Ultraschall Med 2016; 37(02): 140-156
DOI: 10.1055/s-0041-107766
Continuing Medical Education
© Georg Thieme Verlag KG Stuttgart · New York

Ultrasound Artifacts – Part 1

Artefakte bei der Sonografie – Teil 1
J. A. Bönhof
Further Information

Publication History

17 August 2015

19 September 2015

Publication Date:
13 November 2015 (online)

Abstract

Knowledge of artifacts is essential for the competent use of ultrasound. Artifacts are method-based and should be differentiated from image errors of another genesis. They are logical and occur because the conditions required for image generation do not fully correspond to the reality. Artifacts occur due to disregard of the true dimensions of sound lobes (slice-thickness artifacts and bow artifacts, range ambiguities) and due to different types of mirroring with different appearances. There are also comet-tail-like artifacts such as comet-tail and ring-down artifacts.

Zusammenfassung

Kenntnisse über Artefakte sind für jeden wichtig, der Sonografie kompetent betreiben will. Artefakte sind methodeninhärent und von Bildfehlern anderer Genese zu unterscheiden; sie sind logisch und entstehen, weil die Voraussetzungen, die zum Bildaufbau gemacht werden müssen, nicht vollständig der Wirklichkeit entsprechen. Es gibt Artefakte u. a. durch Vernachlässigung der wahren Ausmaße der Schallkeulen (Schichtdicken- und Bogenartefakte, sowie Tiefenartefakte, „Range ambiguities“) und durch Spiegelungen verschiedener Art, mit unterschiedlichem Aussehen; zudem noch Schweifartefakte wie Comet tail- und Ring down-Artefakt.