Whole-Body CT in Multiple Trauma Patients: Clinically Adapted Usage of Differently Weighted CT Protocols
Artikel in mehreren Sprachen: English | deutschPublikationsverlauf
24. Dezember 2017
07. Mai 2018
Publikationsdatum:
12. November 2018 (online)
Abstract
Purpose Whole-body CT (wbCT) has been established as an internationally accepted diagnostic modality in multiple trauma. Until 2011, a uniform CT scanning protocol was used for all multiple trauma patients (pat.) at our hospital (OLD protocol = OP). In 2011, 2 new differently weighted protocols were introduced: TIME protocol (TP) for hemodynamically unstable pat. and DOSE protocol (DP) for pat. with stable vital parameters. The aim of this study was to compare the original “One-fits-all-concept” with the new, clinically oriented approach to wbCT.
Materials and Methods This study retrospectively evaluated 3 distinct wbCT protocols, looking at automatic exposure control variation (AEC; OP/TP) and arm positioning close to the body/overhead (TP/DP). The analysis included waist circumference (WC, cm), injury severity score (ISS), examination time (ET, min), image noise (IN), and effective dose (E, mSv). Normality of distribution was assessed with the Kolmogorov-Smirnov test. Data are given as median and range. Test of significance with Kruskal-Wallis test or Mann-Whitney-U-test. Level of significance: 0.05.
Results 308 pat. were included in the study (77 % m; age: 46 a, 18 – 90 a; WC: 93 cm, 66 – 145 cm). ISS was 14 (OP; n = 104; 0 – 75), 18 (TP; n = 102; 0 – 75) and 9 (DP; n = 102; 0 – 50). ET was 3.9 min (OP; 3.3 – 5.6 min), 4.1 min (TP; 2.8 – 7.2 min) and 7.7 min (DP; 6 – 10 min). IN showed no significant differences when comparing OP/TP but was significantly reduced in DP. For a wbCT (vertex to ischium), E could be reduced from 49.7 mSv to 35.4 mSv by optimizing AEC (OP/TP). Through the overhead repositioning of the arms in DP, a further reduction to 28.2 mSv was achieved.
Conclusion AEC and arm repositioning have a crucial influence on image quality and dose. The presented clinical approach is superior to the original concept.
Key Points:
-
The use of 2 differently weighted wbCT protocols allows a more flexible approach to the patient’s clinical presentation.
-
The clinically adapted concept presented in this study allows trauma care centers to reduce the collective dose.
-
Whole-body CT is leading to exposure to relevant radiation doses – further multicenter research is required.
Citation Format
-
Reske SU, Braunschweig R, Reske AW et al. Whole-Body CT in Multiple Trauma Patients: Clinically Adapted Usage of Differently Weighted CT Protocols. Fortschr Röntgenstr 2018; 190: 1141 – 1151
Key words
multiple trauma - whole-body CT - protocol optimization - arm positioning - dose reduction - clinically adapted approach* Board Member, Working Group Imaging Procedures of the Musculosceletal System of the German Röntgen Society, Berlin, Germany.
-
References
- 1 Statistisches Bundesamt. Gesundheit – Todesursachen in Deutschland 2013. 2014 https://www.destatis.de/DE/Publikationen/Thematisch/Gesundheit/Todesursachen/Todesursachen.html Abgerufen am: 03.12.2014
- 2 Huber-Wagner S, Lefering R, Qvick LM. Working Group on Polytrauma of the German Trauma Society. et al. Effect of whole-body CT during trauma resuscitation on survival: a retrospective, multicentre study. Lancet 2009; 373: 1455-1461
- 3 Salim A, Sangthong B, Martin M. et al. Whole body imaging in blunt multisystem trauma patients without obvious signs of injury: Results of a prospective study. Arch Surg 2006; 141: 468-475
- 4 Hoffstetter P, Herold T, Daneschnejad M. et al. Non-trauma-associated additional findings in whole-body CT examinations in patients with multiple trauma. Fortschr Röntgenstr 2008; 180: 120-126
- 5 TraumaRegister DGU® Jahresbericht 2009. 2009 http://www.traumaregister-dgu.de/de/service/downloads.html Abgerufen am: 15.11.2017
- 6 TraumaRegister DGU® Jahresbericht 2017. 2017 http://www.traumaregister-dgu.de/de/service/downloads.html Abgerufen am: 15.11.2017
- 7 Snyder GE. Whole-body imaging in blunt multisystem trauma patients who were never examined. Ann Emerg Med 2008; 52: 101-103
