic removal of the hepaticogastrostomy stent after dissection, with the fistula between the stomach and the liver being closed surgically.


Quality:

Endoscopists should be aware of this adverse event before performing EUS-HGS.

Endoscopy_UCTN_Code_CPL_1AL_2AD

Endoscopy E-Videos
https://eref.thieme.de/e-videos

Endoscopy E-Videos is a free access online section, reporting on interesting cases and new techniques in gastroenterological endoscopy. All papers include a high
quality video and all contributions are
freely accessible online.

This section has its own submission
website at
https://mc.manuscriptcentral.com/e-videos

 
  • References

  • 1 Ogura T, Higuchi K. Technical tips for endoscopic ultrasound-guided hepaticogastrostomy. World J Gastroenterol 2016; 22: 3945-3951
  • 2 Okuno N, Hara K, Mizuno N. et al. Stent migration into the peritoneal cavity following endoscopic ultrasound-guided hepaticogastrostomy. Endoscopy 2015; 47 (Suppl. 01) E311
  • 3 Uemura RS, Khan MA, Otoch JP. et al. EUS-guided choledochoduodenostomy versus hepaticogastrostomy: a systematic review and meta-analysis. J Clin Gastroenterol 2018; 52: 123-130