Fracture of 19G needle during transrectal endoscopic ultrasound-guided drainage of perirectal collection: an unusual complication

Gianfranco Donatelli
1   Unité d’Endoscopie Interventionnelle, Hôpital Privé des Peupliers, Ramsay Générale de Santé, Paris, France
,
Fabrizio Cereatti
2   Gastroenterologia ed Endoscopia Digestiva ASST Cremona, Cremona, Italy
› Author Affiliations
Further Information

Publication History

Publication Date:
09 May 2019 (online)

A 54-year-old woman was referred to our hospital for general sepsis. Computed tomography (CT) scan highlighted a collection of size 7 cm in the pouch of Douglas. The patient had a medical history of previous salpingo-oophorectomy coupled with appendicectomy for pelvic inflammatory disease.

Endoscopic ultrasound (EUS)-guided drainage of the collection was attempted. The perirectal collection was identified by transrectal EUS and was easily punctured with a 19G needle (Cook Medical). A guidewire was inserted into the collection under fluoroscopic control. Cautery of the entry site was performed with a 10-Fr cystotome, a 10-Fr double-pigtail plastic stent was delivered, and successful drainage of the cavity with outflow of purulent fluid into the rectum was obtained. At the end of the procedure re-evaluation of fluoroscopic images (after scope withdrawal) highlighted a metallic foreign body, 4 cm long, close to the stent ([Fig. 1]).

Zoom Image
Fig. 1 Fluoroscopic image showing a double-pigtail plastic stent draining a perirectal collection, and a metallic foreign body 4 cm in length close to the stent.

When the needle was advanced outside the external sheath, our suspicion of needle fracture was confirmed.

Lower endoscopy visualized the ruptured needle almost fixed against the rectal wall. The needle was retrieved using a foreign body forceps with attention paid to avoid injury to the rectum and the anal canal ([Video 1]).

Video 1 Fracture of a 19G needle during endoscopic ultrasound (EUS)-guided drainage of a perirectal collection. Early detection allowed endoscopic removal of the fractured part without complications.


Quality:

Needle fracture has been reported as a rare complication after EUS-fine needle aspiration (EUS-FNA) of pancreatic and endobronchial masses [1] [2]. In addition an incidental finding of intravascular migration of a fractured needle 6 months after the index EUS-FNA has been reported [3].

Risk factors have been supposed to be stiffness of the target lesion, bent position of the scope, or excessive elevator movement. Needle fracture can easily go unnoticed, since the integrity of the needle after EUS puncture is seldom checked. Whenever therapeutic EUS procedures are performed, concomitant use of fluoroscopy is advisable not only to guide the different procedural steps but also to detect early complications

Although needle fracture is very rare, its early detection is of extreme importance in order to avoid long-term sequelae and re-interventions.

Endoscopy_UCTN_Code_CPL_1AL_2AZ

Endoscopy E-Videos
https://eref.thieme.de/e-videos

Endoscopy E-Videos is a free access online section, reporting on interesting cases and new techniques in gastroenterological endoscopy. All papers include a high
quality video and all contributions are
freely accessible online.
This section has its own submission
website at
https://mc.manuscriptcentral.com/e-videos

 
  • References

  • 1 Rimbaș M, Attili F, Andrade Zurita S. et al. Fractured needle during endoscopic ultrasound-guided fine-needle aspiration of a pancreatic head mass. Endoscopy 2015; 47 (Suppl. 01) E432
  • 2 Vial MR, O’Connell JO, Grosu HB. et al. Needle fracture during endobronchial ultrasound-guided transbronchial needle aspiration. Am J Respir Crit Care Med 2016; 193: 213-214
  • 3 Lake E, Puleston J, Farquharson F. Intra-arterial migration of a fractured endoscopic needle. Endoscopy 2017; 49: E70-E72