- 8 Fabian TC. Whole-body CT in multiple trauma. Lancet 2009; 373: 1408-1409
- 9 Ruchholtz S, Waydhas C, Schroeder T. et al. The value of computed tomography in the early treatment of seriously injured patients. Chirurg 2002; 73: 1005-1012
- 10 Brenner DJ, Hall EJ. Cancer Risks from CT Scans: Now We Have Data, What Next?. Radiology 2012; 265: 330-331
- 11 Laack TA, Thompson KM, Kofler JM. et al. Comparison of trauma mortality and estimated cancer mortality from computed tomography during initial evaluation of intermediate-risk trauma patients. J Trauma 2011; 70: 1362-1365
- 12 AUC TraumaRegister DGU ®. http://www.traumaregister-dgu.de/de/registerstruktur.html Abgerufen am 15.11.2017
- 13 Deutsche Gesellschaft für Unfallchirurgie. S3-Leitlinie Polytrauma/Schwerverletzten-Behandlung. 2016 http://www.awmf.org/leitlinien/detail/ll/012-019/ll-ansicht/autor.html Abgerufen am: 15.11.2017
- 14 Kloth JK, Kauczor HU, Hosch W. Imaging in the emergency room. Med Klin Intensivmed Notfmed 2011; 106: 82-88
- 15 Huber-Wagner S. et al. Whole-body CT Score – Kriterien zur Durchführung einer Ganzkörper-Computertomografie bei potentiell schwerverletzten Patienten. Berlin: Deutscher Kongress für Orthopädie und Unfallchirurgie (DKOU 2015); 2015
- 16 Gupta M, Schriger DL, Hiatt JR. et al. Selective use of computed tomography compared with routine whole body imaging in patients with blunt trauma. Ann Emerg Med 2011; 58: 407-416.e15
- 17 Wurmb TE, Kenn W. The role of early multislice computed tomography in major trauma. Trauma 2012; 14: 301-312
- 18 Baker SP, O’Neill B. The injury severity score: an update. J Trauma 1976; 16: 882-885
- 19 Karlo C, Gnannt R, Frauenfelder T. et al. Whole-body CT in polytrauma patients: effect of arm positioning on thoracic and abdominal image quality. Emerg Radiol 2011; 18: 285-293
- 20 Brink M, de Lange F, Oostveen LJ. et al. Arm Raising at Exposure-controlled Multidetector Trauma CT of Thoracoabdominal Region: Higher Image Quality, Lower Radiation Dose. Radiology 2008; 249: 661-670
- 21 Huda W, Magill D, He W. CT effective dose per dose length product using ICRP 103 weighting factors. Med Phys 2011; 38: 1261-1265
- 22 International Commission on Radiological Protection. The 2007 Recommendations of the ICRP. ICRP Publication 103. Ann ICRP; 2007 37. (2–4).
- 23 Bundesamt für Strahlenschutz Deutschland. Diagnostische Referenzwerte für diagnostische und interventionelle Röntgenanwendungen. 2016 http://www.bfs.de/DE/themen/ion/anwendung-medizin/diagnostik/referenzwerte/referenzwerte_node.html Abgerufen am: 15.11.2017
- 24 Yaniv G, Portnoy O, Simon D. et al. Revised protocol for whole-body CT for multi-trauma patients applying triphasic injection followed by a single-pass scan on a 64-MDCT. Clin Radiol 2013; 68: 668-675
- 25 Fanucci E, Fiaschetti V, Rotili A. et al. Whole body 16-row multislice CT in emergency room: effects of different protocols on scanning time, image quality and radiation exposure. Emerg Radiol 2007; 13: 251-257
- 26 Bayer J, Pache G, Strohm PC. et al. Influence of arm positioning on radiation dose for whole body computed tomography in trauma patients. J Trauma 2011; 70: 900-905
- 27 Loewenhardt B, Buhl M, Gries A. et al. Radiation exposure in whole-body computed tomography of multiple trauma patients: bearing devices and patient positioning. Injury 2012; 43: 67-72
- 28 Kalender WA, Buchenau S, Deak P. et al. Technical approaches to the optimisation of CT. Phys Med 2008; 24: 71-79
- 29 Harrieder A, Geyer LL, Körner M. et al. Evaluation of radiation dose in 64-row whole-body CT of multiple injured patients compared to 4-row CT. Fortschr Röntgenstr 2012; 184: 443-449
- 30 International Commission on Radiological Protection. 1990 Recommendations of the ICRP. ICRP Publication 60. Ann ICRP; 1991 21. (1–3)
- 31 Mettler FA, Huda W, Yoshizumi TT. et al. Effective Doses in Radiology and Diagnostic Nucelar Medicine: A Catalog. Radiology 2008; 248: 254-263
- 32 Loewenhardt B, Huettinger R, Reinert M. et al. Dose effects and image quality: Is there any influence by bearing devices in whole-body computed tomography in trauma patients?. Injury 2014; 45: 170-175